Patrick Z McVeigh1, Raphael Sacho, Robert A Weersink, Vitor M Pereira, Walter Kucharczyk, Eric J Seibel, Brian C Wilson, Timo Krings. 1. *Department of Medical Biophysics, University of Toronto; ‡Department of Medical Imaging, Toronto Western Hospital, University Health Network; §Radiation Medicine Program, Princess Margaret Cancer Centre; ¶Techna Institute, University Health Network, Toronto, Ontario, Canada; ‖Department of Medical Imaging, Toronto General Hospital, University Health Network, Toronto, Ontario, Canada; #Department of Mechanical Engineering, University of Washington, Seattle, Washington; **Department of Surgery, Division of Neurosurgery, University of Toronto, Toronto, Ontario, Canada.
Abstract
BACKGROUND: Endoluminal optical imaging, or angioscopy, has not seen widespread application during neurointerventional procedures, largely as a result of the poor imaging resolution of existing angioscopes. Scanning fiber endoscopes (SFEs) are a novel endoscopic platform that allows high-resolution video imaging in an ultraminiature form factor that is compatible with currently used distal access endoluminal catheters. OBJECTIVE: To test the feasibility and potential utility of high-resolution angioscopy with an SFE during common endovascular neurosurgical procedures. METHODS: A 3.7-French SFE was used in a porcine model system to image endothelial disruption, ischemic stroke and mechanical thrombectomy, aneurysm coiling, and flow-diverting stent placement. RESULTS: High-resolution, video-rate imaging was shown to be possible during all of the common procedures tested and provided information that was complementary to standard fluoroscopic imaging. SFE angioscopy was able to assess novel factors such as aneurysm base coverage fraction and side branch patency, which have previously not been possible to determine with conventional angiography. CONCLUSION: Endovascular imaging with an SFE provides important information on factors that cannot be assessed fluoroscopically and is a novel platform on which future neurointerventional techniques may be based because it allows for periprocedural inspection of the integrity of the vascular system and the deployed devices. In addition, it may be of diagnostic use for inspecting the vascular wall and postprocedure device evaluation.
BACKGROUND: Endoluminal optical imaging, or angioscopy, has not seen widespread application during neurointerventional procedures, largely as a result of the poor imaging resolution of existing angioscopes. Scanning fiber endoscopes (SFEs) are a novel endoscopic platform that allows high-resolution video imaging in an ultraminiature form factor that is compatible with currently used distal access endoluminal catheters. OBJECTIVE: To test the feasibility and potential utility of high-resolution angioscopy with an SFE during common endovascular neurosurgical procedures. METHODS: A 3.7-French SFE was used in a porcine model system to image endothelial disruption, ischemic stroke and mechanical thrombectomy, aneurysm coiling, and flow-diverting stent placement. RESULTS: High-resolution, video-rate imaging was shown to be possible during all of the common procedures tested and provided information that was complementary to standard fluoroscopic imaging. SFE angioscopy was able to assess novel factors such as aneurysm base coverage fraction and side branch patency, which have previously not been possible to determine with conventional angiography. CONCLUSION: Endovascular imaging with an SFE provides important information on factors that cannot be assessed fluoroscopically and is a novel platform on which future neurointerventional techniques may be based because it allows for periprocedural inspection of the integrity of the vascular system and the deployed devices. In addition, it may be of diagnostic use for inspecting the vascular wall and postprocedure device evaluation.
Endovascular neurosurgical therapies and technology continue to evolve and have now been developed for a number of intracranial pathologies. However, the lesions or malformations being treated are only visualized indirectly; treatment guidance and assessment are provided by x-ray fluoroscopy, which typically provides an indirect outline or contrast-filled “cast” of the vessel in the form of a 2-dimensional projection. Because fluoroscopy provides minimal information about the endoluminal surface, it cannot detect changes such as atherosclerosis until they cause structural deformation/luminal encroachment, nor can it assess posttherapeutic changes such as stent endothelialization. These limitations of fluoroscopy during “live” imaging can lead to difficulties when attempting to cross complex and tortuous 3-dimensional anatomy, with a concomitant increase in risk and procedure time as well as radiation dose.To address these issues, direct endoluminal optical visualization or angioscopy has been proposed as an adjuvant imaging modality since the 1980s1 and has been used to visualize the human aorta,2,3 coronary4 and peripheral5 arteries, as well as therapies such as carotid6 and coronary7 angioplasty. Angioscopy has also been used to observe the placement of intravascular occlusion coils and stents in the carotid arteries of porcine and canine animal models,8 as well as to assess the completeness of clip occlusion in a canineaneurysm model.9 Although promising, these early works have all made use of coherent fiber bundle (CFB) endoscopes in which a densely packed array of small-diameter glass fibers relays images from the vessel lumen to an eyepiece or camera mounted on the distal end. CFBs are inherently limited by the restriction that each pixel in an image requires a separate fiber in the bundle, necessitating dense packing for high-resolution imaging; however, additional fibers increase the bundle size and decrease its mechanical flexibility, limiting their utility in the cerebral vasculature. As a result, current angioscopes consist of low-resolution arrays to maintain a small profile and maintain some mechanical flexibility, and thus widespread adoption of angioscopy has been limited to date largely as a result of the mechanical fragility and poor image quality produced by these CFBs.Scanning fiber endoscopes (SFEs) are an emerging technology that produces images at significantly higher resolution than conventional fiberoptic endoscopes and in a miniature form factor, which is far more robust than a CFB.10 To date, SFEs have been used to image the esophagus11 and pancreatic and bile ducts,12 as well as the respiratory tree,13 but there have been no reports of their use for endovascular imaging. In this work, we demonstrate for the first time the feasibility of using the SFE for endoluminal imaging in a porcine model and demonstrate its utility during typical neurointerventional procedures.
METHODS
Scanning Fiber Endoscope
The details of the SFE design have been previously described elsewhere.10 In brief, the SFE consists of a single optical fiber that is scanned by a piezoelectric drive mechanism to illuminate the entire field of view of the endoscope, and return light is collected by a ring of fibers surrounding the scan head. For standard RGB (red, green, blue) imaging, the illumination spot is scanned in an outwardly expanding spiral, and the reflected light is collected and digitized to construct the image point-by-point in a manner analogous to the operation of a laser scanning microscope. The scan engine is contained within a stainless steel tube, which provides radiopacity, and the entire assembly is packaged into 3.7-French (F) (1.22 mm) low-density polyethylene tubing to provide flexibility while maintaining some pushability. The SFE used in this study is shown in Figure 1 and produces 600-line images at a video rate that is displayed on a monitor adjacent to the fluoroscopy display in our integrated interventional setup. Because the laser illumination light is conveyed through the central scanning fiber, the light source can be considered “cold” and does not actively conduct heat into the surrounding vessel. However, the distal tip of the SFE does contain a heating element that maintains the scan engine at physiological temperatures to avoid lens fogging when used in vivo, eliminating the need for heated saline or gas insufflation systems.8
FIGURE 1
A, a 3.7-French (F) scanning fiber endoscope (SFE) used in this study exiting from a 5-F Envoy guide catheter, demonstrating the 8-mm radiopaque tip. Scale bar = 5 mm. B, the SFE deployed in combination with a 9-F coaxial balloon guide catheter and 2.5-F microcatheter for imaging of mechanical thrombectomy with a stent retriever.
A, a 3.7-French (F) scanning fiber endoscope (SFE) used in this study exiting from a 5-F Envoy guide catheter, demonstrating the 8-mm radiopaque tip. Scale bar = 5 mm. B, the SFE deployed in combination with a 9-F coaxial balloon guide catheter and 2.5-F microcatheter for imaging of mechanical thrombectomy with a stent retriever.
In Vivo Imaging Using a Porcine Model
All animal procedures were carried out with institutional approval (University Health Network, Toronto, Ontario, Canada). Three Yorkshire pigs weighing 35 to 45 kg were sedated using ketamine/atropine/midazolam, and after induction and intubation, general anesthesia was maintained using inhaled isoflurane. An indwelling catheter was placed in an ear vein for fluid maintenance, along with electrocardiography, end-tidal CO2, and SpO2 leads for physiological monitoring. Bilateral femoral artery access was obtained using ultrasound guidance, and the renal, common carotid, and cervical arteries (3-5 mm) were selectively catheterized and assessed using standard fluoroscopic angiography. The catheters were continuously flushed using heparinized physiological saline solution (10 U/mL). Autologous thrombi were prepared by allowing unheparinized blood to clot over 24 hours and then immersing the resultant clot in fluoroscopic contrast media for 48 hours. Thrombi were then cut to size and injected into either a porcine lingual artery or an internal carotid artery as previously described.14 Immediately on completion of the study, the animals were euthanized under anesthesia by pentobarbital sodiumoverdose.
Endovascular Imaging Arrangement
To minimize any potential risks to the vessel endothelium during testing, the SFE was always used within a 056/070-inch guide catheter (Envoy; Codman, Raynham, Massachusetts/Neuron; Penumbra, Alameda, California). A clear field of view was provided by continuous heparinized saline solution flush through the outer guide catheter from standard pressure bags. The flush flow rate was manually adjusted to provide clear imaging depending on the diameter of the vessel being imaged and local blood flow conditions. Where necessary, proximal occlusion was provided by a coaxial balloon guide catheter (8-/9-F, Merci; Concentric Medical, Mountain View, California) through which the SFE could be placed within its own guide catheter.Detachable aneurysm coils of various designs were delivered through a microcatheter that was advanced via a second 6-F guide catheter from the contralateral access sheath to allow visualization of coil deployment and detachment in real time. The internal carotid artery immediately proximal to the rete mirabile was embolized using coils to create an area of flow stagnation to test flow-diverter placement. Then, in a manner similar to coil deployment, flow-diverting stents (Pipeline; Covidien, Saint-Laurent, Quebec, Canada) were deployed in the usual fashion through a microcatheter under direct visualization by the SFE in various anatomies including parent vessel bends, straight vessel segments, distally occluded side branches, and overlying patent side branches. Mechanical thrombectomy of injected autologous thrombi was performed using various stent retrievers (Solitaire FR; Covidien, Saint-Laurent/Trevo Provue; Stryker, Hamilton, Ontario, Canada) after lesion crossing with a microwire and microcatheter under SFE observation. Endothelial dissections were created by using a 0.014-inch microwire in reverse with support from the guide catheter to force the rigid tip of the wire under the endothelial surface and advanced to create a dissection flap.
RESULTS
As shown in Figure 1, the SFE was easily integrated with standard guide catheters including flexible distal access catheters used in neurointerventional procedures. The radiopaque distal end (scan engine) of the SFE was clearly visualized during fluoroscopy, and the optical image could be observed to ensure that the SFE did not protrude beyond the orifice of the carrying catheter without the need for x-ray validation. Clear images of portions of the endothelium could be generated using modest saline solution flush rates on the order of 50 mL/min, and with proximal control of blood flow using a coaxial balloon guide catheter, completely clear fields of view could be obtained using minimal saline flow rates as shown in Figure 2 (see Video, Supplemental Digital Content 1-2, which demonstrate renal artery inspection and microwire branch selection, http://links.lww.com/NEU/A627 and http://links.lww.com/NEU/A628). The SFE sheath was found to transmit torque relatively poorly to the distal end; however, rotation of the angled-tip guide catheter containing the SFE allowed for inspection of vessel walls and side braches from multiple angles.
FIGURE 2
A, fluoroscopic image of the scanning fiber endoscope (SFE) in place in the renal artery with a balloon in the descending aorta for proximal flow control. B, view from the SFE at the end of the guide catheter, showing a smooth white endothelium and the first bifurcation of the renal artery. C, SFE image of renal branch selection using a 0.014-inch microwire.
A, fluoroscopic image of the scanning fiber endoscope (SFE) in place in the renal artery with a balloon in the descending aorta for proximal flow control. B, view from the SFE at the end of the guide catheter, showing a smooth white endothelium and the first bifurcation of the renal artery. C, SFE image of renal branch selection using a 0.014-inch microwire.In smaller arteries where the 6-F guide catheters were nearly occlusive, the balloon guide catheter was found to be unnecessary, and clear imaging was obtained using saline solution flush rates less than 50 mL/min. As shown in Figure 3, the SFE was able to provide clear images of a microcatheter delivering aneurysm coils into a side branch and the high-resolution, real-time imaging was extremely sensitive to coil prolapse and microcatheter push-back without the need for continuous fluoroscopy (see Video, Supplemental Digital Content 3, demonstrating coil placement, prolapse, and detachment, http://links.lww.com/NEU/A629). When examining the high-resolution SFE images, it was noted with surprise that despite fastidious preparation, many of the catheters introduced initially contained numerous microbubbles, which could be seen coalescing and being dispersed into the distal circulation during device deployment, examples of which are shown in Figure 4.
FIGURE 3
Real-time scanning fiber endoscope images of aneurysm coil delivery (A) into a side branch from a microcatheter having a distal diameter of 2.8-French. B, post-filling inspection of the coil mass demonstrating a hanging detachment zone end and dense packing of coils at the level of the orificium.
FIGURE 4
Left, scanning fiber endoscope image demonstrating persistent microbubbles within the lumen of a fastidiously prepared guide catheter that were observed to dislodge randomly with increased saline solution flush rates. Right, image showing the displacement of microbubbles trapped within a flow-diverting stent during device deployment.
Real-time scanning fiber endoscope images of aneurysm coil delivery (A) into a side branch from a microcatheter having a distal diameter of 2.8-French. B, post-filling inspection of the coil mass demonstrating a hanging detachment zone end and dense packing of coils at the level of the orificium.Left, scanning fiber endoscope image demonstrating persistent microbubbles within the lumen of a fastidiously prepared guide catheter that were observed to dislodge randomly with increased saline solution flush rates. Right, image showing the displacement of microbubbles trapped within a flow-diverting stent during device deployment.Figure 5 demonstrates the ability of the SFE to visualize the key portions of mechanical thrombectomy with a stent retriever, from initial lesion inspection and microwire crossing to the inspection of the stent contents before withdrawal and assessment of residual thrombus and revascularization (see also Video, Supplemental Digital Content 4, demonstrating stent retriever deployment and withdrawal from a microcatheter allowing for visualization of stent tines and incorporated thrombus material, http://links.lww.com/NEU/A630). It was noted that if the SFE and stent retriever were used in the same sheath, then care had to be taken to avoid incorporation of the SFE tip into the tines of the stent during retrieval. Careful withdrawal of both devices under SFE guidance is necessary to avoid interference, which could lead to dislodgment of incorporated thrombus material from the retriever.
FIGURE 5
Scanning fiber endoscope images of ischemic stroke thrombectomy from initial lesion inspection (A), microwire crossing (B), initial withdrawal (C), inspection of stent retriever contents in situ (D, E), and ex vivo correlative image of retrieved material (F).
Scanning fiber endoscope images of ischemic stroke thrombectomy from initial lesion inspection (A), microwire crossing (B), initial withdrawal (C), inspection of stent retriever contents in situ (D, E), and ex vivo correlative image of retrieved material (F).The SFE was found to be very useful in imaging the placement and deployment of flow diverters as shown in Figure 6. After deployment, the angioscopic images were able to clearly assess factors such as wall apposition, stent coverage of the aneurysm base, protrusion of aneurysm coils, and side branch coverage as well as flow-diverter properties (complete opening of the device, determination of tine opening, and size of the cells). Because of the unique manner in which the SFE generates an image, it is able to realize nearly diffraction-limited image resolution assuming negligible aberration effects from the objective lens. The current SFE design has a spatial resolution less than 20 μm, which may be further decreased when using dynamic zoom10 to magnify areas of interest. In standard imaging operation, the SFE can resolve features not visible on fluoroscopy, such as the secondary windings of coils as seen in Figure 3, and can differentiate between the 28- to 33-μm15 cobalt-chromium-nickel and platinum-tungsten braided strands in the pipeline as shown in Figure 6.
FIGURE 6
Scanning fiber endoscope imaging of flow-diverter placement over a previously coiled side branch showing placement of the stent during exit from the 2.8-French delivery microcatheter (A), the ability to resolve the number of diverter tines overlying the base of the coiled branch and any coil mass movement (B), and inspection of side-wall apposition, wall kinking, and side-branch patency (C).
Scanning fiber endoscope imaging of flow-diverter placement over a previously coiled side branch showing placement of the stent during exit from the 2.8-French delivery microcatheter (A), the ability to resolve the number of diverter tines overlying the base of the coiled branch and any coil mass movement (B), and inspection of side-wall apposition, wall kinking, and side-branch patency (C).Figure 7 illustrates the ability of the SFE to detect endothelial dissections as they are developed and the utility of angioscopic imaging in differentiating the resultant true and false lumen. The SFE was also capable of detecting areas of mild endothelial damage, such as those induced by attempted dissection flap creation, which were not apparent on concurrent fluoroscopy.
FIGURE 7
Scanning fiber endoscope images of the development of an endothelial dissection. A, 0.014-inch microwire being applied in reverse showing the proximal entry point of the wire into the subendothelial space as well as the body of the wire tunneling distally. B, fully developed dissection with a damaged endothelium demonstrating the ability to differentiate the true (black arrow) and false (white arrow) lumens using angioscopy.
Scanning fiber endoscope images of the development of an endothelial dissection. A, 0.014-inch microwire being applied in reverse showing the proximal entry point of the wire into the subendothelial space as well as the body of the wire tunneling distally. B, fully developed dissection with a damaged endothelium demonstrating the ability to differentiate the true (black arrow) and false (white arrow) lumens using angioscopy.
DISCUSSION
To our knowledge, this is the first report of in vivo, high-resolution (>600 lines) endoluminal imaging during endovascular neurosurgical procedures. In cardiac applications, endoluminal imaging was previously shown to be a useful adjunct to fluoroscopy for detecting vascular pathology and guiding interventions, but the poor image quality produced by conventional fiberoptic angioscopes has greatly limited their use in routine practice.This study demonstrates the potential of endovascular intracranial imaging with an SFE for research, teaching, and clinical (diagnostic and peri-interventional) applications. From a research point of view, imaging both current and novel endovascular devices in vivo during application and at follow-up would certainly enhance our understanding of both mechanisms of action and biological responses, such as endothelial healing of aneurysms and stents. When used for long-term follow-up in animal experiments, it may be used to monitor this healing process over time without the need to euthanize the animals. The physical properties of the device can be tested in real time in the live animal, and its response to various challenges can be directly imaged.In this regard, the SFE may also play a role in training physicians in the use of new devices. Currently, training options are based on computer simulations, transparent silicone flow models, or fluoroscopic images of animal models. Although the latter has the inherent problems of only visualizing a 2-dimensional projection of the vessel “cast” or the device deployed, the former models cannot simulate the in vivo properties of vessels, side branches, and so on. Direct inspection of the device and its effects on the vessel will be useful to teach the physician how to use specific devices, what to do, and what to avoid (eg, the effects of torquing a flow diverter can be directly demonstrated).From a clinical perspective, endoluminal optical imaging may prove useful in diagnostic, peri-interventional, and postinterventional applications.
Diagnostic Utility
In the case of atherosclerosis, a significant body of work already exists describing the utility of angioscopic imaging in assessing plaque risk factors such as evidence of hemorrhage and lipid content,4,16,17 which has been collected primarily with crude angioscopes in the coronary arteries. The greatly increased resolution of the SFE may allow for better definition of the marginal extent of an atherosclerotic plaque within the cervical carotid arteries and intracranial vasculature and, with future human studies, may identify features associated with embolic risk beyond the basic size and color metrics known today. From vessel wall imaging, it is known that the vulnerable plaque has specific features including a thin or ruptured fibrous cap, a lipid-rich core, and intraplaque hemorrhage. Endoluminal imaging with an SFE may allow for improved evaluation of at least some of the salient features compared with current angiography. In the intracranial circulation, it will not only be important to visualize the plaque components but also the proximity of the plaque to perforating vessels.As shown in the porcine model, dissections and the surrounding normal structures can be visualized directly using the SFE. Arterial intra-/extracranial dissections may become symptomatic with ischemic symptoms due to various pathomechanisms: distal reopening of the endothelial damage will lead to a false lumen that can lead to arterial thromboembolic events due to slow flow and contact with the subendothelial wall. This false lumen may persist and eventually become endothelialized. False aneurysms can form at the level of the dissection with thrombus formation due to recirculating blood, or the true lumen may be compromised to a degree that hemodynamic infarctions occur. If the dissection (ie, the intramural hematoma) extends intracranially, perforating branches may be occluded. Depending on the pathomechanism of stroke in arterial dissections, different therapies may be used to treat these patients, and the SFE may be able to be of diagnostic use in differentiating between the aforementioned cases in addition to angiography. Endoluminal optical imaging may also allow for improved detection of the type of aneurysms (eg, blister, dissection, saccular) and in cases of SAH with aneurysm multiplicity, detection of the culprit lesion.
Periprocedural Applications
Because of its small size, wide-angle field of view, and high resolution, we were able to use the SFE in a coaxial arrangement with a number of endovascular tools. The SFE was extremely useful in guiding the selection of branches with a microwire, as real-time imaging of tip position was often easier to correlate with proximal end manipulations than the fluoroscopic projection imaging, especially in cases of tortuous anatomy. Following a standard fluoroscopic roadmap, it was frequently the case that branches of interest could be selected extremely quickly under SFE guidance alone, with fluoroscopy being used only for confirmation of final position instead of constant use during the procedure. Although the arteries in the porcine model were largely normal, imaging with the SFE could be of significant benefit in guiding attempts to cross complex stenotic lesions such as those of the carotid bifurcation, where the 3-dimensional structure of the lesion and path of the wire may not be readily apparent on 2-dimensional projection images.As an extension of the navigation assistance previously alluded to, we have shown that endovascular imaging with the SFE is helpful in cases in which anatomic pathologies such as a dissection may make lumen selection with fluoroscopy alone challenging. The high-resolution images provided by the SFE allowed for straightforward diagnosis of a dissection, even in early stages where they were not readily apparent on angiography. The SFE aided significantly in the differentiation between true and false lumens. SFE angioscopy could also be useful in cases of chronic dissection to determine whether the dissected portion has become endothelialized or whether it contains small amounts of thrombus not yet apparent on other imaging modalities. Considering the ability of the SFE to visualize endothelial damage directly, concurrent imaging may also aid in the prevention of iatrogenic dissection creation by allowing the operator to see when devices may be beginning to tunnel under the endothelium far before the lumen changes are visible on fluoroscopy.In cases of ischemic stroke as modeled here, the SFE is limited to en face imaging of the thrombus and does not provide any depth-resolved imaging as is possible with optical coherence tomography or intravascular ultrasound. As such, we found the greatest utility of angioscopy to be in the inspection of stent retriever integration into the clot and in monitoring the contents of the stent as they were withdrawn from the body for possible detachment and distal embolization. After thrombectomy, even in cases in which flow was returned to normal by angiographic assessment, the SFE was able to detect small areas of residual thrombus that could be removed to prevent future ischemic events, although these were not observed in this initial feasibility study. Combining the SFE with vacuum-based thrombectomy systems may allow for more complete removal of thrombus in such situations. Endoluminal imaging may also prove useful for detecting cases in which the thrombus lodges on underlying intracranial atherosclerotic plaque. Finally, the SFE can be used after stent retriever removal to determine the integrity of the wall, for example, if the endothelium was stripped during the clot retrieval, acetylsalicylic acid treatment may be considered.Perhaps the most promising initial applications suggested by this work is the use of the SFE in guiding aneurysm therapy using both coils and flow diverters. In the case of aneurysmal rupture, angioscopy could allow the identification of the rupture point and guide the placement of coils to avoid intraprocedural rupture. Similarly, it was noted in our tests that the real-time imaging could identify coil loops on the verge of prolapsing into the vessel lumen and the microcatheter could be adjusted so as to direct those coils back into the body of the aneurysm. As the SFE was generally held outside the lumen of the aneurysm after framing with coils to avoid incorporation with coils, fluoroscopic inspection was generally preferred for the assessment of coil packing inside the aneurysm body. The ability of the SFE to inspect the density of coil packing in the neck of the aneurysm at the completion of coiling, however, far exceeds the limited detail available with fluoroscopy, and the high-resolution imaging is able to detect subtle coil prolapse or loose detachment zone ends that are not readily visible on fluoroscopy.The addition of endovascular imaging during flow-diverter placement was found to be especially helpful in assessing the degree of coverage of the target lesion, which in turn can inform the need for telescoping multiple-flow diverters to achieve sufficient coverage. The resolution of the images was sufficient to visualize all of the tines of the flow diverter after deployment to the point that a metric such as the number of tine intersections over the aneurysm base could be calculated, perhaps allowing for a more quantitative approach to deciding how much coverage is “enough,” especially in the case of wide-necked aneurysms. In cases in which re-entry was necessary for diverter overlay, the SFE imaging was found to be much more sensitive than fluoroscopy to interference between the delivery catheter and tines of a previously deployed stent. The SFE was also able to assess features such as stent wall apposition, kinking when deployed around a curve, and side-branch patency, which are not possible with fluoroscopy alone. In cases in which an undersized stent was placed in a vessel, angioscopic imaging could demonstrate migration of the stent if pressure was applied using the microcatheter. From a technical standpoint, the SFE proved to be useful in visualizing and guiding the resheathing of the capture coil of the Pipeline device, a procedure that may sometimes be difficult due to tortuous anatomy and large bore of the delivery catheter with kinking of the capture coil at its distal opening. If the flow diverter had anchor tines, the SFE would also have the ability to assess their embedding into the endothelium.
Postprocedural Applications
Endoluminal imaging at the resolution afforded by the SFE may also be useful at follow-up for both coil and flow-diverter therapy assessment. We have demonstrated the ability of the system to image the endothelium in normal and traumatic vessels, and it could similarly be used to assess re-endothelialization of stented or coiled lesions to guide the withdrawal of antiplatelet therapy. In the same way, SFE angioscopy could be used to assess complications such as in-stent restenosis or anchor tine damage in the hope of improving management or identification of possible mechanisms of pathogenesis.
Limitations and Future Applications
Although we believe that these results have demonstrated a number of potential uses for high-resolution angioscopy with an SFE in endovascular applications, the SFE probe design used for these tests is the same general-purpose device used in previous gastrointestinal imaging studies and, as such, could be specifically modified for optimal intracranial use. The major limiting factor is the necessity to carry the SFE, a 3.7-F device, inside a 6-F carrying catheter to provide an atraumatic tip and saline solution flush lumen. Repackaging the device into an integrated assembly specifically designed for endovascular use would greatly reduce the diameter and allow the SFE to access even more distal regions of the circulation and provide more opportunities for use with other devices without the need for bilateral access. The rigid tip portion of the SFE can be redesigned to a 30% reduction in size (0.85-mm diameter and 6.0-mm length) without any sacrifice in performance from the current design. The angle of viewing of the endoscope may also be adjusted from the present 0° arrangement to 30°/45°/90° to allow for oblique viewing of the wall of a vessel or the contents of an aneurysm without needing to exit the lumen of the vessel.High-quality angioscopic imaging necessitates an essentially blood-free field, and, as such, many of the images presented in this paper were acquired with near-total proximal occlusion to permit inspection of the entire vessel surface simultaneously. In practice, this may be difficult to achieve in regions of intracranial circulation with large collateral supply while keeping flush volumes low. We have observed, however, that laminar blood flow in vessels tends to deflect around the tip of the guide catheter, allowing inspection of a small portion of the vessel wall with the SFE without the need for flow arrest. By rotating the guide catheter, the entire circumference of the vessel can be examined in this way, and a revised catheter tip design may allow large areas to be imaged rapidly in this manner.Endovascular imaging as proposed here carries with it risks associated with the invasiveness of the procedure and standard endovascular procedure risks, as well as the need for blood flow disruption for clear imaging. The current SFE has a shaft stiffness roughly equivalent to that of a microcatheter and, as such, is not anticipated to increase the possibility of vessel damage when introduced through a guide catheter into the intracranial circulation. However, this will need to be conclusively shown with ex vivo force testing and in vivo histopathological examination before use in humans. The perceived utility of the SFE for tasks such as postcoiling follow-up assessment will require more rigorous comparison with other approaches such as high-resolution computed tomography to evaluate the potential risk-benefit ratio.It is also worthwhile to note that the SFE represents a novel optical platform in endovascular applications and is not restricted to white-light imaging alone. Simultaneous fluorescence imaging and spectroscopy have previously been demonstrated,18,19 as have SFE probes that are capable of optical coherence tomography scanning.20 In addition to the possibility of multimodal imaging, the central scanning fiber may also be used for the delivery of light for endoluminal photodynamic therapy or higher power applications such as laser plaque ablation or cautery.10 Further technical developments, such as a steerable tip end and integrated working channel, are planned to address particular endovascular tasks.
CONCLUSION
We have demonstrated the feasibility of real-time, high-resolution endovascular imaging during common neurointerventional procedures using a novel SFE. We have shown that the SFE can detect endothelial changes and common procedural complications with much more sensitivity than is possible with fluoroscopy and that the SFE can add useful information during challenging interventional tasks. This work has only begun to demonstrate what may be possible with the fusion of fluoroscopy and next-generation angioscopy, and we anticipate future applications beyond those shown here to leverage the strengths of both modalities to improve patient care.
Disclosure
This work is supported in part by funding from the NIH including grant R01EB016457 and the Canadian Institutes of Health Research grant RMF111623. Dr Seibel is a co-inventor on a number of patents in SFE technology. The other authors have no personal, financial, or institutional interest in any of the drugs, materials, or devices described in this article.
Authors: Adam W Templeton; Kevin Webb; Joo Ha Hwang; Eric J Seibel; Michael Saunders Journal: Gastrointest Endosc Date: 2014-01-23 Impact factor: 9.427
Authors: Patrick Z McVeigh; Raphael Sacho; Robert A Weersink; Vitor M Pereira; Walter Kucharczyk; Eric J Seibel; Brian C Wilson; Timo Krings Journal: Neurosurgery Date: 2014-08 Impact factor: 4.654
Authors: A Rouchaud; C Ramana; W Brinjikji; Y-H Ding; D Dai; T Gunderson; J Cebral; D F Kallmes; R Kadirvel Journal: AJNR Am J Neuroradiol Date: 2016-07-07 Impact factor: 3.825
Authors: Ivan L Yeoh; Per G Reinhall; Martin C Berg; Howard J Chizeck; Eric J Seibel Journal: J Dyn Syst Meas Control Date: 2017-06-05 Impact factor: 1.372
Authors: Matthew J Gounis; Kajo van der Marel; Miklos Marosfoi; Mary L Mazzanti; Frédéric Clarençon; Ju-Yu Chueh; Ajit S Puri; Alexei A Bogdanov Journal: Stroke Date: 2015-09-08 Impact factor: 7.914
Authors: Luis E Savastano; Quan Zhou; Arlene Smith; Karla Vega; Carlos Murga-Zamalloa; David Gordon; Jon McHugh; Lili Zhao; Michael Wang; Aditya Pandey; B Gregory Thompson; Jie Xu; Jifeng Zhang; Y Eugene Chen; Eric J Seibel; Thomas D Wang Journal: Nat Biomed Eng Date: 2017-02-10 Impact factor: 25.671
Authors: Patrick Z McVeigh; Raphael Sacho; Robert A Weersink; Vitor M Pereira; Walter Kucharczyk; Eric J Seibel; Brian C Wilson; Timo Krings Journal: Neurosurgery Date: 2014-08 Impact factor: 4.654