Test-retest reliability of retinal oxygen saturation measurement.

Rachael A O'Connell1, Andrew J Anderson, Sarah L Hosking, Abrez H Batcha, Bang V Bui.   

Abstract

PURPOSE: To determine intrasession and intersession repeatability of retinal vessel oxygen saturation from the Oxymap Retinal Oximeter using a whole image-based analysis technique and so determine optimal analysis parameters to reduce variability.
METHODS: Ten fundus oximetry images were acquired through dilated pupils from 18 healthy participants (aged 22 to 38) using the Oxymap Retinal Oximeter T1. A further 10 images were obtained 1 to 2 weeks later from each individual. Analysis was undertaken for subsets of images to determine the number of images needed to return a stable coefficient of variation (CoV). Intrasession and intersession variability were quantified by evaluating the CoV and establishing the 95% limits of agreement using Bland and Altman analysis. Retinal oxygenation was derived from the distribution of oxygenation values from all vessels of a given width in an image or set of images, as described by Paul et al. in 2013.
RESULTS: Grouped in 10-μm-wide bins, oxygen saturation varied significantly for both arteries and veins (p < 0.01). Between 110 and 150 μm, arteries had the least variability between individuals, with average CoVs less than 5% whose confidence intervals did not overlap with the greater than 10% average CoVs for veins across the same range. Bland and Altman analysis showed that there was no bias within or between recording sessions and that the 95% limits of agreement were generally lower in arteries.
CONCLUSIONS: Retinal vessel oxygen saturation measurements show variability within and between clinical sessions when the whole image is used, which we believe more accurately reflects the true variability in Oxymap images than previous studies on select image segments. Averaging data from vessels 100 to 150 μm in width may help to minimize such variability.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24811846     DOI: 10.1097/OPX.0000000000000257

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  9 in total

1.  Retinal vessel oxygen saturation in healthy subjects and early branch retinal vein occlusion.

Authors:  Jing-Yan Yang; Bing You; Qian Wang; Szy Yann Chan; Jost B Jonas; Wen-Bin Wei
Journal:  Int J Ophthalmol       Date:  2017-02-18       Impact factor: 1.779

Review 2.  Imaging Retinal Activity in the Living Eye.

Authors:  Jennifer J Hunter; William H Merigan; Jesse B Schallek
Journal:  Annu Rev Vis Sci       Date:  2019-09-15       Impact factor: 6.422

3.  Retinal Oximetry in a Healthy Japanese Population.

Authors:  Yuki Nakano; Takeru Shimazaki; Nobuko Kobayashi; Yukiko Miyoshi; Aoi Ono; Mamoru Kobayashi; Chieko Shiragami; Kazuyuki Hirooka; Akitaka Tsujikawa
Journal:  PLoS One       Date:  2016-07-19       Impact factor: 3.240

4.  Early hyperoxia burden detected by cerebral near-infrared spectroscopy is superior to pulse oximetry for prediction of severe retinopathy of prematurity.

Authors:  Z A Vesoulis; C E Lust; S M Liao; S B Trivedi; A M Mathur
Journal:  J Perinatol       Date:  2016-08-25       Impact factor: 2.521

5.  Inter-visit variability of conjunctival microvascular hemodynamic measurements in healthy and diabetic retinopathy subjects.

Authors:  Maziyar M Khansari; Michael Tan; Preny Karamian; Mahnaz Shahidi
Journal:  Microvasc Res       Date:  2018-02-10       Impact factor: 3.514

6.  Retinal oxygen saturation in 1461 healthy children aged 7-19 and its associated factors.

Authors:  Xiaoxiao Liu; Xiangui He; Yao Yin; Bo Zhang; Sifei Sun; Jianfeng Zhu; Haidong Zou; Ling Wang; Xun Xu
Journal:  Acta Ophthalmol       Date:  2019-02-03       Impact factor: 3.761

7.  The Changes of Retinal Saturation after Long-Term Tamponade with Silicone Oil.

Authors:  Bingsheng Lou; Zhaohui Yuan; Liwen He; Lixia Lin; Qianying Gao; Xiaofeng Lin
Journal:  Biomed Res Int       Date:  2015-10-18       Impact factor: 3.411

8.  Normative Values of Retinal Oxygen Saturation in Rhesus Monkeys: The Beijing Intracranial and Intraocular Pressure (iCOP) Study.

Authors:  Jing Li; Yiquan Yang; Diya Yang; Xiangxiang Liu; Yunxiao Sun; Shifei Wei; Ningli Wang
Journal:  PLoS One       Date:  2016-03-01       Impact factor: 3.240

9.  Retinal Oximetry and Vessel Diameter Measurements With a Commercially Available Scanning Laser Ophthalmoscope in Diabetic Retinopathy.

Authors:  Norman P Blair; Justin Wanek; Anthony E Felder; Charlotte E Joslin; Jacob K Kresovich; Jennifer I Lim; Felix Y Chau; Yannek Leiderman; Mahnaz Shahidi
Journal:  Invest Ophthalmol Vis Sci       Date:  2017-10-01       Impact factor: 4.799

  9 in total