Literature DB >> 24866836

Effect of different sterilization modes on the surface morphology, ion release, and bone reaction of retrieved micro-implants.

Noha El-Wassefy1, Abeer El-Fallal, Mahasen Taha.   

Abstract

OBJECTIVE: To compare as-received and sterilized micro-implants in order to assess the prospects of reusing them.
MATERIALS AND METHODS: Forty micro-implants from a single manufacturing lot were used in the study. Thirty were retrieved from patients after successful service in their mouth and with no signs of failure. The retrieved micro-implants were divided into three groups, according to method of sterilization: autoclave, gamma radiation, or ultraviolet radiation. All groups were subjected to scanning electron microscope analysis for surface morphology assessment. The specimens were immersed in a standard simulated body-fluid solution kept at 37°C in an incubator; the solution was then withdrawn at 24 hours and 30 days to evaluate aluminum and vanadium ion release by atomic absorption spectrophotometer in parts per billion. The micro-implants were then surgically implanted into the tibia of rabbits for a 1-month healing period, and the bone-implant blocks were processed for routine histologic examination.
RESULTS: This study revealed that sterilized micro-implants had altered surface topography, different ion release values, and different histologic cell reactions than the as-received micro-implants.
CONCLUSIONS: Within the limitations of this study, it can be concluded that retrieved self-drilling micro-implants have tip sharpness variations that require correction before insertion by bone drilling. The autoclave-sterilized micro-implants showed better histologic results than micro-implants sterilized by gamma or ultraviolet rays.

Entities:  

Keywords:  Bone interface histology; Ion release; Micro-implants; Sterilizations; Surface morphology

Mesh:

Substances:

Year:  2015        PMID: 24866836     DOI: 10.2319/012014-56.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  3 in total

1.  A comparison of decontamination methods of tried-in preformed metal crowns: an in-vivo study.

Authors:  V Darshan; K R Indushekar; B G Saraf; N Sheoran; B Sharma; D Sardana
Journal:  Eur Arch Paediatr Dent       Date:  2019-03-25

2.  Effect of steam and dry heat sterilization on the insertion and fracture torque of orthodontic miniscrews.

Authors:  Shiva Alavi; Farnaz Asadi; Seyed Amir Hossein Raji; Soroor Samie
Journal:  Dent Res J (Isfahan)       Date:  2020-05-23

3.  An evaluation of two types of nickel-titanium wires in terms of micromorphology and nickel ions' release following oral environment exposure.

Authors:  Abdul Razzak A Ghazal; Mohammad Y Hajeer; Rabab Al-Sabbagh; Ibrahim Alghoraibi; Ahmad Aldiry
Journal:  Prog Orthod       Date:  2015-05-20       Impact factor: 2.750