Impact of coronary stent designs on acute stent recoil.

Tomoyuki Ota1, Hideki Ishii2, Takuya Sumi3, Takuya Okada3, Hisashi Murakami3, Susumu Suzuki4, Kenji Kada3, Naoya Tsuboi3, Toyoaki Murohara4.   

Abstract

BACKGROUND: Acute stent recoil has been often observed following stent delivery balloon deflation in coronary arteries and the recoil rate varies by stent design. Accordingly, the purpose of the present study was to evaluate the impact of stent designs on acute stent recoil after new generation drug-eluting stent implantation. METHODS AND
RESULTS: A total of 154 lesions [56 treated with biolimus-eluting stent (BES), 46 with cobalt chromium everolimus-eluting stent (CoCr-EES), and 52 with platinum chromium everolimus-eluting stent (PtCr-EES)] were evaluated. Quantitative coronary angiography was used to measure the minimal lumen diameter (MLD). MLD1 was defined as a MLD of complete expansion of the last stent delivery balloon at the highest pressure. MLD2 was defined as a MLD immediately after the last stent delivery balloon deflation. Acute stent recoil was determined by the calculation as (MLD1-MLD2)/MLD1. Acute stent recoil was significantly higher in the CoCr-EES group versus the BES group and PtCr-EES group (10.1 ± 6.9%, 6.7 ± 5.5%, and 6.5 ± 4.8%, respectively, p = 0.01). Multivariate linear regression analysis demonstrated that the use of CoCr-EES and the number of stent delivery balloon inflations were independent predictors of acute stent recoil (r = 0.26, β = 0.21, p = 0.01 and r = -0.51, β = -0.58, p < 0.01, respectively).
CONCLUSION: Acute stent recoil occurred more frequently with the CoCr-EES compared with both BES and PtCr-EES. Strategies with multiple balloon inflation might be needed to overcome this recoil phenomenon.
Copyright © 2014. Published by Elsevier Ltd.

Entities:  

Keywords:  Drug-eluting stent; Percutaneous coronary intervention; Stent recoil

Mesh:

Substances:

Year:  2014        PMID: 24679906     DOI: 10.1016/j.jjcc.2014.02.013

Source DB:  PubMed          Journal:  J Cardiol        ISSN: 0914-5087            Impact factor:   3.159


  4 in total

1.  Predictors of acute scaffold recoil after implantation of the everolimus-eluting bioresorbable scaffold: an optical coherence tomography assessment in native coronary arteries.

Authors:  Takao Sato; John Jose; Mohamed El-Mawardy; Dmitriy S Sulimov; Ralph Tölg; Gert Richardt; Mohamed Abdel-Wahab
Journal:  Int J Cardiovasc Imaging       Date:  2016-10-19       Impact factor: 2.357

Review 2.  Methods to assess bioresorbable vascular scaffold devices behaviour after implantation.

Authors:  Alberto Pernigotti; Elisabetta Moscarella; Giosafat Spitaleri; Claudia Scardino; Kohki Ishida; Salvatore Brugaletta
Journal:  J Thorac Dis       Date:  2017-08       Impact factor: 2.895

3.  Multi-objective optimisation of material properties and strut geometry for poly(L-lactic acid) coronary stents using response surface methodology.

Authors:  Ross W Blair; Nicholas J Dunne; Alex B Lennon; Gary H Menary
Journal:  PLoS One       Date:  2019-08-26       Impact factor: 3.240

4.  Elastic stent recoil in coronary total occlusions: Comparison of durable-polymer zotarolimus eluting stent and ultrathin strut bioabsorbable-polymer sirolimus eluting stent.

Authors:  Riccardo Improta; Paola Scarparo; Jeroen Wilschut; Quinten Wolff; Joost Daemen; Wijnand K Den Dekker; Felix Zijlstra; Nicolas M Van Mieghem; Roberto Diletti
Journal:  Catheter Cardiovasc Interv       Date:  2021-05-07       Impact factor: 2.585

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.