Longitudinal deformation of drug-eluting stents: evaluation by multislice computed tomography.

Rafael Romaguera1, Gerard Roura, Josep Gomez-Lara, Jose L Ferreiro, Montserrat Gracida, Luis Teruel, Matias de Albert, Albert Ariza, Joan A Gomez-Hospital, Angel Cequier.   

Abstract

BACKGROUND: Some modifications introduced in the design of the new generation of drug-eluting stent (DES) to improve their flexibility may entail a reduction in their longitudinal strength. This study sought to evaluate the longitudinal deformation of DESs by multislice computed tomography (MSCT).
METHODS: This study included DESs that could have been potentially deformed by mechanical actions such as: (1) catheter impingement; (2) postdilation; (3) kissing balloon; and (4) intravascular imaging after implantation. Patients on atrial fibrillation or with overlapping stents were excluded. All patients underwent stent length evaluation by MSCT 9-12 months after implantation.
RESULTS: Forty-five stents were included: 15 platinum chromium (PtCr-DES), 15 cobalt chromium (CoCr-DES), and 15 stainless-steel (SS-DES). The relative longitudinal deformation by stent type was 6.93 ± 5.82% for PtCr-DES, 6.19 ± 5.79% for CoCr- DES, and 4.03 ± 4.07% for SS-DES (P=.31). Among the mechanical actions studied, only catheter impingement was related to longitudinal stent deformation (P<.01). After adjustment, only catheter impingement (P<.01) and nominal stent length (P=.049) were independently related to longitudinal deformation. There were no stent fractures.
CONCLUSIONS: Longitudinal deformation of DESs is common in all the studied platforms when subject to longitudinal forces. Guiding catheter impingement is the only mechanical action significantly associated with DES shortening.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24717272

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  3 in total

1.  Stent fracture and longitudinal compression detected on coronary CT angiography in the first- and new-generation drug-eluting stents.

Authors:  Mi Sun Chung; Dong Hyun Yang; Young-Hak Kim; Jae-Hyung Roh; Jihyun Song; Joon-Won Kang; Jung-Min Ahn; Duk-Woo Park; Soo-Jin Kang; Seung-Whan Lee; Cheol Whan Lee; Seong-Wook Park; Seung-Jung Park; Tae-Hwan Lim
Journal:  Int J Cardiovasc Imaging       Date:  2015-10-26       Impact factor: 2.357

2.  Left Circumflex Coronary Artery as the Culprit Vessel in ST-Segment-Elevation Myocardial Infarction.

Authors:  Diab Ghanim; Fabio Kusniec; Wadi Kinany; Dahud Qarawani; David Meerkin; Khaled Taha; Offer Amir; Shemy Carasso
Journal:  Tex Heart Inst J       Date:  2017-10-01

3.  Longitudinal stent elongation or shortening after deployment in the coronary arteries: which is dominant?

Authors:  Magdy Algowhary; Mohammed Aboel-Kassem F Abdelmegid
Journal:  Egypt Heart J       Date:  2021-05-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.