BACKGROUND: Concurrent validity and intra-rater reliability using a customized Android phone application to measure cervical-spine range-of-motion (ROM) has not been previously validated against a gold-standard three-dimensional motion analysis (3DMA) system.Entities:
Mesh:
Year: 2014 PMID: 24742001 PMCID: PMC4021613 DOI: 10.1186/1743-0003-11-65
Source DB: PubMed Journal: J Neuroeng Rehabil ISSN: 1743-0003 Impact factor: 4.262
Figure 1Experimental setup. This shows the starting position of the participant and the equipment set-up.
Figure 2Comparison of lateral flexion assessment techniques. (A) Right lateral flexion in a single axis. Note that the red dot is aligned with the horizontal line. (B) Right lateral flexion with some cervical flexion. Note that the red dot is not aligned with the horizontal line.
Validity of the phone compared to 3DMA using 3 repetitions of each cervical movement (n = 21)
| Flexion | 52.0 ± 8.7 | 49.9 ± 8.8 | 0.98 | 0.99 | 0.30 to 0.996 | None | 2.3 | 2 | N | N |
| Extension | 79.3 ± 8.0 | 80.4 ± 9.9 | 0.92 | 0.83 | 0.80 to 0.97 | None | 9.6 | 6 | N | N |
| Right Lateral Flexion¥ | 45.0 ± 7.3 | 43.0 ± 7.0 | 0.96 | 0.93 | 0.71 to 0.99 | None | 4.6 | 5 | N | N |
| Left Lateral Flexion | 48.8 ± 8.8 | 47.8 ± 8.0 | 0.95 | 0.92 | 0.89 to 0.98 | None | 7.1 | 7 | N | N |
| Right Rotation | 57.1 ± 9.7 | 70.9 ± 7.2 | 0.53 | 0.81 | -0.13 to 0.85 | -33.7 + 0.31 | 9.6 | 8 | Y | Y |
| Left Rotation | 65.3 ± 15.1 | 71.4 ± 5.8 | 0.53 | 0.52 | -0.60 to 0.80 | -71.2 + 0.95 | 18.6 | 14 | Y | Y |
SD = standard deviation; 3DMA = three dimensional motion analysis; ICC = intra-class coefficients; CI = confidence interval; LoA = limits of agreement; Prop Bias = proportional bias; N = no; Y = yes.
*All correlations were p < 0.001 except Left Rotation (p = 0.02).
¥Based on n = 20 as markers on one subject were missing.
†% Error = 0.5*Width of 95% LOA/[(MeanPhone + MeanVicon)/2].
€Proportional and fixed bias were determined from ordinary least products analysis.
Intra-rater reliability of the phone (n = 16)
| Flexion | 51.3 ± 7.9 | 54.9 ± 7.5 | 0.86 | 0.38-0.96 | N | 9.1 | 9 | N | N | 3.1 | 9.2 | -12.84 to 5.48 |
| Extension¥ | 79.0 ± 7.6 | 80.8 ± 7.03 | 0.82 | 0.49-0.94 | N | 11.8 | 7 | N | N | 5.0 | 11.9 | -13.67 to 10.3 |
| Right Lateral Flexion | 43.5 ± 6.7 | 44.9 ± 7.0 | 0.90 | 0.73-0.97 | N | 8.3 | 9 | N | N | 2.8 | 8.3 | -9.73 to 6.95 |
| Left Lateral Flexion | 49.1 ± 8.8 | 51.2 ± 7.4 | 0.85 | 0.57-0.95 | N | 11.8 | 12 | N | N | 4.1 | 12.2 | -14.20 to 10.16 |
| Right Rotation | 50.0 ± 17.1 | 70.5 ± 22.7 | 0.33 | -0.34-0.73 | N | 48.2 | 40 | N | N | 16.4 | 48.7 | -70.79 to 29.81 |
| Left Rotation | 64.3 ± 16.3 | 69.8 ± 15.6 | 0.05 | -1.7-0.67 | N | 46.8 | 35 | N | N | 15.8 | 46.9 | -52.38 to 41.36 |
SD = standard deviation; ICC = intra-class coefficients; CI = confidence interval; Sys Bias = systematic bias; LoA = limits of agreement; Prop Bias = proportional bias; N = no; Y = yes; SEM = standard error of measurement; MDC = minimal detectable change; diff = difference.
¥Based on n = 15 as one subject’s thoracic spine was not well stabilized.
†% Error = 0.5*Width of 95% LOA/[(MeanPhone + MeanVicon)/2].
€Proportional and fixed bias were determined from ordinary least products analysis.
Reliability of the 3DMA (n = 16)
| Flexion | 48.9 ± 7.7 | 51.9 ± 6.9 | 0.88 | 0.54-0.96 | N | 8.7 | 9 | N | N | 3.0 | 8.91 | -14.77 to 8.90 |
| Extension¥ | 79.1 ± 9.9 | 81.6 ± 9.2 | 0.88 | 0.67-0.96 | N | 10.0 | 6 | N | N | 3.4 | 10.1 | -12.69 to 7.56 |
| Right Lateral Flexion | 41.7 ± 6.7 | 42.9 ± 6.9 | 0.94 | 0.82-0.98 | N | 6.8 | 8 | N | N | 2.3 | 6.83 | -8.01 to 5.51 |
| Left Lateral Flexion | 46.7 ± 7.6 | 47.1 ± 6.6 | 0.92 | 0.78-0.97 | N | 8.0 | 9 | N | N | 2.8 | 8.32 | -8.57 to 7.89 |
| Right Rotation | 68.8 ± 5.1 | 72.7 ± 5.9 | 0.64 | -0.010-0.88 | N | 10.4 | 7 | N | N | 3.6 | 10.69 | -14.45 to 6.81 |
| Left Rotation | 70.2 ± 6.7 | 73.4 ± 6.7 | 0.77 | 0.32-0.92 | N | 11.2 | 8 | N | N | 3.8 | 11.29 | -14.35 to 8.03 |
SD = standard deviation; ICC = intra-class coefficients; CI = confidence interval; Sys Bias = systematic bias; LoA = limits of agreement; Prop Bias = proportional bias; N = no; Y = yes; SEM = standard error of measurement; MDC = minimal detectable change; diff = difference.
¥Based on n = 15 as one subject’s thoracic spine could not be effectively stabilized using the experimental technique utilized.
†% Error = 0.5*Width of 95% LOA/[(MeanPhone + MeanVicon)/2].
€Proportional and fixed bias were determined from ordinary least products analysis.