Rodolfo Ghirlando1, Huaying Zhao2, Andrea Balbo3, Grzegorz Piszczek4, Ute Curth5, Chad A Brautigam6, Peter Schuck7.
Abstract
Accurate measurements of rotor temperature are critical for the interpretation of hydrodynamic parameters in analytical ultracentrifugation. We have recently developed methods for a more accurate determination of the temperature of a spinning rotor using iButton temperature loggers. Here we report that the temperature measured with the iButton on the counterbalance of a resting rotor, following thermal equilibration under high vacuum, closely corresponded to the temperature of the spinning rotor with a precision better than 0.2°C. This strategy offers an inexpensive and straightforward approach to monitor the accuracy of the temperature calibration and determine corrective temperature offsets. Published by Elsevier Inc.Entities:
Keywords: Hydrodynamics; Sedimentation velocity; Temperature calibration
Mesh:
Substances:
Year: 2014 PMID: 24799348 PMCID: PMC4080836 DOI: 10.1016/j.ab.2014.04.029