Literature DB >> 24613513

A dosimetric comparison of IMRT versus helical tomotherapy for brain tumors.

M Skórska1, T Piotrowski2, J Kaźmierska3, K Adamska4.   

Abstract

BACKGROUND AND
PURPOSE: Helical tomotherapy (HT) can deliver highly conformal, uniform doses to the target volume. However, HT can only be delivered in a coplanar mode. The purpose of this study was to perform a dosimetric comparison of HT versus coplanar (cIMRT) and non-coplanar (n-cIMRT) beam arrangements on a conventional linear accelerator in a diverse group of brain tumors.
MATERIALS AND METHODS: A total of 45 treatment plans were calculated retrospectively for 15 cases. For each case, 3 different delivery techniques (n-cIMRT, cIMRT and HT) were used. The treatment plans were compared using the parameters of the target coverage (conformity index; CI) and homogeneity (HI) for the planning target volume (PTV) and the maximum and mean doses for organs at risk (OARs).
RESULTS: Median HI and CI were the best for HT plans and the worst for cIMRT. The largest reduction of maximum dose for lenses and mean dose for both eyes was achieved for n-cIMRT plans. Mean dose for chiasm and the ipsilateral optic nerve were the lowest for HT. The contralateral optic nerve was most spared with n-cIMRT. For D1% in the brain stem, there was no significant difference between HT and the IMRT plans.
CONCLUSIONS: Both HT and n-cIMRT are capable of producing conformal and homogeneous treatment plans with a good sparing of OARs. However, due to the non-coplanar capabilities of IMRT, n-cIMRT led to a superior dose reduction to the lenses.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Brain tumor; Plan comparison; Radiotherapy; Tomotherapy; Treatment planning

Mesh:

Year:  2014        PMID: 24613513     DOI: 10.1016/j.ejmp.2014.02.002

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  6 in total

1.  Relationships between various indices of doses distribution homogeneity.

Authors:  Marzena Mrozowska; Paweł Kukołowicz
Journal:  Rep Pract Oncol Radiother       Date:  2015-05-01

2.  Dosimetric comparison of conformal technique (3D) with volumetric modulated arc therapy with respect to doses obtained in the temporal lobe area in patients irradiated for brain meningioma.

Authors:  Łukasz Trembecki; Tomasz Siudziński; Aleksandra Sztuder; Adam Maciejczyk
Journal:  Rep Pract Oncol Radiother       Date:  2019-05-31

3.  Performance of the eclipse monitor unit objective tool utilizing volumetric modulated arc therapy for rectal cancer.

Authors:  Alejandro Prado; Ángel Gaitán; Mario Leonor; Marta Manzano; Eduardo Cabello; Raúl Díaz; Alejandro Ferrando; Ana Milanés; Gustavo Pozo
Journal:  Rep Pract Oncol Radiother       Date:  2019-02-25

4.  Intensity-Modulated Radiotherapy (IMRT) vs Helical Tomotherapy (HT) in Concurrent Chemoradiotherapy (CRT) for Patients with Anal Canal Carcinoma (ACC): an analysis of dose distribution and toxicities.

Authors:  Rosanna Yeung; Yarrow McConnell; Heather Warkentin; Darren Graham; Brad Warkentin; Kurian Joseph; Corinne M Doll
Journal:  Radiat Oncol       Date:  2015-04-17       Impact factor: 3.481

5.  Helical tomotherapy: Comparison of Hi-ART and Radixact clinical patient treatments at the Technical University of Munich.

Authors:  K M Kraus; S Kampfer; J J Wilkens; L Schüttrumpf; S E Combs
Journal:  Sci Rep       Date:  2020-03-18       Impact factor: 4.379

6.  Dosimetric comparisons of craniospinal axis irradiation using helical tomotherapy, volume-modulated arc therapy and intensity-modulated radiotherapy for medulloblastoma.

Authors:  Yangqing Sun; Gui Liu; Wen Chen; Taili Chen; Pei Liu; Qian Zeng; Jidong Hong; Rui Wei
Journal:  Transl Cancer Res       Date:  2019-02       Impact factor: 1.241

  6 in total