| Literature DB >> 24714654 |
Kristel Schaap1, Yvette Christopher-de Vries, Catherine K Mason, Frank de Vocht, Lützen Portengen, Hans Kromhout.
Abstract
OBJECTIVES: Limited data is available about incidence of acute transient symptoms associated with occupational exposure to static magnetic stray fields from MRI scanners. We aimed to assess the incidence of these symptoms among healthcare and research staff working with MRI scanners, and their association with static magnetic field exposure.Entities:
Mesh:
Year: 2014 PMID: 24714654 PMCID: PMC4033112 DOI: 10.1136/oemed-2013-101890
Source DB: PubMed Journal: Occup Environ Med ISSN: 1351-0711 Impact factor: 4.402
Symptoms and symptom categories included in the study
| Symptom as described in diary | Shortened name | Category |
|---|---|---|
| Sensation of dizziness or vertigo | Vertigo | SMF target—core |
| Nausea | Nausea | SMF target—core |
| Tinnitus/sensation of head ringing | Head ringing | SMF target—core |
| Seeing light spots or light flashes | Magnetophosphenes | SMF target—core |
| Metallic taste | Metallic taste | SMF target—core |
| Headache | Headache | SMF target |
| Tiredness or sleepiness | Tiredness | SMF target |
| Concentration problems | Concentration problems | SMF target |
| Vomiting | Vomiting | SMF target |
| Feeling of instability when standing, walking or moving | Instability | SMF target |
| Feeling lightheaded or weightless | Lightheadedness | SMF target |
| Blurred or double vision | Blurred vision | SMF target |
| A strange smell sensation | Strange smell | SMF target |
| Tingling sensation in the body | Tingling | SGF target |
| Involuntary muscle contractions | Muscle contractions | SGF target |
| Seeing black spots or having a temporary loss of vision | Black spots | A priori unrelated |
| Itchy, watery or red eyes | Irritated eyes | A priori unrelated |
| Sensation of glowing, burning or irritated skin | Irritated skin | A priori unrelated |
| Suddenly feeling warm or hot, hot flashes | Hot flashes | A priori unrelated |
| Earache | Earache | A priori unrelated |
| Palpitation | Palpitation | A priori unrelated |
SGF, switched gradient field; SMF, static magnetic field.
Description of the collected data
| nsubj (%) Total=331 | nobs (%) Total=633 | |||
|---|---|---|---|---|
| Job title | ||||
| Radiographer, radiography student or intern* | 197 | (60) | 418 | (66) |
| Scientific staff or research student | 101 | (31) | 172 | (27) |
| Clinical staff | 19 | (6) | 21 | (3) |
| Technical staff | 10 | (3) | 17 | (3) |
| Unknown | 4 | (1) | 5 | (1) |
| Gender | ||||
| Female | 188 | (57) | 354 | (56) |
| Male | 139 | (42) | 274 | (43) |
| Unknown | 4 | (1) | 5 | (1) |
| Scanner category† | ||||
| Unexposed | 129 | (39) | 134 | (20) |
| 1.5 T closed bore‡ | 142 | (43) | 259 | (38) |
| 3.0 T closed bore§ | 89 | (27) | 131 | (19) |
| 7.0 T closed bore¶ | 25 | (8) | 31 | (5) |
| Various scanner types up to 1.5 T** | 34 | (10) | 49 | (7) |
| Small-bore animal scanners >4.7 T†† | 34 | (10) | 57 | (8) |
| Unknown | 13 | (4) | 14 | (2) |
| Reported workload level† | ||||
| Low | 135 | (41) | 180 | (28) |
| Medium | 226 | (68) | 366 | (58) |
| High | 72 | (22) | 86 | (14) |
| Unknown | 1 | (0) | 1 | (0) |
| Used disinfectants or solvents during work† | ||||
| Yes | 235 | (71) | 417 | (66) |
| No | 141 | (43) | 210 | (33) |
| Unknown | 5 | (2) | 6 | (1) |
| Consumed alcohol during 24 h prior to shift† | ||||
| Yes | 124 | (38) | 165 | (26) |
| No | 257 | (78) | 438 | (69) |
| Unknown | 22 | (7) | 30 | (5) |
*This group includes 68 observations obtained from 67 CT radiographers and CT radiography interns. The remaining observations are from MRI radiographers, students and interns.
†Due to multiple observations per subject, a subject can appear in more than one category. Therefore, the sum might exceed 100%. This also applies to the number of shifts per scanner category, since within a shift a subject could have worked with more than one type of scanner.
‡Includes all large-bore cylindrical systems of 1.5 Tesla.
§Includes all large-bore cylindrical systems of 3.0 Tesla.
¶Includes all large-bore cylindrical systems of 7.0 Tesla.
**Includes extremity scanners, upright scanners and open (ie, transversal field) scanners up to 1.5 Tesla and large-bore cylindrical systems below 1.5 Tesla.
††Includes small-bore cylindrical animal scanners with field strengths between 4.7 and 11.7 Tesla.
Nsubj, number of subjects; Nobs, number of observations; T, Tesla.
Reported symptoms by scanner category*
| Reporting of at least one symptom within the specified category (nobs(%)) | |||||||
|---|---|---|---|---|---|---|---|
| Scanner category | Overall nobs | SMF target symptoms (%) | SMF target—core symptoms (%) | A priori unrelated symptoms (%) | |||
| Unexposed | 134 | 25 | (19) | 1 | (1) | 4 | (3) |
| 1.5 T closed bore† | 259 | 72 | (28) | 14 | (5) | 17 | (7) |
| 3 T closed bore‡ | 131 | 46 | (35) | 12 | (9) | 3 | (2) |
| 7 T closed bore§ | 31 | 12 | (39) | 9 | (29) | 0 | (0) |
| <1.5 T various types¶ | 49 | 13 | (27) | 2 | (4) | 7 | (14) |
| >4.7 T small bore** | 57 | 9 | (16) | 0 | (0) | 4 | (7) |
| Overall | 619 | 162 | (26) | 35 | (6) | 33 | (5) |
*Scanner category was missing for 14 of the 633 observations, resulting in 619 valid observations. One shift can satisfy more than one scanner category; therefore, the sum of observations exceeds the total number of 619 observations. Results are based on an unbalanced study sampling design with repeated measures per subject.
†Includes all large-bore cylindrical systems of 1.5 Tesla.
‡Includes all large-bore cylindrical systems of 3.0 Tesla.
§Includes all large-bore cylindrical systems of 7.0 Tesla.
¶Includes extremity scanners, upright scanners and open (ie, transversal field) scanners up to 1.5 Tesla and large-bore cylindrical systems below 1.5 Tesla.
**Includes small-bore cylindrical animal scanners with field strengths between 4.7 and 11.7 Tesla.
Nobs, number of observations; SMF, static magnetic field; T, Tesla.
Adjusted* ORs for scanner category in the ‘symptom-reporting’ class
| SMF target symptoms | SMF target—core symptoms | A priori unrelated symptoms | ||||
|---|---|---|---|---|---|---|
| Scanner category | OR | 95% CI | OR | 95% CI | OR | 95% CI |
| Unexposed (reference) | 1 | 1 | 1 | |||
| 1.5 T closed bore† | 1.88 | 1.07–3.31 | 2.41 | 0.72–8.13 | 0.75 | 0.17–3.34 |
| 3.0 T closed bore‡ | 2.14 | 1.13–4.03 | 4.92 | 1.21–19.97 | 0.28 | 0.05–1.62 |
| 7.0 T closed bore§ | 4.17 | 1.30–13.35 | 66.82 | 6.39–698.97 | n.a. | n.a. |
| <1.5 T various types¶ | 1.47 | 0.59–3.64 | 0.84 | 0.14–5.19 | 1.53 | 0.25–9.41 |
| >4.7 T small bore** | 0.72 | 0.27–1.94 | n.a. | n.a. | 17.70 | 0.04–7379.05 |
*Adjusted for gender, age (18–27, 28–40, 40–65 years), workload, use of solvents and alcohol consumption during previous 24 h.
†Includes all large-bore cylindrical systems of 1.5 Tesla.
‡Includes all large-bore cylindrical systems of 3.0 Tesla.
§Includes all large-bore cylindrical systems of 7.0 Tesla.
¶Includes extremity scanners, upright scanners and open (ie, transversal field) scanners up to 1.5 Tesla and large-bore cylindrical systems below 1.5 Tesla.
**Includes small-bore cylindrical animal scanners with field strengths between 4.7 and 11.7 Tesla.
SMF, static magnetic field; T, Tesla.