Miniature forward-imaging B-scan optical coherence tomography probe to guide real-time laser ablation.

Zhuoyan Li1, Jin H Shen, John A Kozub, Ratna Prasad, Pengcheng Lu, Karen M Joos.   

Abstract

BACKGROUND AND
OBJECTIVE: Investigations have shown that pulsed lasers tuned to 6.1 µm in wavelength are capable of ablating ocular and neural tissue with minimal collateral damage. This study investigated whether a miniature B-scan forward-imaging optical coherence tomography (OCT) probe can be combined with the laser to provide real-time visual feedback during laser incisions. STUDY DESIGN/METHODS AND MATERIALS: A miniature 25-gauge B-scan forward-imaging OCT probe was developed and combined with a 250 µm hollow-glass waveguide to permit delivery of 6.1 µm laser energy. A gelatin mixture and both porcine corneal and retinal tissues were simultaneously imaged and lased (6.1 µm, 10 Hz, 0.4-0.7 mJ) through air. The ablation studies were observed and recorded in real time. The crater dimensions were measured using OCT imaging software (Bioptigen, Durham, NC). Histological analysis was performed on the ocular tissues.
RESULTS: The combined miniature forward-imaging OCT and mid-infrared laser-delivery probe successfully imaged real-time tissue ablation in gelatin, corneal tissue, and retinal tissue. Application of a constant number of 60 pulses at 0.5 mJ/pulse to the gelatin resulted in a mean crater depth of 123 ± 15 µm. For the corneal tissue, there was a significant correlation between the number of pulses used and depth of the lased hole (Pearson correlation coefficient = 0.82; P = 0.0002). Histological analysis of the cornea and retina tissues showed discrete holes with minimal thermal damage.
CONCLUSIONS: A combined miniature OCT and laser-delivery probe can monitor real-time tissue laser ablation. With additional testing and improvements, this novel instrument has the future possibility of effectively guiding surgeries by simultaneously imaging and ablating tissue.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  ablation; cornea; eye; gelatin; mid-infrared; ophthalmology; retina; waveguide

Mesh:

Year:  2014        PMID: 24648326      PMCID: PMC3962956          DOI: 10.1002/lsm.22214

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  43 in total

1.  Guidance of aortic ablation using optical coherence tomography.

Authors:  Nirlep A Patel; Xingde Li; Debra L Stamper; James G Fujimoto; Mark E Brezinski
Journal:  Int J Cardiovasc Imaging       Date:  2003-04       Impact factor: 2.357

2.  Toward guidance of epicardial cardiac radiofrequency ablation therapy using optical coherence tomography.

Authors:  Christine P Fleming; Kara J Quan; Andrew M Rollins
Journal:  J Biomed Opt       Date:  2010 Jul-Aug       Impact factor: 3.170

3.  Optical coherence tomography for the diagnosis of cervical intraepithelial neoplasia.

Authors:  Julia K S Gallwas; Lisa Turk; Herbert Stepp; Susanna Mueller; Robert Ochsenkuehn; Klaus Friese; Christian Dannecker
Journal:  Lasers Surg Med       Date:  2011-03       Impact factor: 4.025

4.  Cortical bone healing following laser osteotomy using 6.1 microm wavelength.

Authors:  J T Payne; G M Peavy; L Reinisch; D C Van Sickle
Journal:  Lasers Surg Med       Date:  2001       Impact factor: 4.025

5.  Vitreoretinal surgery of the posterior segment for explosive trauma in terrorist warfare.

Authors:  Boris Bajaire; Elena Oudovitchenko; Edgar Morales
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-01-27       Impact factor: 3.117

6.  Perforating globe injuries during operation Iraqi Freedom.

Authors:  Marcus H Colyer; Dal W Chun; Kraig S Bower; John S B Dick; Eric D Weichel
Journal:  Ophthalmology       Date:  2008-07-31       Impact factor: 12.079

7.  Structural markers observed with endoscopic 3-dimensional optical coherence tomography correlating with Barrett's esophagus radiofrequency ablation treatment response (with videos).

Authors:  Tsung-Han Tsai; Chao Zhou; Yuankai K Tao; Hsiang-Chieh Lee; Osman O Ahsen; Marisa Figueiredo; Tejas Kirtane; Desmond C Adler; Joseph M Schmitt; Qin Huang; James G Fujimoto; Hiroshi Mashimo
Journal:  Gastrointest Endosc       Date:  2012-07-24       Impact factor: 9.427

8.  Near-infrared image-guided laser ablation of artificial caries lesions.

Authors:  You-Chen Tao; Kenneth Fan; Daniel Fried
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2007-03-08

9.  Role of optical coherence tomography on corneal surface laser ablation.

Authors:  Bruna V Ventura; Haroldo V Moraes; Newton Kara-Junior; Marcony R Santhiago
Journal:  J Ophthalmol       Date:  2012-09-05       Impact factor: 1.909

10.  Raman-shifted alexandrite laser for soft tissue ablation in the 6- to 7-µm wavelength range.

Authors:  John Kozub; Borislav Ivanov; Aroshan Jayasinghe; Ratna Prasad; Jin Shen; Marc Klosner; Donald Heller; Marcus Mendenhall; David W Piston; Karen Joos; M Shane Hutson
Journal:  Biomed Opt Express       Date:  2011-04-19       Impact factor: 3.732

View more
  10 in total

1.  Efficacy and predictability of soft tissue ablation using a prototype Raman-shifted alexandrite laser.

Authors:  John A Kozub; Jin-H Shen; Karen M Joos; Ratna Prasad; M Shane Hutson
Journal:  J Biomed Opt       Date:  2015-10       Impact factor: 3.170

2.  Motorized Micro-Forceps with Active Motion Guidance based on Common-Path SSOCT for Epiretinal Membranectomy.

Authors:  Gyeong Woo Cheon; Berk Gonenc; Russell H Taylor; Peter L Gehlbach; Jin U Kang
Journal:  IEEE ASME Trans Mechatron       Date:  2017-09-05       Impact factor: 5.303

3.  Evaluation of microsurgical tasks with OCT-guided and/or robot-assisted ophthalmic forceps.

Authors:  Haoran Yu; Jin-Hui Shen; Rohan J Shah; Nabil Simaan; Karen M Joos
Journal:  Biomed Opt Express       Date:  2015-01-09       Impact factor: 3.732

4.  Review of intraoperative optical coherence tomography: technology and applications [Invited].

Authors:  Oscar M Carrasco-Zevallos; Christian Viehland; Brenton Keller; Mark Draelos; Anthony N Kuo; Cynthia A Toth; Joseph A Izatt
Journal:  Biomed Opt Express       Date:  2017-02-21       Impact factor: 3.732

5.  Development and preliminary results of bimanual smart micro-surgical system using a ball-lens coupled OCT distance sensor.

Authors:  Dongwoo Koo; Hyun-Cheol Park; Peter L Gehlbach; Cheol Song
Journal:  Biomed Opt Express       Date:  2016-10-31       Impact factor: 3.732

6.  Correlation of spectral domain optical coherence tomography with histology and electron microscopy in the porcine retina.

Authors:  Wankun Xie; Min Zhao; Shu-Huai Tsai; William L Burkes; Luke B Potts; Wenjuan Xu; H Ross Payne; Travis W Hein; Lih Kuo; Robert H Rosa
Journal:  Exp Eye Res       Date:  2018-08-16       Impact factor: 3.467

7.  Optic nerve sheath fenestration using a Raman-shifted alexandrite laser.

Authors:  John Kozub; Jin H Shen; Karen M Joos; Ratna Prasad; M Shane Hutson
Journal:  Lasers Surg Med       Date:  2015-12-14       Impact factor: 4.025

8.  A novel integration of spectral-domain optical-coherence-tomography and laser-ablation system for precision treatment.

Authors:  Yingwei Fan; Boyu Zhang; Wei Chang; Xinran Zhang; Hongen Liao
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-09-09       Impact factor: 2.924

Review 9.  A Review of Robotic and OCT-Aided Systems for Vitreoretinal Surgery.

Authors:  Elan Z Ahronovich; Nabil Simaan; Karen M Joos
Journal:  Adv Ther       Date:  2021-04-03       Impact factor: 3.845

10.  Laser brain cancer surgery in a xenograft model guided by optical coherence tomography.

Authors:  Nitesh Katta; Arnold D Estrada; Austin B McElroy; Aleksandra Gruslova; Meagan Oglesby; Andrew G Cabe; Marc D Feldman; Ry Declan Fleming; Andrew J Brenner; Thomas E Milner
Journal:  Theranostics       Date:  2019-05-26       Impact factor: 11.556

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.