| Literature DB >> 24892346 |
Sarah J Castillo1, Richard Castillo, Peter Balter, Tinsu Pan, Geoffrey Ibbott, Brian Hobbs, Ying Yuan, Thomas Guerrero.
Abstract
The benefits of four-dimensional computed tomography (4D CT) are limited by the presence of artifacts that remain difficult to quantify. A correlation-based metric previously proposed for ciné 4D CT artifact identification was further validated as an independent artifact evaluator by using a novel qualitative assessment featuring a group of observers reaching a consensus decision on artifact location and magnitude. The consensus group evaluated ten ciné 4D CT scans for artifacts over each breathing phase of coronal lung views assuming one artifact per couch location. Each artifact was assigned a magnitude score of 1-5, 1 indicating lowest severity and 5 indicating highest severity. Consensus group results served as the ground truth for assessment of the correlation metric. The ten patients were split into two cohorts; cohort 1 generated an artifact identification threshold derived from receiver operating characteristic analysis using the Youden Index, while cohort 2 generated sensitivity and specificity values from application of the artifact threshold. The Pearson correlation coefficient was calculated between the correlation metric values and the consensus group scores for both cohorts. The average sensitivity and specificity values found with application of the artifact threshold were 0.703 and 0.476, respectively. The correlation coefficients of artifact magnitudes for cohort 1 and 2 were 0.80 and 0.61, respectively, (p < 0.001 for both); these correlation coefficients included a few scans with only two of the five possible magnitude scores. Artifact incidence was associated with breathing phase (p < 0.002), with presentation less likely near maximum exhale. Overall, the correlation metric allowed accurate and automated artifact identification. The consensus group evaluation resulted in efficient qualitative scoring, reduced interobserver variation, and provided consistent identification of artifact location and magnitudes.Entities:
Mesh:
Year: 2014 PMID: 24892346 PMCID: PMC4048877 DOI: 10.1120/jacmp.v15i3.4718
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Artifact evaluation software showing: T0% of a 4D CT scan (left) with a highlighted identified artifact at couch position 13 indicating a saved artifact location; and corresponding deep‐inspiration breath‐hold scan (right) used as an anatomic reference for artifact identification in the 4D CT (left).
Figure 2Example of coronal slices taken from T10% (left) and T90% (right) of patient (NCM values shown in Fig. 3.), with couch positions indicated on the left side of each coronal view and consensus group identified scores per couch position shown on the right side of each coronal view. The artifact in T10% at couch position 11 was scored as a more severe artifact based on the higher interference of the artifact with anatomy.
Figure 3Normalized correlation metric (NCM) vs. couch positions for patient case . Breathing phases T10% and T90% are displayed for comparison with coronal views of T10% and T90% in Fig. 2.
Figure 4Estimated probability of artifacts as a function of phase and associated 95% confidence intervals (grey bars). The risk of an artifact decreased for exhale phase images. The p‐value derives from the likelihood ratio test of the global null hypothesis of the absence of association with phase.
Figure 5Example ROC curve from case of cohort 1, with area under the curve (AUC), Youden index, and corresponding NCM threshold indicated.
Cohort 1 ROC parameters
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| AUC (area under curve) | 0.756 | 0.525 | 0.769 | 0.709 | 0.801 |
| Youden's index | 0.446 | 0.103 | 0.507 | 0.461 | 0.545 |
| NCM threshold | 125% | 73% | 93% | 93% | 81% |