The aim of this study was to compare the therapeutic effects of rhGH administered either by subcutaneous needle-injection (pens) or subcutaneous needle-free jet-injection (VISION(®)). Furthermore, a survey was carried out after using VISION(®) for 12 mo. A needle-free injection group consisting of 18 subjects (11 males and 7 females, mean age 5.87 ± 2.05 yr at the start of hGH therapy) who have not used pen injectors to date, were allowed to use VISION(®) in their third to fifth years of GH therapy. In addition, a group of 8 subjects who had been using pen injectors at our clinic (6 males and 2 females, mean age 6.54 ± 2.78 at the start of GH therapy) was monitored as a control. The results indicate that there are no significant differences between the mean growth rates, growth rate SD scores or height SD scores when comparing injection devices. Furthermore, the survey of VISION(®) revealed that 70% of the subjects found it slightly or not painful at or after injection, 70% found VISION(®) very easy or easy to use, and 80% found the weight of the device appropriate. All subjects expressed a desire to continue using VISION(®) in the future. Our results suggest that there are no problems with the effectiveness of hGH treatment with VISION(®), a needle-free jet-injection device and that VISION(®) is an effective device for children who have an aversion to needle injection.
The aim of this study was to compare the therapeutic effects of rhGH administered either by subcutaneous needle-injection (pens) or subcutaneous needle-free jet-injection (VISION(®)). Furthermore, a survey was carried out after using VISION(®) for 12 mo. A needle-free injection group consisting of 18 subjects (11 males and 7 females, mean age 5.87 ± 2.05 yr at the start of hGH therapy) who have not used pen injectors to date, were allowed to use VISION(®) in their third to fifth years of GH therapy. In addition, a group of 8 subjects who had been using pen injectors at our clinic (6 males and 2 females, mean age 6.54 ± 2.78 at the start of GH therapy) was monitored as a control. The results indicate that there are no significant differences between the mean growth rates, growth rate SD scores or height SD scores when comparing injection devices. Furthermore, the survey of VISION(®) revealed that 70% of the subjects found it slightly or not painful at or after injection, 70% found VISION(®) very easy or easy to use, and 80% found the weight of the device appropriate. All subjects expressed a desire to continue using VISION(®) in the future. Our results suggest that there are no problems with the effectiveness of hGH treatment with VISION(®), a needle-free jet-injection device and that VISION(®) is an effective device for children who have an aversion to needle injection.
Syringe materials have changed from glass to plastic, while the shape of the syringe itself
has not changed much. The biggest change, has been in the size of the needle. In particular,
needles have become thinner, reaching the present size of 32 gauge.On the other hand, new syringe design has been sought. In order to eliminate the fear, pain
and infection of needle injection, needle-free injection systems have been designed and put
into clinical use.Needle-free injection devices used in Japan include the Twin-Jector EZ and EZ II for humangrowth hormone use, SYRIJET for dental surgery/surgical management, and ShimaJET, recently
developed and commercialized by a Japanese company for insulin use.Several studies have compared needle-free injectors to conventional syringes with needles.
So far, the results have been conflicting: the needle-free injector has been reported to be
both less painful and more readily accepted by the patients and their families than
conventional needle injection (1, 2), while other reports found no differences between
needle-free injection and conventional syringes (3).In 1993 we investigated the use of a needle-free injection system for GH therapy and
reported that it was beneficial in improving patients’ adherence to treatment regimes and
that no differences were found in the therapeutic effects, compared with needle injection
(4). In our clinic, we carried out GH therapy using
a needle-free injector for children who feared needle injection. However, due to a product
recall in November 2000, an alternative needle-free injection system, VISION®,
was imported directly from ANTARES PHARMA Co., U.S.A. and used for GH therapy.We have collected data from 18 subjects using the VISION® needle-free injector
and 8 subjects using conventional needle pen-injectors, and compared the data. Furthermore,
a survey of the VISION® users was carried out. The results are reported
below.
Subjects and Methods
The needle-free injection (VISION®) manufactured by Antares Pharma Co. is shown
in Fig. 1. The needle-free syringe was exchanged with a new one, once a week.
Fig. 1
Needle-free injection device, VISION®.
Needle-free injection device, VISION®.A needle-free injection group consisting of 18 subjects (11 males and 7 females, mean age
5.87 ± 2.05 yr at the start of hGH therapy) who have not used needle injectors to date, were
allowed to use VISION® in their third to fifth years of GH therapy. In addition,
a group of 8 subjects who had been using needle injectors at our clinic (6 males and 2
females, mean age 6.54 ± 2.78 at the start of GH therapy) was monitored as a control (Table 1). All subjects were pediatric patients with GH deficiency (GHD).
Table 1
Patient background of needle-free injection group and needle injection
group
Growth rates, growth rate SD scores and height SD scores from the needle-free injection
group were compared with the control group and evaluated. Numerical values are shown as mean
± SD. Using the t-test, probabilities of not more than 5% were determined as significant
difference.A survey was carried out after using VISION® for 12 mo (Table 2), in which the operation and effectiveness of the device were evaluated.
Table 2
Survey contents
Written consent to participate in this survey was obtained from the parents of the
patients.
Results
For each parameter, as shown in Table 1, no
differences were observed in the backgrounds of the needle and needle-free groups of
patients.Figure 2 shows growth rates. The growth rates for both groups over 8 yr from the start of GH
treatment were compared, with no significant differences observed between the needle and the
needle-free groups. Figures 3 and 4 show growth rate SD scores and height SD scores, respectively. No differences were
observed between the two groups in either the growth rate SD scores or the height SD
scores.
Fig. 2
Mean (SD) growth rates during duration of treatment by needle-free injection or
traditional needle injection.
Fig. 3
Mean (SD) growth rate SD scores during duration of treatment by needle-free
injection or traditional needle injection.
Fig. 4
Mean (SD) height SD scores during duration of treatment by needle-free injection or
traditional needle injection.
Mean (SD) growth rates during duration of treatment by needle-free injection or
traditional needle injection.Mean (SD) growth rate SD scores during duration of treatment by needle-free
injection or traditional needle injection.Mean (SD) height SD scores during duration of treatment by needle-free injection or
traditional needle injection.The survey was given to the 18 subjects who had experienced needle-free injection using
Twin-Jector EZ, Twin-Jector EZ II and/or VISION®. Answers were obtained from 17
subjects (6 girls and 11 boys). The response rate was 94.4%. Figures 5, 6, 7, 8,9 show pain at injection, pain after injection, how to use, size/weight and desire to
continue use, respectively. At injection, 70% felt slight or no pain. After injection,
approximately 70% felt no pain. In comparison with the previous needle-free injectors,
Twin-Jector EZ and/or Twin-Jector EZ II, about 70% responded that the directions for use of
VISION were easy or very easy to understand. They also found that VISION® itself
was easy to use. The subjects found the VISION injector to be appropriately sized and easy
to handle. In addition, 80% or more answered that VISION® was not heavy. All
subjects expressed a desire to continue using VISION in the future.
Fig. 5
Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.
Fig. 6
Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.
Fig. 7
Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.
Fig. 8
Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.
Fig. 9
Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.
Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.Responses to the descriptive questionnaire after using the needle-free injection
(VISION®) for 12 mo.Regarding breakage, in 4.5 yr of use by 18 subjects, a total of 3 syringes were broken by 2
subjects at our clinic. All incidents occurred at the initial stage of using
VISION® and could have been caused by inappropriate operation.
Discussion
GH deficiency was treated by intramuscular injection of GH extracted from human pituitary
glands in the early 1960s. Since then, there have been many advances in treatment
encompassing administration, the injection product and the injection device.In our clinic, we use a pen needle device for the treatment of GH. But the needle-free
alternative is an effective solution for children who have an aversion to needle injection,
and it is useful for increasing the dose of GH treatment.In the United States, VISION® has been approved for subcutaneous
self-administration of insulin by insulin-dependent diabetes mellituspatients. In Europe,
VISION® (ZomaJect 2) has been used for subcutaneous self-injection of growth
hormone.Several studies have been carried out comparing pharmacokinetic and pharmacodynamic
characteristics of needle-free systems with traditional needle injection (5,6,7,8,9). These studies support the use of jet injection as a
viable alternative to the traditional injection pens with needles.This study was designed to compare the therapeutic effects of rhGH administered by
subcutaneous needle-injection (pens) with subcutaneous needle-free jet-injection
(VISION®). The results showed that no significant differences were found in the
growth rates, growth rate SD scores or height SD scores, when comparing injection devices.
Furthermore, the survey of VISION® revealed that 70% found it slightly or not
painful at or after injection, 70% found VISION® very easy or easy to use, and
80% found the weight of the device appropriate.As for complaints on using VISION®, only 3 syringes were damaged by 2 subjects
in 4.5 yr of use of VISION®. All breakages occurred at the initial stage of using
VISION® and could have been caused by inappropriate handling. Subsequently,
there was no breakage observed up to the end of study. There were no other complaints.
Therefore, it was concluded that there are no problems regarding safety and operationality
for hGH treatment with VISION®, a needle-free injector.The results indicate that there are no problems regarding effectiveness of the
VISION® needle-free injector for hGH treatment. The results also indicate that
VISION® is an effective device for children who have an aversion to needle
injection.
Authors: H G Dörr; S Zabransky; E Keller; B J Otten; C J Partsch; L Nyman; B K Gillespie; N R Lester; A M Wilson; Cecilia Hyrén; M A van Kuijck; P Schuld; S L Schoenfeld Journal: J Pediatr Endocrinol Metab Date: 2003-03 Impact factor: 1.634