Dose and image quality of cone-beam computed tomography as compared with conventional multislice computed tomography in abdominal imaging.

Alexander A Schegerer1, Ursula Lechel, Manuel Ritter, Gerald Weisser, Christian Fink, Gunnar Brix.   

Abstract

OBJECTIVES: Recent technical developments have facilitated the application of cone-beam computed tomography (CBCT) for interventional and intraoperative imaging. The aim of this study was to compare the radiation doses and image quality in CBCT with those of conventional multislice spiral computed tomography (MSCT) for abdominal and genitourinary imaging.
METHODS: Different CBCT and MSCT protocols for imaging soft tissues and hard-contrast objects at different dose levels were investigated in this study. Local skin and organ doses were measured with thermoluminescent dosimeters placed in an anthropomorphic phantom. Moreover, the contrast-to-noise ratio, the noise-power spectrum, and the high-contrast resolution derived from the modulation transfer function were determined in a phantom with the same absorption properties as those of anthropomorphic phantom.
RESULTS: The effective dose of the examined abdominal/genitourinary CBCT protocols ranged between 0.35 mSv and 18.1 mSv. As compared with MSCT, the local skin dose of CBCT examinations could locally reach much higher doses up to 190 mGy. The effective dose necessary to realize the same contrast-to-noise ratio with CBCT and MSCT depended on the MSCT convolution kernel: the MSCT dose was smaller than the corresponding CBCT dose for a soft kernel but higher than that for a hard kernel. The noise-power spectrum of the CBCT images at tube voltages of 85/90 kV(p) is at least half of that of images measured at 103/115 kV(p) at any arbitrarily chosen spatial frequency. Although the pixel size and slice thickness of CBCT were half of those of the MSCT images, high-contrast resolution was inferior to the MSCT images reconstructed with a hard convolution kernel.
CONCLUSIONS: As compared with MSCT using a medium-hard convolution kernel, CBCT produces images at medium noise levels and, simultaneously, medium spatial resolution at approximately the same dose. It is well suited for visualizing hard-contrast objects in the abdomen with relatively low image noise and patient dose. For the detection of low-contrast objects at standard tube voltages of approximately 120 kV(p), however, MSCT should be preferred.

Entities:  

Mesh:

Year:  2014        PMID: 24853071     DOI: 10.1097/RLI.0000000000000069

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  9 in total

1.  Radiation exposure of the interventional radiologist during percutaneous biopsy using a multiaxis interventional C-arm CT system with 3D laser guidance: a phantom study.

Authors:  Nils Rathmann; Michael Kostrzewa; Kerim Kara; Soenke Bartling; Holger Haubenreisser; Stefan O Schoenberg; Steffen J Diehl
Journal:  Br J Radiol       Date:  2015-09-15       Impact factor: 3.039

2.  Prototype metal artefact reduction algorithm in flat panel computed tomography - evaluation in patients undergoing transarterial hepatic radioembolisation.

Authors:  Qeumars Mustafa Hamie; Adrian Raoul Kobe; Leif Mietzsch; Michael Manhart; Gilbert Dominique Puippe; Thomas Pfammatter; Roman Guggenberger
Journal:  Eur Radiol       Date:  2017-07-14       Impact factor: 5.315

3.  Balloon pulmonary angioplasty: applicability of C-Arm CT for procedure guidance.

Authors:  Jan B Hinrichs; Julius Renne; Marius M Hoeper; Karen M Olsson; Frank K Wacker; Bernhard C Meyer
Journal:  Eur Radiol       Date:  2016-02-23       Impact factor: 5.315

4.  C-Arm Cone-Beam CT Virtual Navigation-Guided Percutaneous Mediastinal Mass Biopsy: Diagnostic Accuracy and Complications.

Authors:  Hyungjin Kim; Chang Min Park; Sang Min Lee; Jin Mo Goo
Journal:  Eur Radiol       Date:  2015-04-28       Impact factor: 5.315

Review 5.  Interventional Oncology in Hepatocellular Carcinoma: Progress Through Innovation.

Authors:  Lin Mu; Julius Chapiro; Jeremiah Stringam; Jean-François Geschwind
Journal:  Cancer J       Date:  2016 Nov/Dec       Impact factor: 3.360

6.  Initial evaluation of image performance of a 3-D x-ray system: phantom-based comparison of 3-D tomography with conventional computed tomography.

Authors:  Robyn Melanie Benz; Meritxell Alzamora Garcia; Felix Amsler; Johannes Voigt; Andreas Fieselmann; Anna Lucja Falkowski; Bram Stieltjes; Anna Hirschmann
Journal:  J Med Imaging (Bellingham)       Date:  2018-03-06

7.  Three-dimensional assessment of the favorability of maxillary posterior teeth intrusion in different facial patterns limited by the vertical relationship with the maxillary sinus floor.

Authors:  Ehab A Abdulghani; Maged Sultan Alhammadi; Abeer A Al-Sosowa; Abeer A Almashraqi; Hasan M Sharhan; Hanan Al-Fakeh; BaoCheng Cao
Journal:  Clin Oral Investig       Date:  2022-03-20       Impact factor: 3.606

8.  CT-Guided Percutaneous Transthoracic Needle Biopsy Using the Additional Laser Guidance System by a Pulmonologist with 2 Years of Experience in CT-Guided Percutaneous Transthoracic Needle Biopsy.

Authors:  Min Cheol Jeon; Ju Ock Kim; Sung Soo Jung; Hee Sun Park; Jeong Eun Lee; Jae Young Moon; Chae Uk Chung; Da Hyun Kang; Dong Il Park
Journal:  Tuberc Respir Dis (Seoul)       Date:  2018-06-19

9.  Short-term outcomes of radiofrequency ablation for hepatocellular carcinoma using cone-beam computed tomography for planning and image guidance.

Authors:  Xue-Song Yao; Dong Yan; Xian-Xian Jiang; Xiao Li; Hui-Ying Zeng; Huai Li
Journal:  World J Clin Cases       Date:  2021-03-06       Impact factor: 1.337

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.