Ali Reza Pournajafian1, Mohammad Reza Ghodraty1, Seyed Hamid Reza Faiz2, Poupak Rahimzadeh2, Hamidreza Goodarzynejad3, Enseyeh Dogmehchi2.
Abstract
BACKGROUND: To determine if the GlideScope® videolaryngoscope (GVL) could attenuate the hemodynamic responses to orotracheal intubation compared with conventional Macintosh laryngoscope.Entities:
Keywords: Hemodynamic Responses; Intubation; Laryngoscopes
Year: 2014 PMID: 24910788 PMCID: PMC4028761 DOI: 10.5812/ircmj.12334
Source DB: PubMed Journal: Iran Red Crescent Med J ISSN: 2074-1804 Impact factor: 0.611
Figure 1.A GlideScope Video Laryngoscope With Its Specific Rigid Stylet
Figure 2.Flow Diagram of Patients Recruitment
Baseline Characteristics of the Study Population (n= 95) [a]
| All | Macintosh (n = 49) | GlideScope (n = 46) | P Value [ | |
|---|---|---|---|---|
|
| 34.8 ± 11.1 | 33.7 ± 10.6 | 36.1 ± 11.6 | 0.292 |
|
| 24.5 ± 3.4 | 24.1 ± 3.3 | 24.9 ± 3.5 | 0.226 |
|
| 166.7 ± 8.2 | 165.9 ± 7.5 | 167.5 ± 8.9 | 0.345 |
|
| 67.9 ± 9.8 | 66.2 ± 9.8 | 69.7 ± 9.1 | 0.079 |
|
| 0.503 | |||
| Male | 38 (40.0) | 18 (36.7) | 20 (43.5) | |
| Female | 57 (60.0) | 31 (63.3) | 26 (56.5) | |
|
| 0.133 | |||
| I | 72 (75.8) | 34 (69.4) | 38 (82.6) | |
| II | 23 (24.2) | 15 (30.6) | 8 (17.4) | |
|
| 0.008 | |||
| I | 80 (84.2) | 46 (93.9) | 34 (73.9) | |
| II | 15 (15.8) | 3 (6.1) | 12 (26.1) | |
|
| 0.383 | |||
| I | 66 (69.5) | 36 (73.5) | 30 (65.2) | |
| II | 29 (30.5) | 13 (26.5) | 16 (34.8) |
a Data are presented in Mean ± SD or No. (%).
b P values for Macintosh vs. GlideScope ASA (American Society of Anesthesiologists).
Figure 3.The Box-plot Graph of Comparison of Intubation Time (Seconds) Between Macintosh Group and Glidescope Group
Comparison of Hemodynamic Changes Associated With Orotracheal Intubation Between the Two Groups [a, b]
| Variables | Baseline | Post-induction | Pre-laryngoscopy | After Intubation, min | P Value | ||
|---|---|---|---|---|---|---|---|
| 1 | 3 | 5 | |||||
|
| 0.863 | ||||||
| Macintosh | 125.3 ± 10.8 | 111.5 ± 12.9 [ | 109.1 ± 14.1 | 135.2 ± 19.5 [ | 128.7 ± 16.8 | 118.3 ± 13.9 | |
| GlideScope | 127.0 ± 12.3 | 109.4 ± 16.7 [ | 107.0 ± 14.5 | 140.6 ± 20.8 [ | 128.5 ± 12.9 | 118.2 ± 16.6 | |
|
| 0.410 | ||||||
| Macintosh | 80.2 ± 9.0 | 70.5 ± 12.1 [ | 67.4 ± 12.6 | 91.0 ± 14.6 [ | 86.8 ± 16.0 | 77.3 ± 12.3 | |
| GlideScope | 80.2 ± 7.4 | 66.5 ± 12.9 [ | 66.0 ± 10.9 | 94.5 ± 13.0 [ | 82.2 ± 13.0 | 72.7 ± 12.1 | |
|
| 0.811 | ||||||
| Macintosh | 94.5 ± 8.7 | 83.8 ± 11.4 [ | 80.2 ± 11.8 | 105.4 ± 16.3 [ | 99.8 ± 15.2 | 90.5 ± 11.9 | |
| GlideScope | 94.4 ± 8.1 | 82.3 ± 14.8 [ | 79.0 ± 11.3 | 110.4 ± 15.1 [ | 97.4 ± 15.1 | 88.1 ± 12.0 | |
|
| 0.160 | ||||||
| Macintosh | 92.0 ± 16.1 | 88.9 ± 13.5 | 83.1 ± 12.1 | 92.1 ± 14.3 [ | 88.5 ± 14.9 | 83.4 ± 12.8 | |
| GlideScope | 84.5 ± 18.7 | 86.0 ± 17.2 | 79.0 ± 13.1 | 90.5 ± 15.0 [ | 85.8 ± 15.0 | 80.6 ± 14.9 | |
a Abbreviations: AP, arterial pressure; BP, blood pressure; SD, standard deviation.
b Data are presented as Mean ± SD.
c P value < 0.05 as compared to respective baseline values.
d P value < 0.05 as compared to respective pre-laryngoscopy values.
Figure 4.Changes of Mean Rate Pressure Product (Rpp) Before and After Intubation (Mean ± SD) in Two Various Laryngoscopy Methods