Implications of CT noise and artifacts for quantitative 99mTc SPECT/CT imaging.

K W Hulme1, S C Kappadath1.   

Abstract

PURPOSE: This paper evaluates the effects of computed tomography (CT) image noise and artifacts on quantitative single-photon emission computed-tomography (SPECT) imaging, with the aim of establishing an appropriate range of CT acquisition parameters for low-dose protocols with respect to accurate SPECT attenuation correction (AC).
METHODS: SPECT images of two geometric and one anthropomorphic phantom were reconstructed iteratively using CT scans acquired at a range of dose levels (CTDIvol = 0.4 to 46 mGy). Resultant SPECT image quality was evaluated by comparing mean signal, background noise, and artifacts to SPECT images reconstructed using the highest dose CT for AC. Noise injection was performed on linear-attenuation (μ) maps to determine the CT noise threshold for accurate AC.
RESULTS: High levels of CT noise (σ ∼ 200-400 HU) resulted in low μ-maps noise (σ ∼ 1%-3%). Noise levels greater than ∼ 10% in 140 keV μ-maps were required to produce visibly perceptible increases of ∼ 15% in (99m)Tc SPECT images. These noise levels would be achieved at low CT dose levels (CTDIvol = 4 μGy) that are over 2 orders of magnitude lower than the minimum dose for diagnostic CT scanners. CT noise could also lower (bias) the expected μ values. The relative error in reconstructed SPECT signal trended linearly with the relative shift in μ. SPECT signal was, on average, underestimated in regions corresponding with beam-hardening artifacts in CT images. Any process that has the potential to change the CT number of a region by ∼ 100 HU (e.g., misregistration between CT images and SPECT images due to motion, the presence of contrast in CT images) could introduce errors in μ140 keV on the order of 10%, that in turn, could introduce errors on the order of ∼ 10% into the reconstructed (99m)Tc SPECT image.
CONCLUSIONS: The impact of CT noise on SPECT noise was demonstrated to be negligible for clinically achievable CT parameters. Because CT dose levels that affect SPECT quantification is low (CTDIvol ∼ 4 μGy), the low dose limit for the CT exam as part of SPECT/CT will be guided by CT image quality requirements for anatomical localization and artifact reduction. A CT technique with higher kVp in combination with lower mAs is recommended when low-dose CT images are used for AC to minimize beam-hardening artifacts.
© 2014 American Association of Physicists in Medicine.

Mesh:

Substances:

Year:  2014        PMID: 24694155     DOI: 10.1118/1.4868511

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Practical reconstruction protocol for quantitative (90)Y bremsstrahlung SPECT/CT.

Authors:  W Siman; J K Mikell; S C Kappadath
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

2.  Radionuclide bone scan SPECT-CT: lowering the dose of CT significantly reduces radiation dose without impacting CT image quality.

Authors:  Sandeep Kumar Gupta; Scott Trethewey; Bree Brooker; Natalie Rutherford; Jenny Diffey; Suresh Viswanathan; John Attia
Journal:  Am J Nucl Med Mol Imaging       Date:  2017-04-15

3.  Robust, fully automatic delineation of the head contour by stereotactical normalization for attenuation correction according to Chang in dopamine transporter scintigraphy.

Authors:  Catharina Lange; Jens Kurth; Anita Seese; Sarah Schwarzenböck; Karen Steinhoff; Bert Umland-Seidler; Bernd J Krause; Winfried Brenner; Osama Sabri; Swen Hesse; Ralph Buchert
Journal:  Eur Radiol       Date:  2015-02-26       Impact factor: 5.315

4.  Evaluation of quantitative accuracy among different scatter corrections for quantitative bone SPECT/CT imaging.

Authors:  Kenta Miwa; Reo Nemoto; Hirotsugu Masuko; Tensho Yamao; Rinya Kobayashi; Noriaki Miyaji; Kosuke Inoue; Hiroya Onodera
Journal:  PLoS One       Date:  2022-06-06       Impact factor: 3.752

5.  Quantitative imaging of bone remodeling in patients with a unicompartmental joint unloading knee implant (ATLAS Knee System)-effect of metal artifacts on a SPECT-CT-based quantification.

Authors:  Oliver S Grosser; Marcus Klutzny; Heiko Wissel; Dennis Kupitz; Michael Finger; Simone Schenke; Jan Wuestemann; Christoph H Lohmann; Christoph Hoeschen; Maciej Pech; Christian Staerke; Michael C Kreissl
Journal:  EJNMMI Phys       Date:  2021-02-17

Review 6.  Best practice for the nuclear medicine technologist in CT-based attenuation correction and calcium score for nuclear cardiology.

Authors:  Luca Camoni; Andrea Santos; Marieclaire Attard; Marius Ovidiu Mada; Agata Karolina Pietrzak; Sonja Rac; Sebastijan Rep; Christelle Terwinghe; Pedro Fragoso Costa
Journal:  Eur J Hybrid Imaging       Date:  2020-07-06

7.  On the optimization of bone SPECT/CT in terms of image quality and radiation dose.

Authors:  Monika Tulik; Piotr Tulik; Teresa Kowalska
Journal:  J Appl Clin Med Phys       Date:  2020-10-27       Impact factor: 2.243

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.