First performance tests of a digital photon counter (DPC) array coupled to a CsI(Tl) crystal matrix for potential use in SPECT.

Maria Georgiou1, Giacomo Borghi, Spiridon V Spirou, George Loudos, Dennis R Schaart.   

Abstract

The digital photon counter (DPC) is a recently developed type of digital silicon photomultiplier that combines low dark count rates, low readout noise, and fully digital, integrated readout circuitry with neighbor logic capability, system scalability, and MR compatibility. These are desirable properties for application in scintillation detectors for single photon emission computed tomography (SPECT). In this work, the feasibility of using a DPC array in combination with a CsI(Tl) crystal matrix as a potential detector for SPECT is investigated for the first time. Given the relatively long decay time of CsI(Tl), an important consideration is the influence on the detector performance of the DPC dark count rate as a function of temperature. We present a preliminary characterization of a detector assembled with an array of 2 × 2 × 3 mm(3) CsI(Tl) crystals. Preparatory measurements were acquired with a (57)Co source in order to optimize the light-guide thickness and the sensor settings. The spatial resolution of the detector was tested by acquiring flood maps with (57)Co as well as (99m)Tc sources. Three crystal identification algorithms were compared for the reconstruction of the flood maps. All crystal elements could be visualized clearly and high values of peak-to-valley ratios were achieved. Energy resolutions of ∼18.5% FWHM and ∼15% FWHM were measured at 122 keV and 140 keV, respectively. Temperature-dependent measurements indicate that the detector can work satisfactorily up to about 15 °C.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24743522     DOI: 10.1088/0031-9155/59/10/2415

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

Review 1.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.

Authors:  Karen Van Audenhaege; Roel Van Holen; Stefaan Vandenberghe; Christian Vanhove; Scott D Metzler; Stephen C Moore
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

2.  Initial PET performance evaluation of a preclinical insert for PET/MRI with digital SiPM technology.

Authors:  David Schug; Christoph Lerche; Bjoern Weissler; Pierre Gebhardt; Benjamin Goldschmidt; Jakob Wehner; Peter Michael Dueppenbecker; Andre Salomon; Patrick Hallen; Fabian Kiessling; Volkmar Schulz
Journal:  Phys Med Biol       Date:  2016-03-17       Impact factor: 3.609

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.