Literature DB >> 24697405

Fatigue and nanomechanical properties of K3XF nickel-titanium instruments.

Y Shen1, H Zhou, L Campbell, Z Wang, R Wang, T Du, M Haapasalo.   

Abstract

AIM: To examine the fatigue behaviour of heat-treated NiTi instruments when immersed in aqueous media and to determine the effect of cyclic fatigue on the hardness and elastic modulus of NiTi instruments using a nanoindentation technique.
METHODOLOGY: K3XF and K3 NiTi instruments, both in sizes 25, 0.04 taper and 40, 0.04 taper, were subjected to rotational bending at a curvature of 42° either in air or under deionized water, and the number of revolutions to fracture (Nf ) was recorded. The fracture surface of all fragments was examined with a scanning electron microscope. The hardness and elastic modulus of the fracture surface of instruments sized 25, 0.04 taper were then measured using a nanoindentation test.
RESULTS: The K3XF instruments had a fatigue resistance superior to K3 instruments under dry and aqueous environments (P < 0.05). The fatigue life of K3 instruments was similar under both conditions, whereas the Nf of K3XF was greater under water than in air, especially at the size 40, 0.04 taper (P < 0.05). The values for the fraction of the area occupied by the dimple region were significantly smaller in K3XF instruments than in K3 instruments, especially under water (P < 0.05). There was no difference in hardness on K3XF instruments between new files and instruments subjected to the fatigue process. The hardness of instruments subjected to the fatigue process was significantly lower in K3XF than in K3 instruments (P < 0.05).
CONCLUSION: The fatigue life of K3XF instruments under water is longer than it is for K3XF instruments in air. There was no work-hardening effect on K3XF instruments subjected to the fatigue process.
© 2014 International Endodontic Journal. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  K3XF; cyclic fatigue; hardness; nanoindentation; nickel-titanium instrument

Mesh:

Substances:

Year:  2014        PMID: 24697405     DOI: 10.1111/iej.12265

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

2.  Comparative evaluation of cyclic fatigue resistance of four different nickel-titanium rotary files with different cross-sectional designs and alloy properties.

Authors:  Mehmet Emin Kaval; Ismail Davut Capar; Hüseyin Ertas; Bilge Hakan Sen
Journal:  Clin Oral Investig       Date:  2016-07-26       Impact factor: 3.573

Review 3.  Rotary science and its impact on instrument separation: A focused review.

Authors:  Sandhya Anand Khasnis; Prem Prakash Kar; Apoorva Kamal; Jayaprakash D Patil
Journal:  J Conserv Dent       Date:  2018 Mar-Apr

Review 4.  Effect of body temperature on the cyclic fatigue resistance of the nickel-titanium endodontic instruments: A systematic review and meta-analysis of in vitro studies.

Authors:  Selventhra Savitha; Sidhartha Sharma; Vijay Kumar; Amrita Chawla; Perumal Vanamail; Ajay Logani
Journal:  J Conserv Dent       Date:  2022-08-02

5.  Fracture Resistance of K3 Nickel-Titanium Files Made from Different Thermal Treatments.

Authors:  JinWook Choi; Soram Oh; Yu-Chan Kim; Kwang-Koo Jee; KeeYeon Kum; SeokWoo Chang
Journal:  Bioinorg Chem Appl       Date:  2016-11-14       Impact factor: 7.778

  5 in total