Reliability of pH measurement and the auscultatory method to confirm the position of a nasogastric tube.

Kurt Boeykens1, Els Steeman2, Ivo Duysburgh3.   

Abstract

BACKGROUND: Blind placement of a nasogastric feeding tube is a common nursing procedure. Confirmation of the correct position in the stomach is warranted to avoid serious complications such as misplacement in the lung. Testing pH of aspirate from a tube is one of the techniques to confirm the tip position. The purpose of this study was to evaluate the auscultatory method and pH measurement with a pH cut-off point of 5.5 after tube insertion and to compare this with the 'gold standard': an abdominal X-ray. Also the feasibility of the pH method was evaluated.
MATERIALS AND METHODS: Large prospective observational study in a general hospital. In adult hospitalised patients, the positioning of 331 feeding tubes was tested using two different methods to predict tube position in the stomach.
RESULTS: In 98.9% (n=178) of aspirate samples with a pH ≤ 5.5, the tube was located in the stomach. If an aspirate could be obtained, the results of pH measurements showed a sensitivity of 78.4% and a specificity of 85.7%. Obtaining aspirate initially after placement was possible in approximately half of cases but after taking additional measures (including administration of air into the tube, side-positioning of the patient and re-aspiration after one hour) this increased to 81.6%. The sensitivity of the auscultatory method was 79% while the specificity was 61%.
CONCLUSIONS: A pH of ≤ 5.5 from tube aspirate is adequate to check the position of the tube in the stomach. Additional measures improve the success to obtain an aspirate from the tube. The auscultatory method is unreliable.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Auscultatory method; Nasogastric feeding; Nasogastric tube; X-ray; pH; pH measurement

Mesh:

Year:  2014        PMID: 24731474     DOI: 10.1016/j.ijnurstu.2014.03.004

Source DB:  PubMed          Journal:  Int J Nurs Stud        ISSN: 0020-7489            Impact factor:   5.837


  8 in total

1.  Novel Visual Nasogastric Tube Insertion System: A Feasibility and Efficiency Study in a Manikin.

Authors:  Qiaoya Li; Juan Xie; Jinxing Wu; Rui Guo; Wenwen Ma; Gang Xu; Min Yang; Huisheng Deng
Journal:  Gastroenterol Res Pract       Date:  2016-11-23       Impact factor: 2.260

2.  Diagnostic accuracy of a pH stick, modified to detect gastric lipase, to confirm the correct placement of nasogastric tubes.

Authors:  Anne M Rowat; Catriona Graham; Martin Dennis
Journal:  BMJ Open Gastroenterol       Date:  2018-07-31

3.  Study to determine the likely accuracy of pH testing to confirm nasogastric tube placement.

Authors:  Anne M Rowat; Catriona Graham; Martin Dennis
Journal:  BMJ Open Gastroenterol       Date:  2018-06-09

4.  Development and validation of ester impregnated pH strips for locating nasogastric feeding tubes in the stomach-a multicentre prospective diagnostic performance study.

Authors:  Melody Ni; Mina E Adam; Fatima Akbar; Jeremy R Huddy; Simone Borsci; Peter Buckle; Francesca Rubulotta; Reuben Carr; Ian Fotheringham; Claire Wilson; Matthew Tsang; Susan Harding; Nichola White; George B Hanna
Journal:  Diagn Progn Res       Date:  2021-12-14

5.  A new strategy for enteral nutrition using a deflection flexible visual gastric tube: A randomized crossover manikin trial.

Authors:  Jie Li; Yan-Mei Feng; Dong Wan; Hui-Sheng Deng; Rui Guo
Journal:  Medicine (Baltimore)       Date:  2018-05       Impact factor: 1.889

6.  Improving knowledge, attitudes, and practice of nurses in medication administration through enteral feeding tubes by clinical pharmacists: a case-control study.

Authors:  Samira Hossaini Alhashemi; Raana Ghorbani; Afsaneh Vazin
Journal:  Adv Med Educ Pract       Date:  2019-07-09

7.  Accuracy of the combined method (auscultation and pH measurement) and ultrasonography for confirmation of gastric tube placement: a study protocol for a prospective study.

Authors:  Mayara Carvalho Godinho Rigobello; Jorge Elias Junior; Carlos Alberto Grespan Bonacim; Renata Cristina de Campos Pereira Silveira; Fernanda Caroline Bonardi; Roosevelt Santos Nunes; Rosana Aparecida Pereira; Fernanda Raphael Escobar Gimenes
Journal:  BMJ Open       Date:  2020-09-30       Impact factor: 2.692

8.  Ultrasonographic Confirmation of Nasogastric Tube Placement in the COVID-19 Era.

Authors:  Vasiliki Tsolaki; George E Zakynthinos; Paris Zygoulis; Fotini Bardaka; Aikaterini Malita; Vasileios Aslanidis; Epaminondas Zakynthinos; Demosthenes Makris
Journal:  J Pers Med       Date:  2022-02-23
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.