Simple, inexpensive, and effective injector for descemet membrane endothelial keratoplasty.

Eun Chul Kim1, Gustavo Bonfadini, Leisha Todd, Angela Zhu, Albert S Jun.   

Abstract

PURPOSE: We describe an inexpensive, simple, and effective endothelium-Descemet membrane (EDM) graft injector assembled from regular operating room supplies in Descemet membrane endothelial keratoplasty (DMEK).
METHODS: To assemble the injector, standard intravenous tubing was cut approximately 2 inches from the Luer lock end, leaving a steep bevel. The cut end of the tubing was firmly wedged bevel up and advanced into the back of an Alcon IOL B cartridge. The Luer lock end of the tubing was then attached to a 5- or 10-mL syringe filled with BSS Plus. The EDM graft was then placed into a Petri dish filled with BSS. After the graft was sucked into an injector with bevel-side up under the surgical microscope, the graft was then inserted into the anterior chamber with the injector through the main incision in the superotemporal quadrant.
RESULTS: In seven eyes of seven patients with Fuchs endothelial corneal dystrophy treated with DMEK using our injector, clear attached grafts and improved visual acuity were achieved.
CONCLUSIONS: This simple, inexpensive, and effective injector is a safe and viable device to facilitate this part of DMEK surgery.

Entities:  

Mesh:

Year:  2014        PMID: 24763123     DOI: 10.1097/ICO.0000000000000121

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  9 in total

1.  Commentary: Modified intraocular lens injector assisted rescue technique for failed viscoexpression in a case of intracameral cysticercosis.

Authors:  Ashok Sharma
Journal:  Indian J Ophthalmol       Date:  2020-05       Impact factor: 1.848

2.  Modified intraocular lens injector assisted rescue technique for failed viscoexpression in a case of intracameral cysticercosis.

Authors:  Rouli Sud; Pallavi Sharma; Brijesh Takkar; Sumeet Khanduja
Journal:  Indian J Ophthalmol       Date:  2020-05       Impact factor: 1.848

3.  Safety analysis and results of a borosilicate glass cartridge for no-touch graft loading and injection in Descemet membrane endothelial keratoplasty.

Authors:  Annekatrin Rickmann; Silke Wahl; Alisa Katsen-Globa; Peter Szurman
Journal:  Int Ophthalmol       Date:  2019-01-04       Impact factor: 2.031

4.  "Endothelium-Out" and "Endothelium-In" Descemet Membrane Endothelial Keratoplasty (DMEK) Graft Insertion Techniques: A Systematic Review With Meta-Analysis.

Authors:  Hon Shing Ong; Hla M Htoon; Marcus Ang; Jodhbir S Mehta
Journal:  Front Med (Lausanne)       Date:  2022-06-14

5.  Aspiration of Tri-folded, Endothelium-In Grafts for Descemet Membrane Endothelial Keratoplasty.

Authors:  Sezen Karakus; Mehrnaz Ighani; Puntakarn Noparat; Marjan Tofigh; Eric Chiang; Kali Barnes; Conan Y Chen; Tiffany S Liu; Allen O Eghrari
Journal:  Cornea       Date:  2019-05       Impact factor: 2.651

6.  Clinical Outcomes Of Descemet Membrane Endothelial Keratoplasty Using The Bonfadini-Todd Injector For Graft Insertion.

Authors:  Mehrnaz Ighani; Sezen Karakus; Allen O Eghrari
Journal:  Clin Ophthalmol       Date:  2019-09-20

Review 7.  Evolution of therapies for the corneal endothelium: past, present and future approaches.

Authors:  Hon Shing Ong; Marcus Ang; Jodhbir Mehta
Journal:  Br J Ophthalmol       Date:  2020-07-24       Impact factor: 4.638

8.  Comparing the effect of three Descemet membrane endothelial keratoplasty injectors on endothelial damage of grafts.

Authors:  Elizabeth Shen; Adam Fox; Brian Johnson; Marjan Farid
Journal:  Indian J Ophthalmol       Date:  2020-06       Impact factor: 1.848

9.  Two-year clinical outcome after Descemet membrane endothelial keratoplasty using a standardized protocol.

Authors:  K S Siddharthan; Varna Shet; Anushri Agrawal; Jagdeesh Kumar Reddy
Journal:  Indian J Ophthalmol       Date:  2020-11       Impact factor: 1.848

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.