Literature DB >> 10411251

The incidence and cause of emergency oxygenator changeovers.

A R Fisher1.   

Abstract

Emergency oxygenator changeover presents as an important potential hazard during cardiopulmonary bypass. In order to assess the frequency and possible causes of this event a retrospective survey was conducted of cardiothoracic units in the UK and Ireland. The survey was sent to these units in 1993 to cover the period from 1990 to 1992 inclusive and again in 1997 to cover the period from 1994 to 1996 inclusive. The total number of procedures reported in 1993 was 68937 (response of 63%), among which there were 17 emergency oxygenator changeovers representing an incidence of one for every 4055 cases. The use of aprotinin was associated with nine of these incidents. Two other incidents occurred due to leaking oxygenators. The total number of procedures reported in 1997 was 97313 (response of 70%), among which there were 21 emergency oxygenator changeovers, representing an incidence of one for every 4634 cases. The use of aprotinin was only associated with two of these incidents while seven were due to the development of a high trans-oxygenator pressure gradient and six occurred due to leaking oxygenators.

Mesh:

Year:  1999        PMID: 10411251     DOI: 10.1177/026765919901400309

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  4 in total

1.  Membrane oxygenator heat exchanger failure detected by unique blood gas findings.

Authors:  Justin L Hawkins
Journal:  J Extra Corpor Technol       Date:  2014-03

2.  Isoflurane cracks the polycarbonate connector of extra-corporeal circuit -A case report-.

Authors:  Hyung Sun Lim; Sung Hun Cho; Deok Kyu Kim; Sang Kyi Lee; Seong Hoon Ko
Journal:  Korean J Anesthesiol       Date:  2010-03-29

3.  Oxygenation failure during cardiopulmonary bypass prompts new safety algorithm and training initiative.

Authors:  David P Webb; Robert J Deegan; James P Greelish; John G Byrne
Journal:  J Extra Corpor Technol       Date:  2007-09

4.  Anesthetic vaporizer mount malfunction resulting in oxygenation failure after initiating cardiopulmonary bypass: specific recommendations for the pre-bypass checklist.

Authors:  Nischal K Gautam; Michael L Schmitz; Luis M Zabala; Michael W White; Wesley A Mckamie; Alyssa Lutz; Charles E Johnson
Journal:  J Extra Corpor Technol       Date:  2009-09
  4 in total