The new mid-scala electrode array: a radiologic and histologic study in human temporal bones.

Frederike Hassepass1, Stefan Bulla, Wolfgang Maier, Roland Laszig, Susan Arndt, Rainer Beck, Lousia Traser, Antje Aschendorff.   

Abstract

HYPOTHESIS: To analyze the quality of insertion of the newly developed midscala (MS) electrode, which targets a midscalar electrode position to reduce the risk of trauma to the lateral wall and the modiolus.
BACKGROUND: Modern cochlear implant surgery aims for a safe intracochlear placement of electrode arrays with an ongoing debate regarding cochleostomy or round window (RW) insertion and the use of lateral wall or perimodiolar electrode placement. Intracochlear trauma after insertion of different electrodes depends on insertion mode and electrode design and may result in trauma to the delicate structures of the cochlear.
METHODS: We performed a temporal bone (TB) trial with insertion of the MS electrode in n = 20 TB's after a mastoidectomy and posterior tympanotomy. Insertion was performed either via the RW or a cochleostomy. Electrode positioning, length of insertion, and angle of insertion were analyzed with rotational tomography (RT). TBs were histologically analyzed. Results of RT and histology were compared.
RESULTS: Scala tympani (ST) insertion could be accomplished reliably by both RW and via a cochleostomy approach. In 20 TBs, 1 scala vestibuli insertion, 1 incomplete (ST), and 1 elevation of basilar membrane were depicted. No trauma was found in 94.7% of all ST insertions. RT allowed determination of the intracochlear electrode position, which was specified by histologic sectioning.
CONCLUSION: The new MS electrode seems to fulfill reliable atraumatic intracochlear placement via RW and cochleostomy approaches. RT is available for evaluation of intracochlear electrode position, serving as a potential quality control instrument in human implantation.

Entities:  

Mesh:

Year:  2014        PMID: 24836594     DOI: 10.1097/MAO.0000000000000412

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  14 in total

1.  Multicenter surgical experience evaluation on the Mid-Scala electrode and insertion tools.

Authors:  Dzemal Gazibegovic; Eva M Bero
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-08-11       Impact factor: 2.503

2.  Insertion characteristics and placement of the Mid-Scala electrode array in human temporal bones using detailed cone beam computed tomography.

Authors:  Aarno Dietz; Dzemal Gazibegovic; Jyrki Tervaniemi; Veli-Matti Vartiainen; Heikki Löppönen
Journal:  Eur Arch Otorhinolaryngol       Date:  2016-05-19       Impact factor: 2.503

3.  Microtopographical features generated by photopolymerization recruit RhoA/ROCK through TRPV1 to direct cell and neurite growth.

Authors:  Shufeng Li; Bradley W Tuft; Linjing Xu; Marc A Polacco; Joseph C Clarke; C Allan Guymon; Marlan R Hansen
Journal:  Biomaterials       Date:  2015-03-12       Impact factor: 12.479

4.  A New Application of CBCT Image Fusion in Temporal Bone Studies.

Authors:  Matti Iso-Mustajärvi; Sini Sipari; Antti Lehtimäki; Jyrki Tervaniemi; Heikki Löppönen; Aarno Dietz
Journal:  J Int Adv Otol       Date:  2019-12       Impact factor: 1.017

5.  Predictive factors for short- and long-term hearing preservation in cochlear implantation with conventional-length electrodes.

Authors:  George B Wanna; Brendan P O'Connell; David O Francis; Rene H Gifford; Jacob B Hunter; Jourdan T Holder; Marc L Bennett; Alejandro Rivas; Robert F Labadie; David S Haynes
Journal:  Laryngoscope       Date:  2017-06-22       Impact factor: 3.325

6.  Electrode Location and Angular Insertion Depth Are Predictors of Audiologic Outcomes in Cochlear Implantation.

Authors:  Brendan P O'Connell; Ahmet Cakir; Jacob B Hunter; David O Francis; Jack H Noble; Robert F Labadie; Geraldine Zuniga; Benoit M Dawant; Alejandro Rivas; George B Wanna
Journal:  Otol Neurotol       Date:  2016-09       Impact factor: 2.311

7.  Intracellular calcium and cyclic nucleotide levels modulate neurite guidance by microtopographical substrate features.

Authors:  Shufeng Li; Bradley Tuft; Linjing Xu; Marc Polacco; Joseph C Clarke; C Allan Guymon; Marlan R Hansen
Journal:  J Biomed Mater Res A       Date:  2016-04-19       Impact factor: 4.396

Review 8.  On the Horizon: Cochlear Implant Technology.

Authors:  Joseph P Roche; Marlan R Hansen
Journal:  Otolaryngol Clin North Am       Date:  2015-10-09       Impact factor: 3.346

9.  Hearing Preservation Outcomes With a Mid-Scala Electrode in Cochlear Implantation.

Authors:  Jacob B Hunter; René H Gifford; George B Wanna; Robert F Labadie; Marc L Bennett; David S Haynes; Alejandro Rivas
Journal:  Otol Neurotol       Date:  2016-03       Impact factor: 2.311

Review 10.  The importance of electrode location in cochlear implantation.

Authors:  Brendan P O'Connell; Jacob B Hunter; George B Wanna
Journal:  Laryngoscope Investig Otolaryngol       Date:  2016-11-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.