Tracked ultrasonography snapshots enhance needle guidance for percutaneous renal access: a pilot study.

Tamas Ungi1, Darren Beiko, Michael Fuoco, Franklin King, Matthew S Holden, Gabor Fichtinger, D Robert Siemens.   

Abstract

BACKGROUND AND
PURPOSE: Although ultrasonography-guided percutaneous nephrostomy is relatively safe, a number of factors make it challenging for inexperienced operators. A computerized needle navigation technique using tracked ultrasonography snapshots was investigated to determine whether performance of percutaneous nephrostomy by inexperienced users could be improved.
METHODS: Ten operators performed the procedure on a phantom model with alternating needle guidance between conventional ultrasonography and tracked ultrasonography snapshots. The needle was reinserted until fluid backflow confirmed calyceal access. Needle trajectories were recorded using the real time needle navigation system for offline evaluation of operator performance. Recorded needle trajectories were used to measure needle motion path length inside the phantom tissue, number of reinsertions, total procedure time, and needle insertion time as end points of this study.
RESULTS: Needle path length measured inside the phantom tissue was significantly lower with ultrasonography snapshots guidance (295.0±23.1 mm, average±standard error of the mean) compared with control procedures (977.9±144.4 mm, P<0.01). This was associated with a significantly lower number of needle insertion attempts with ultrasonography snapshots (average 1.27±0.10 vs 2.83±0.31, P<0.01). The total procedure time and the needle insertion time were also significantly lower with ultrasonography snapshots guidance.
CONCLUSION: Tracked ultrasonography snapshots appear to improve the performance of percutaneous nephrostomy in these preliminary investigations, justifying further validation studies. The presented navigation system is reproducible because of commercially available hardware and open-source software components, facilitating its potential role in clinical practice.

Mesh:

Year:  2014        PMID: 24745550     DOI: 10.1089/end.2014.0011

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  1 in total

1.  Augmented reality surgical navigation with ultrasound-assisted registration for pedicle screw placement: a pilot study.

Authors:  Longfei Ma; Zhe Zhao; Fang Chen; Boyu Zhang; Ligong Fu; Hongen Liao
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-08-05       Impact factor: 2.924

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.