Literature DB >> 24705004

Validation study of Nexfin® continuous non-invasive blood pressure monitoring in critically ill adult patients.

K Ameloot1, K Van De Vijver, N Van Regenmortel, I De Laet, K Schoonheydt, H Dits, O Broch, B Bein, M L Malbrain.   

Abstract

BACKGROUND: Nexfin® (BMEYE, Amsterdam, The Netherlands) is a totally non-invasive blood pressure and cardiac output (CO) monitor based on finger arterial pulse contour analysis.
METHODS: We performed an open observational study in a mix of medical-surgical-burns critically ill patients (N.=45) to validate Nexfin obtained blood pressures (MAPnex) against PiCCO (MAPfem) derived blood pressure measurements. MAPnex, MAPfem and corresponding systolic (SBP) and diastolic (DBP) blood pressures were measured continuously and registered with a 2 hour interval during the 8-hour study period. Statistical analysis was performed by Pearson regression, Bland and Altman, Concordance plot and Polar plot analysis.
RESULTS: MAPnex shows excellent correlation with MAPfem (R² 0.88, mean bias ± LA -2.3±12.4 mmHg, 14.7% error) and may be used interchangeably with invasive monitoring. The excellent MAPnex -MAPfem correlation was preserved in subgroup analysis for patients with severe hypotension, high systemic vascular resistance, low CO, hypothermia and in patients supported by inotropic/vasopressive agents. MAPnex is able to follow changes in MAPfem during the same time interval (level of concordance 85.5%). Nexfin SBP and DBP show significant correlation with PiCCO but the criteria for interchangeability were not met. Finally, polar plot analysis showed that trending capabilities were excellent when changes in MAPnexMAPnex) were compared to ΔMAPfem (96.1% of changes were within the level of 10% limits of agreement).
CONCLUSION: In this sample of critically ill patients we found a good correlation between MAPnex and invasive blood pressures obtained by PiCCO.

Entities:  

Mesh:

Year:  2014        PMID: 24705004

Source DB:  PubMed          Journal:  Minerva Anestesiol        ISSN: 0375-9393            Impact factor:   3.051


  11 in total

Review 1.  [Goal-directed hemodynamic therapy: Concepts, indications and risks].

Authors:  S A Haas; B Saugel; C J Trepte; D A Reuter
Journal:  Anaesthesist       Date:  2015-07       Impact factor: 1.041

2.  Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Miho Sumiyoshi; Takuma Maeda; Erika Miyazaki; Naoshi Hotta; Hitoshi Sato; Eisuke Hamaguchi; Hiroko Kanazawa; Yoshihiko Ohnishi; Masataka Kamei
Journal:  J Anesth       Date:  2019-03-23       Impact factor: 2.078

3.  Continuous noninvasive monitoring of arterial pressure using the vascular unloading technique in comparison to the invasive gold standard in elderly comorbid patients: A prospective observational study.

Authors:  Phil Klose; Ulf Lorenzen; Rouven Berndt; Christoph Borzikowsky; Moritz Hill; Matthias Gruenewald; Gunnar Elke; Jochen Renner
Journal:  Health Sci Rep       Date:  2020-11-10

4.  Enhancing Cardiac Rehabilitation With Stress Management Training: A Randomized, Clinical Efficacy Trial.

Authors:  James A Blumenthal; Andrew Sherwood; Patrick J Smith; Lana Watkins; Stephanie Mabe; William E Kraus; Krista Ingle; Paula Miller; Alan Hinderliter
Journal:  Circulation       Date:  2016-03-21       Impact factor: 29.690

Review 5.  Multi-Site Photoplethysmography Technology for Blood Pressure Assessment: Challenges and Recommendations.

Authors:  Gabriel Chan; Rachel Cooper; Manish Hosanee; Kaylie Welykholowa; Panayiotis A Kyriacou; Dingchang Zheng; John Allen; Derek Abbott; Nigel H Lovell; Richard Fletcher; Mohamed Elgendi
Journal:  J Clin Med       Date:  2019-11-01       Impact factor: 4.241

6.  Comparison of the non-invasive Nexfin® monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients.

Authors:  Felix Balzer; Marit Habicher; Michael Sander; Julian Sterr; Stephanie Scholz; Aarne Feldheiser; Michael Müller; Carsten Perka; Sascha Treskatsch
Journal:  J Int Med Res       Date:  2016-05-02       Impact factor: 1.671

Review 7.  Improving Perioperative Outcomes Through Minimally Invasive and Non-invasive Hemodynamic Monitoring Techniques.

Authors:  Takashige Yamada; Susana Vacas; Yann Gricourt; Maxime Cannesson
Journal:  Front Med (Lausanne)       Date:  2018-05-17

8.  Validation of noninvasive continuous arterial pressure measurement by ClearSight System™ during induction of anesthesia for cardiovascular surgery.

Authors:  Tadashi Tanioku; Akari Yoshida; Yuichi Aratani; Keisuke Fujii; Tomoyuki Kawamata
Journal:  BMC Anesthesiol       Date:  2020-07-20       Impact factor: 2.217

9.  Augmented Cardiopulmonary Baroreflex Sensitivity in Intradialytic Hypertension.

Authors:  Sook H Park; Ida T Fonkoue; Yunxiao Li; Dana R DaCosta; Holly R Middlekauff; Jeanie Park
Journal:  Kidney Int Rep       Date:  2018-08-17

10.  Multimodal Monitoring of Cardiovascular Responses to Postural Changes.

Authors:  Arjen Mol; Andrea B Maier; Richard J A van Wezel; Carel G M Meskers
Journal:  Front Physiol       Date:  2020-03-03       Impact factor: 4.566

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.