Measuring level of agreement between values obtained by directly measured blood pressure and ultrasonic Doppler flow detector in cats.

Anderson F da Cunha1, Katrin Saile, Hugues Beaufrère, Wendy Wolfson, Diana Seaton, Mark J Acierno.   

Abstract

OBJECTIVE: To determine if blood pressure measured with an ultrasonic Doppler flow detector (Doppler) is in good agreement with directly measured blood pressures in anesthetized cats.
DESIGN: Prospective observational study.
SETTING: University veterinary teaching hospital. ANIMALS: Thirty-nine cats undergoing routine neutering.
INTERVENTIONS: Cats were divided into 2 groups; 19 cats enrolled in Group A had a 24-Ga catheter inserted into a dorsal pedal artery; 20 cats in Group B had a 20-Ga catheter placed in a femoral artery. In both groups, systolic, diastolic, and mean arterial pressures were directly measured using a validated pressure measurement system. Indirect values were compared against direct blood pressure measurements.
RESULTS: There was no difference between groups. Overall, there was poor agreement with a significant bias observed between Doppler and directly measured blood pressures. For the systolic arterial pressure the bias was -8.8 with limits of agreements (LOA) of -39.3 and 21.7. For the mean arterial pressure, the bias was 14.0 with LOA of -13.9 and 41.9. For the diastolic arterial pressure, the bias was 27.9 with LOA of -4.4 and 60.2. Methodology, weight, sex, and replicates did not have a significant effect on the difference between indirect and direct measurements in any model.
CONCLUSIONS: Results suggest poor agreement between Doppler values and directly measured blood pressures in anesthetized cats. Use of Doppler in cats could be misleading and readings should be interpreted with caution in a clinical context. © Veterinary Emergency and Critical Care Society 2014.

Entities:  

Keywords:  Doppler; blood pressure; cardiovascular; cat; monitoring

Mesh:

Year:  2014        PMID: 24697972     DOI: 10.1111/vec.12161

Source DB:  PubMed          Journal:  J Vet Emerg Crit Care (San Antonio)        ISSN: 1476-4431


  6 in total

1.  Comparison of Direct and Indirect Methods of Measuring Arterial Blood Pressure in Healthy Male Rhesus Macaques (Macaca mulatta).

Authors:  Logan K France; Meghan S Vermillion; Caroline M Garrett
Journal:  J Am Assoc Lab Anim Sci       Date:  2018-01-01       Impact factor: 1.232

2.  Vasopressor use in 41 critically ill cats (2007-2016).

Authors:  Nikki Licht; Elizabeth A Rozanski; John E Rush
Journal:  Can Vet J       Date:  2018-11       Impact factor: 1.008

3.  Oscillometric and Doppler arterial blood pressure measurement in conscious goats.

Authors:  Olga Szaluś-Jordanow; Michał Czopowicz; Anna Świerk; Oktawia Szpinda; Magdalena Garncarz; Marcin Mickiewicz; Agata Moroz; Emilia Bagnicka; Jarosław Kaba
Journal:  Can J Vet Res       Date:  2018-10       Impact factor: 1.310

4.  Blood Pressure Measurements in 780 Apparently Healthy Cats.

Authors:  J R Payne; D C Brodbelt; V Luis Fuentes
Journal:  J Vet Intern Med       Date:  2016-12-01       Impact factor: 3.333

5.  Comparison of oscillometric, Doppler and invasive blood pressure measurement in anesthetized goats.

Authors:  Olga Szaluś-Jordanow; Michał Czopowicz; Agata Moroz; Marcin Mickiewicz; Magdalena Garncarz; Emilia Bagnicka; Tadeusz Frymus; Jarosław Kaba
Journal:  PLoS One       Date:  2018-05-23       Impact factor: 3.240

6.  'The Mercury Challenge': feline systolic blood pressure in primary care practice - a European survey.

Authors:  Andrew Sparkes; Catherine Garelli-Paar; Thomas Blondel; Emilie Guillot
Journal:  J Feline Med Surg       Date:  2022-06-27       Impact factor: 1.971

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.