| Literature DB >> 24785441 |
Marek Belohlavek1, Minako Katayama2, David Zarbatany3, F David Fortuin2, Mostafa Fatemi4, Ivan Z Nenadic4, Eileen M McMahon2.
Abstract
Catheters are increasingly used therapeutically and investigatively. With complex usage comes a need for more accurate intracardiac localization than traditional guidance can provide. An injection catheter navigated by ultrasound was designed and then tested in an open-chest model of acute ischemia in eight pigs. The catheter is made "acoustically active" by a piezo-electric crystal near its tip, electronically controlled, vibrating in the audio frequency range and uniquely identifiable using pulsed-wave Doppler. Another "target" crystal was sutured to the epicardium within the ischemic region. Sonomicrometry was used to measure distances between the two crystals and then compared with measurements from 2-D echocardiographic images. Complete data were obtained from seven pigs, and the correlation between sonomicrometry and ultrasound measurements was excellent (p < 0.0001, ρ = 0.9820), as was the intraclass correlation coefficient (0.96) between two observers. These initial experimental results suggest high accuracy of ultrasound navigation of the acoustically active catheter prototype located inside the beating left ventricle.Entities:
Keywords: Acoustic catheter; Pulsed-wave Doppler; Ultrasound navigation
Mesh:
Year: 2014 PMID: 24785441 PMCID: PMC4051318 DOI: 10.1016/j.ultrasmedbio.2014.02.032
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998