Literature DB >> 24772597

Non-invasive estimated continuous cardiac output (escCO) during severe sepsis and septic shock resuscitation.

Chairat Permpikul, Thongchai Leelayuthachai.   

Abstract

BACKGROUND: Cardiac output (CO) is an important hemodynamic parameter during sepsis and septic shock resuscitation. Conventionally, this value is obtained at bedside by the thermodilution technique, which requires a pulmonary artery catheter (PAC). Estimated Continuous Cardiac Output (esCCO, Nihon Kohden, Japan) calculated from pulse-wave transit time (PWTT) was examined here as an alternative. MATERIAL AND
METHOD: A prospective study was performed in a 14-bed ICU facility. Patients with severe sepsis and septic shock who had PAC placed were included. Serial thermodilution cardiac outputs (CO(IBT)) and esCCO (COesCCO) were obtained at the beginning of resuscitation (t0), at 48 hours (t48), and at 72 hours (t72). Other parameters from the tested device; namely, estimated continuous cardiac index (esCCI), estimated stroke volume (esSV), and estimated stroke volume index (esSVI), were also achieved.
RESULTS: A total of 90-paired readings from ten ICU patients were collected. The overall correlation coefficient (R) between COesCCO and CO(IBT) was 0.76. When focusing on the correlation from each time point, we found R at t0 = 0.65, t48 = 0.74, and t72 = 0.84 (all p < 0.001). Bland and Altman analysis corrected for repeated measures showed a bias of 1.2 liter/min and limits of agreement from -2.8 to +5.2 liter/min. Results also showed fair to poor correlation with other parameters that derived from this device (esCCI, esSV esSVI).
CONCLUSION: The estimated continuous cardiac output (esCCO) correlated well with the cardiac output obtained by thermodilution techniques, especially when patients were out of shock.

Entities:  

Mesh:

Year:  2014        PMID: 24772597

Source DB:  PubMed          Journal:  J Med Assoc Thai        ISSN: 0125-2208


  1 in total

1.  Estimated continuous cardiac output based on pulse wave transit time in off-pump coronary artery bypass grafting: a comparison with transpulmonary thermodilution.

Authors:  Alexey A Smetkin; Ayyaz Hussain; Evgenia V Fot; Viktor I Zakharov; Natalia N Izotova; Angelika S Yudina; Zinaida A Dityateva; Yanina V Gromova; Vsevolod V Kuzkov; Lars J Bjertnæs; Mikhail Y Kirov
Journal:  J Clin Monit Comput       Date:  2016-03-07       Impact factor: 2.502

  1 in total