Autorefraction versus subjective refraction in a radially asymmetric multifocal intraocular lens.

Jan Willem van der Linden1, Violette Vrijman, Rana Al-Saady, Rana El-Saady, Ivanka J van der Meulen, Maarten P Mourits, Ruth Lapid-Gortzak.   

Abstract

PURPOSE: To evaluate whether the automated refraction (AR) correlates with subjective manifest (MR) refraction in eyes implanted with radially asymmetric multifocal intraocular lens (IOLs).
METHODS: This retrospective study evaluated 52 eyes (52 patients) implanted with a radially asymmetric multifocal IOL (LS-312 MF30, Oculentis, Germany). At 3 months postoperatively, the AR and MR values were compared to determine the correlation between the sphere (S), the spherical equivalent (SE) and the astigmatic components J0 and J45.
RESULTS: The difference of mean spherical measurement was +0.98D ± 0.62, with the AR measuring more myopic. The difference of the mean spherical equivalent was +1.11D ± 0.57, again with AR being more myopic. Both these differences were statistically significant (p < 0.001). The astigmatic components showed less differences, with the mean difference of the J0 being -0.09D ± 0.43, and the J45 of +0.04D ± 0.47, which were both not statistically significant (p = 0.123 and p = 0.531, respectively). Correlation analysis of the refractive parameters showed r(2) = 0.067, r(2) = 0.078, r(2) = 0.018 and r(2) = 0.015, respectively, all of which point to a low correlation between the AR and the MR.
CONCLUSION: Autorefraction shows poor correlation to manifest subjective refraction with these radially asymmetric multifocal IOLs. The autorefraction systematically underestimates the spherical and spherical equivalent power, while the correlation between the astigmatic components was also low. Autorefraction seems not a valid starting point for manifest subjective refraction with these types of lenses, unless a corrective factor of about +1 dioptre is used.
© 2014 Acta Ophthalmologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  LS-312; M Plus; autorefraction; cataract; multifocal intraocular lens; radially asymmetric multifocal intraocular lens; refractive lens exchange; residual refraction

Mesh:

Year:  2014        PMID: 24767468     DOI: 10.1111/aos.12410

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  7 in total

1.  Error induced by the estimation of the corneal power and the effective lens position with a rotationally asymmetric refractive multifocal intraocular lens.

Authors:  David P Piñero; Vicente J Camps; María L Ramón; Verónica Mateo; Rafael J Pérez-Cambrodí
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

2.  Validity of automated refraction after segmented refractive multifocal intraocular lens implantation.

Authors:  César Albarrán-Diego; Gonzalo Muñoz; Stephanie Rohrweck; Santiago García-Lázaro; José Ricardo Albero
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

3.  Comparative analysis of visual quality between unilateral implantation of a trifocal intraocular lens and a rotationally asymmetric refractive multifocal intraocular lens.

Authors:  Na Hui; Mei-Fang Chu; Yan Li; Cong-Yi Wang; Lei Yu; Bo Ma
Journal:  Int J Ophthalmol       Date:  2022-09-18       Impact factor: 1.645

4.  A comparative study on early vision quality after implantation of refractive segmental and diffractive multifocal intraocular lens.

Authors:  Huifang Lian; Weisong Ma; Qiuhong Wei; Xiaoyong Yuan
Journal:  Pak J Med Sci       Date:  2020 Nov-Dec       Impact factor: 1.088

5.  Factors affecting prediction error after cataract surgery with implantation of various multifocal IOLs in patients with previous refractive laser surgery.

Authors:  Young-Sik Yoo; Min Chae Kang; Jongyeop Park; Hyung-Goo Kwon; Eui-Sang Chung; Dong Hui Lim; Tae-Young Chung
Journal:  Ann Transl Med       Date:  2021-12

6.  Comparison of Visual Outcomes and Quality of Life in Patients with High Myopic Cataract after Implantation of AT LISA Tri 839MP and LS-313 MF30 Intraocular Lenses.

Authors:  Jiying Shen; Limei Zhang; Shuang Ni; Lei Cai; Haike Guo; Jin Yang
Journal:  J Ophthalmol       Date:  2022-02-28       Impact factor: 1.909

7.  Subjective and objective refractions in eyes with extended-depth-of-focus intraocular lenses using echelette optics: clinical and experimental study.

Authors:  Yuka Ota; Keiichiro Minami; Shinichi Oki; Hiroko Bissen-Miyajima; Keiichiro Okamoto; Masaya Nakashima; Kazuo Tsubota
Journal:  Acta Ophthalmol       Date:  2020-11-16       Impact factor: 3.761

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.