Membrane oxygenator heat exchanger failure detected by unique blood gas findings.

Justin L Hawkins.   

Abstract

Failure of components integrated into the cardiopulmonary bypass circuit, although rare, can bring about catastrophic results. One of these components is the heat exchanger of the membrane oxygenator. In this compartment, unsterile water from the heater cooler device is separated from the sterile blood by stainless steel, aluminum, or by polyurethane. These areas are glued or welded to keep the two compartments separate, maintaining sterility of the blood. Although quality control testing is performed by the manufacturer at the factory level, transport presents the real possibility for damage. Because of this, each manufacturer has included in the instructions for use a testing procedure for testing the integrity of the heat exchanger component. Water is circulated through the heat exchanger before priming and a visible check is made of the oxygenator bundle to check for leaks. If none are apparent, then priming of the oxygenator is performed. In this particular case, this procedure was not useful in detecting communication between the water and blood chambers of the oxygenator.

Entities:  

Mesh:

Year:  2014        PMID: 24779125      PMCID: PMC4557518     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  4 in total

1.  The incidence and cause of emergency oxygenator changeovers.

Authors:  A R Fisher
Journal:  Perfusion       Date:  1999-05       Impact factor: 1.972

2.  Simple modified ultrafiltration.

Authors:  G J Myers; R B Leadon; L B Mitchell; D B Ross
Journal:  Perfusion       Date:  2000-09       Impact factor: 1.972

3.  Testing of heat exchangers in membrane oxygenators using air pressure.

Authors:  Carole Hamilton; Jutta Stein; Rainer Seidler; Robert Kind; Karin Beck; Jürgen Tosok; Jörg Upterfofel
Journal:  Perfusion       Date:  2006-03       Impact factor: 1.972

4.  Study on causes and prevention of electrostatic charge build-up during extracorporeal circulation.

Authors:  J Snijders; P de Bruijn; M Bergmans; G Bastianen
Journal:  Perfusion       Date:  1999-09       Impact factor: 1.972

  4 in total
  2 in total

1.  It is Time to Update Membrane Oxygenator Testing Standards and Their Instructions for Use.

Authors:  Gregory S Matte
Journal:  J Extra Corpor Technol       Date:  2016-09

Review 2.  Sedative and Analgesic Pharmacokinetics During Pediatric ECMO.

Authors:  Kanecia O Zimmerman; Samantha H Dallefeld; Christoph P Hornik; Kevin M Watt
Journal:  J Pediatr Pharmacol Ther       Date:  2020-11-13
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.