| Literature DB >> 24876416 |
Kimberly Englert1, Katrina Ruedy2, Julie Coffey3, Kimberly Caswell3, Amy Steffen4, Lucy Levandoski5.
Abstract
The purpose of this article is to describe challenges associated with successful use of continuous glucose monitoring (CGM) by young children with type 1 diabetes (T1D) and to detail the techniques and products used to improve the duration of sensor wear. The DirecNet Study Group conducted 2 studies in 169 children with T1D between the ages of 1 and 9 years who were instructed to wear a CGM device daily. Problems related to skin irritation and sensor adhesiveness in these young children presented challenges to daily use of the CGM. Study coordinators instituted a variety of techniques using commercially available products to attempt to overcome these problems. Three primary factors that contributed to reduced CGM use were identified: the limited body surface area in smaller children, ambient temperature and humidity, as well as the type and duration of physical activity. Using supplemental products to minimize the impact of these factors resulted in improved adherence and reduced skin irritation. Achieving satisfactory adhesion of the CGM sensor and transmitter may involve finding the right supplemental product or combination of products through trial and error. Optimizing adhesion and minimizing skin irritation can significantly improve duration of use and tolerability of CGM devices by young children.Entities:
Keywords: CGM; adhesive; children; irritation
Mesh:
Substances:
Year: 2014 PMID: 24876416 PMCID: PMC4764227 DOI: 10.1177/1932296814529893
Source DB: PubMed Journal: J Diabetes Sci Technol ISSN: 1932-2968