Diana Mortensen1, Katrine Dannemand2, Svante Twetman1, Mette Kirstine Keller1.
Abstract
OBJECTIVE: To evaluate the performance of an impedance spectroscopy technology for detecting non-cavitated occlusal caries lesions in permanent teeth in vitro. The method was compared with a commonly used laser fluorescence device and validated against histology. MATERIAL ANDEntities:
Keywords: Caries detection; impedance spectroscopy; laser fluorescence; molar teeth.
Year: 2014 PMID: 24799965 PMCID: PMC4009735 DOI: 10.2174/1874210601408010028
Source DB: PubMed Journal: Open Dent J ISSN: 1874-2106
Mean values (SD) of readings in relation to the Downer classification [Downer].
| Method | D1 (n=25) | D2 (n=33) | D3 (n=33) | D4 (n=8) |
|---|---|---|---|---|
| ACIS | 33.1 (20.3) | 43.9 (26.3) | 46.0* (23.9) | 52.8* (15.1) |
| LF pen | 13.8 (19.5) | 16.7 (23.2) | 16.5 (20.6) | 11.8 (10.8) |
Validity of AC impedance spectroscopy (ACIS, CarieScan PRO) and laser fluorescence (DIAGNOdent pen) in detecting occlusal caries on dentin (D2) level. Values in the table are based on the median of three independent examiners.
| Method | Cut-Off Reading | Sensitivity | Specificity | Accuracy | Area Under ROC |
|---|---|---|---|---|---|
| ACIS | 40 | 0.76 | 0.54 | 0.70 | 0.64 |
| ACIS | 50 | 0.34 | 0.96 | 0.49 | 0.63 |
| ACIS | 70 | 0.12 | 1.00 | 0.35 | 0.56 |
| ACIS | 90 | 0.08 | 1.00 | 0.32 | 0.54 |
| LF pen | 10 | 0.47 | 0.54 | 0.49 | 0.54 |
| LF pen | 20 | 0.23 | 0.89 | 0.40 | 0.56 |
| LF pen | 40 | 0.08 | 0.96 | 0.31 | 0.52 |
Validity of AC impedance spectroscopy (ACIS, CarieScan Pro) and laser fluorescence (DIAGNOdent pen) in detecting oc-clusal caries on dentin (D3) level. Values in the table are based on the median of three independent examiners.
| Method | Cut-Off Reading | Sensitivity | Specificity | Accuracy | Area under ROC |
|---|---|---|---|---|---|
| ACIS | 40 | 0.78 | 0.39 | 0.55 | 0.59 |
| ACIS | 50 | 0.39 | 0.81 | 0,64 | 0.60 |
| ACIS | 70 | 0.12 | 0.93 | 0.60 | 0.53 |
| ACIS | 90 | 0.05 | 0.98 | 0.60 | 0.49 |
| LF pen | 10 | 0.49 | 0.54 | 0.52 | 0.52 |
| LF pen | 20 | 0.24 | 0.83 | 0.59 | 0.54 |
| LF pen | 40 | 0.07 | 0.93 | 0.58 | 0.50 |
Intra- and inter-examiner reproducibility using laser fluorescence (LF pen) and AC impedance spectroscopy (ACIS). Val-ues in table denote the mean ICC value and 95% CI.
| LF Pen | ACIS | |
|---|---|---|
| Examiner 1 | 0.95 (0.93-0.97) | 0.79 (0.70-0.85) |
| Examiner 2 | 0.90 (0.86-0.94) | 0.64 (0.51-0.74) |
| Examiner 3 | 0.87 (0.81-0.91) | 0.47 (0.30-0.61) |
| Examiner 1 vs. 2 | 0.98 (0.97-0.99) | 0.83 (0.75-0.89) |
| Examiner 1 vs. 3 | 0.98 (0.97-0.99) | 0.81 (0.71-0.87) |
| Examiner 2 vs. 3 | 0.98 (0.97-0.99) | 0.78 (0.68-0.75) |