Abstract
PURPOSE: To determine the reproducibility of automated segmentation of the three-dimensional (3D) lamina cribrosa (LC) microarchitecture scanned in-vivo using optical coherence tomography (OCT).Entities:
Mesh:
Year: 2014 PMID: 24747957 PMCID: PMC3991692 DOI: 10.1371/journal.pone.0095526
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Figure 1Two consecutive optical coherence tomography scans of the lamina cribrosa of the same eye.
Original C-mode (left) and segmentation overlain (right) where beams (green) and pores (red) were marked. The automated demarcation of the pores was highly repeatable between the scans.
Figure 2Two consecutive optical coherence tomography scans of the lamina cribrosa of the same eye.
Original C-mode (left) and segmentation overlain (right) where beams (green) and pores (red) are marked. Differences in segmentations between the two scans (red arrows) primarily existed due to local disparities in regions analyzed.
Figure 3Shifts in the scanning angle alter B-scan (A) and C-mode (B) images and alter the microarchitecture seen on a single frame.
OCT volume was rotated ±10° with respect to the slow scanning axis using image processing software (FIJI).
Figure 43D view of lamina cribrosa beams demonstrated the reproducibility of the microarchitectural features (same eye as Figure 1).
Average and imprecision for lamina cribrosa parameters.
| Parameters | Average (SD) | Imprecision | Relative Imprecision |
| Pore Diameter (µm) | 24.2 (1.9) | 0.4 | 1.8% |
| Pore Diameter SD (µm) | 9.8 (0.8) | 0.2 | 2.0% |
| Pore Aspect Ratio | 2.00 (0.11) | 0.04 | 1.8% |
| Pore Area (µm2) | 1660 (206) | 50 | 3.0% |
| Beam Thickness (µm) | 48.8 (2.7) | 1.0 | 2.0% |
| Beam Thickness SD (µm) | 16.1 (1.7) | 0.7 | 4.2% |
| Beam Thickness to Pore Diameter Ratio | 2.0 (0.1) | 0.1 | 1.8% |
Relative imprecision of lamina cribrosa parameters in relation to clinical diagnosis.
| Parameters | Healthy (n = 8) | Glaucoma Suspect (n = 19) | Glaucoma(n = 12) |
| Pore Diameter (µm) | 2.8% | 1.5% | 1.2% |
| Pore Diameter SD (µm) | 2.7% | 1.6% | 2.2% |
| Pore Aspect Ratio | 2.1% | 1.7% | 1.8% |
| Pore Area (µm2) | 3.0% | 2.4% | 4.2% |
| Beam Thickness (µm) | 2.0% | 2.2% | 1.8% |
| Beam Thickness SD (µm) | 4.7% | 4.7% | 2.7% |
| Beam Thickness to Pore Diameter Ratio | 1.9% | 1.9% | 1.4% |