| Literature DB >> 24872611 |
L Nithya1, N Arunai Nambi Raj2, Arulraj Kumar1, Sasikumar Rathinamuthu1, Manish Bhushan Pandey1.
Abstract
The aim of this study was to compare the various dosimetric parameters of dynamic multileaf collimator (MLC) intensity modulated radiation therapy (IMRT) plans with volumetric modulated arc therapy (VMAT) plans for base of tongue cases. All plans were done in Monaco planning system for Elekta synergy linear accelerator with 80 MLC. IMRT plans were planned with nine stationary beams, and VMAT plans were done for 360° arc with single arc or dual arc. The dose to the planning target volumes (PTV) for 70, 63, and 56 Gy was compared. The dose to 95, 98, and 50% volume of PTV were analyzed. The homogeneity index (HI) and the conformity index (CI) of the PTV70 were also analyzed. IMRT and VMAT plan showed similar dose coverage, HI, and CI. Maximum dose and dose to 1-cc volume of spinal cord, planning risk volume (PRV) cord, and brain stem were compared. IMRT plan and VMAT plan showed similar results except for the 1 cc of PRV cord that received slightly higher dose in VMAT plan. Mean dose and dose to 50% volume of right and left parotid glands were analyzed. VMAT plan gave better sparing of parotid glands than IMRT. In normal tissue dose analyses VMAT was better than IMRT. The number of monitor units (MU) required for delivering the good quality of the plan and the time required to deliver the plan for IMRT and VMAT were compared. The number of MUs for VMAT was higher than that of IMRT plans. However, the delivery time was reduced by a factor of two for VMAT compared with IMRT. VMAT plans yielded good quality of the plan compared with IMRT, resulting in reduced treatment time and improved efficiency for base of tongue cases.Entities:
Keywords: Base of tongue; intensity-modulated radiation therapy; volumetric modulated arc therapy
Year: 2014 PMID: 24872611 PMCID: PMC4035616 DOI: 10.4103/0971-6203.131288
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Treatment planning objectives
Figure 1Isofill comparison of IMRT (left side) and VMAT (right side) for one base of tongue case. Red color: 68 Gy; yellow color: 63 Gy; cyan color: 56 Gy
The dosimetric comparison of IMRT and VMAT
Figure 2DVH comparison of IMRT and VMAT of PTVs. Solid lines: VMAT plan; dashed line: IMRT plan
Figure 3DVH comparison of spinal cord, brain stem, and PRV cord. Solid lines: VMAT; dashed lines: IMRT; blue color: Spinal cord; brown color: PRV cord; orange color: Brainstem
Figure 4DVH comparison of parotid glands. Solid lines: VMAT; dashed lines: IMRT; blue color: left parotid; green color: right parotid. DVH: Dose volume histogram, IMRT: Intensity-modulated radiation therapy, VMAT: Volumetric-modulated arc therapy