Literature DB >> 24599315

Evaluation of Abbott RealTime CT/NG assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in cervical swabs from female sex workers in China.

Yan Han1, Yue-ping Yin1, Mei-qin Shi1, Bing-jie Zheng1, Ming-ying Zhong1, Ning Jiang1, Shao-chun Chen1, Xiang-sheng Chen1.   

Abstract

BACKGROUND: To evaluate the performance of the Abbott RealTime CT/NG assay for detection of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) among female sex workers (FSWs) in China.
METHODS: Cervical swabs from 997 participants were blindly detected by the Abbott RealTime CT/NG assay on the automated m2000 molecular platform and Roche Cobas Amplicor CT/NG assay. Discrepant analysis were confirmed by the Qiagen care CT PCR assay. The sample was defined as candidate nvCT-positive if it was CT positive in the Abbott m2000 assay, but CT negative in the other two assays.
RESULTS: 25 specimens that were discordant for CT and 26 specimens that were discordant for NG between the two assays were resolved by Qiagen care CT & NG PCR assays. The sensitivity and specificity, respectively, for Abbott m2000 assay were 92.59% and 100% for CT and 95.45% and 99.90% for NG. The positive predictive value (PPV) and negative predictive value (NPV) of Abbott m2000 assay were100% and 98.52% for CT and 95.5% and 99.90% for NG, respectively. No candidate new-variant CT(nvCT)specimens were identified.
CONCLUSION: Abbott RealTime CT/NG assay were more specify for CT and NG detection, however, its sensitivity for CT and NG were a little bit lower than Roche Cobas Amplicor CT/NG assay. Abbott RealTime CT/NG assay had higher PPV for NG detection than Roche Cobas Amplicor CT/NG assay; it would be more suitable for screening for population with low-prevalence NG. There is currently no evidence that nvCT is present in FSWs in China.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24599315      PMCID: PMC3943787          DOI: 10.1371/journal.pone.0089658

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) are among the most common causes of sexually transmitted infections [1]. Mounting evidence suggests co-infection of CT and NG has been become an urgent phenomenon in many countries. The number of new cases in adults of the CT and NG in 2008 was estimated to be 105.7 million and 106.1 million by the World Health Organization (WHO) [2]. Since 1999 our laboratory has conducted studies to estimate the prevalence of CT and NG infections among high risk groups including FSWs in China using the highly sensitive and specific NAATs-Roche Amplicor PCR assay [3]. This assay is well-established, FDA-cleared method for detection of CT and NG [4]. Despite its advantages, this kit has two major limitations. First, the primers used in its NG analysis, which target the putative cytosine DNA methyltransferase gene, cross react with strains of several commensal Neisseria species. Second, its CT assay, which targets cryptic plasmid, is unable to detect the new variant of CT (nvCT). And this test would be no longer available, and has been replaced with the next generation PCR (Roche Cobas 4800) that detects the CT variant, and has no cross-reactions with other Neisseria. But this Cobas 4800 were not available in China market that time. The FDA-cleared new Abbott RealTime CT/NG assay consists of two different targets for CT detection, targeting both the cryptic plasmid and the chromosomal omp1 gene. Its NG analysis detects a region in the NG opacity (Opa) gene. The Abbott RealTime CT/NG assay, which recently became available in China, has been reported to have high sensitivity and specificity, with the advantage of simultaneously detecting both NG and CT (including wild type CT and nvCT) in cervical swab specimens [4]. The nvCT with a 377-base pair (bp) deletion in the cryptic plasmid has been reported in Sweden in 2006 [5].Although the nvCT has been detected in Norway, Finland and Denmark, only a few cases of nvCT have been reported outside the eastern countries [6]. No screening for nvCT have been performed previously in China, despite the fact that the ever-increasing of cross-border travel between China and Sweden. The report incidence of NG for man and women in American were much higher than that in China in 2011 (98.7 and 108.9 per 100,000 people vs. 17.13 and 14.50 per 100,000 people) [7], [8]. This big difference in report incidence of NG between the United States and China is most likely due to different screening methods, the United States expanded use of more sensitive tests such as nucleic acid amplification testing methods (NAATs) [7] while China routinely based on the less sensitive test such as the Clearview test [9]. As China's National STD Reference Laboratory, we are obligated to recommend the more effective assay as reference test for CT and NG detection. The present study aimed to compare the performances of Abbott RealTime CT/NG assay for the duplex testing of both pathogens from cervical swabs among female sex workers (FSWs) in China. Another goal was to estimate the prevalence of nvCT among this high-risk population.

Materials and Methods

Study population

FSWs were recruited from varying work venues in two provinces of southern China (Guangdong and Hainan provinces) from June to September, 2009. A convenience sample of FSWs from each selected venue participated in a questionnaire survey and cervical specimen collection. Females who were older than 18 years and reported providing commercial sex in the past year were eligible for this study. ALL participants provide their written informed consent to participate in this study. The study protocol was reviewed and approved by the ethics committee of Institute of dermatology, Chinese academy of medical sciences and Peking union medical college, Nanjing, China.

Laboratory test

Two cervical swab specimens were collected by trained nurses from each of the 997 eligible participants. Cervical swab samples were shipped to the National Reference Laboratory of the National Center for STD Control in Nanjing and stored at −70°C until the following analyses. Collection of cervical swabs and preparation of swab eluent were performed according to methods described previously [10]. Of the two swab eluent samples from each participant, one was subjected to the Roche Amplicor CT/NG PCR assay for detection of CT and NG according to the manufacturer's instructions in our laboratory; and the other one was sent to Chinese Abbott consignee's commission laboratory in Beijing, in which it was blind tested with the Abbott RealTime CT/NG assay that uses the automated m2000 molecular platform. In Roche Amplicor CT/NG PCR system, specimens yielding signals below the negative cutoff for both CT/NG and the internal control (IC) (optical density (OD) <0.2) were interpreted as inhibitory. An inhibitory specimen was retested by processing another aliquot of the original specimen. The Abbott RealTime CT/NG assay is a qualitative assay. Samples with a cycle number less than or equal to the assay cutoff (CO) were interpreted as “positive.” Samples with a cycle number beyond the assay CO were interpreted as “equivocal.” Samples with no evidence of amplification were interpreted as “negative.” Samples with an initial interpretation of “equivocal” for C. trachomatis were retested. All samples were tested once initially by both test systems as described above. The samples yielding discrepant results in the two assays were further tested by another technician who had no idea of the primary results using the Qiagen care CT&NG PCR assay. This assay was approved by the State Food and Drug Administration commercially available in China market. Reference standards were set by combining the results of the Roche and Qiagen assays. Samples with an indeterminate CT or NG infection status were considered discrepant results and included in the final assay for evaluation of its sensitivity and specificity. The Abbott RealTime CT/NG assay was able to detect the nv-CT, while the Roche Amplicor assay and Qiagen care CT assay were unable to detect the nv-CT. The sample was defined as candidate nvCT-positive if it was CT positive in the Abbott m2000 assay, but CT negative in the other two assays. The candidate nvCT-positive will be confirmed by full gene sequence.

Statistical analysis

Repeatedly inhibitory specimens were excluded from the terminal calculations. All data analyses were performed by IBM SPSS Statistic 20 (SPSS Inc, Chicago, Ill, USA). The sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the Abbott assay were calculated using standard methods. The McNemar test was used to compare the results of the Abbott RealTime CT/NG assay with Roche Cobas Amplicor CT/NG assay.

Results

A total of 997 female cervix swabs were blindly evaluated for CT/NG using the Abbott RealTime CT/NG assay and Roche Cobas Amplicor CT/NG assay. After inhibited samples were retested using the two assays with the same batch of reagents as that applied initially, 0.8% (8/997) of these specimens gave repetitive inhibitory test results for CT and NG with the Roche Cobas Amplicor CT/NG assay, so these eight specimens were excluded from total CT/NG statistical analysis. Three specimens gave repetitive inhibitory test results for CT with the Abbott RealTime CT/NG assay, so these three specimens were excluded from CT statistical analysis. Table 1 shows results for 997 cervix swabs tested with both systems for the presence of CT and NG. Using the Roche assay system, 200 specimens were positive for CT (20.2% prevalence) and 46 were positive for NG (4.7% prevalence). Using the Abbott assay, 175 specimens were positive for CT (17.7% prevalence) and 22 were also positive for NG (2.2% prevalence). A total of 13 specimens were positive for both CT and NG using Roche Cobas Amplicor CT/NG assay, 10 were positive for both CT and NG using Abbott RealTime CT/NG assay. Through Roche Cobas Amplicor CT/NG assay, about 28.3% (13.46) NG positive specimens were co-infected with CT, and 6.5% (13/200) CT positive specimens were co-infected with NG. Meantime, when Abbott ReatTime CT/NG assay was used, these rates were 45.5% (10/22) and 5.7% (10/175) respectively.
Table 1

Comparison of test results for cervical swabs between Abbott m2000 CT/NG system and Roche Amplicor CT/NG PCR systems.

Abbott CTAbbott NG
Roche CTNegativePositiveIndeterminateTotalRoche NGNegativePositiveIndeterminateTotal
Negative78600786Negative94210943
Positive251753203Positive2521046
Indeterminate8008Indeterminate8008
Total8191753997Total975220997
We next calculated concordance of CT between the two assays. There were concordant CT-positive results for 175 specimens and concordant negative results for 786 specimens. 25 specimens were discordant for CT, as they were tested positive with Roche Cobas Amplicor CT/NG assay but negative with the Abbott RealTime CT/NG assay. The positive and negative agreement for CT were 87.5% (95%CI 81.92%, 91.60%) and 100% (95%CI 99.39%, 100%) between the Abbott and Roche assays. The P values from McNemar test for CT was lower than 0.01 between the two systems. The Kappa coefficient for CT was 0.92. There were concordant NG-positive results for 21 specimens and concordant negative results for 942 specimens. 26 specimens were discordant for NG between the 2 systems: 25 specimens were positive for NG with the Roche Cobas Amplicor CT/NG assay but negative with the Abbott RealTime CT/NG assay; one specimen was negative for NG with Roche Cobas Amplicor CT/NG assay but positive for NG with the Abbott RealTime CT/NG assay. The positive and negative agreement for NG were 45.65% (31.18%, 60.84%) and 99.89% (99.31%, 100%) between the two assays systems. The P values from McNemar test for NG was lower than 0.01. The value of Kappa coefficient for NG was 0.61. Discrepant analysis of 25 CT Roche positive/Abbott negative specimens; confirmed 12 as true positives by the Qiagen care CT PCR assay. One specimen was confirmed as true positive by the Qiagen care GC PCR assay in discrepant analysis of 25 NG Roche positive/Abbott negative specimens. Discrepant analysis of 1 NG Roche negative/Abbott positive specimens; confirmed 1 as a true negative by the Qiagen care GC PCR assay (Table 2). The overall true positive prevalence of CT and that of NG were 18.97% (187/986, 95% CI, 16.59–21.58%) and 2.22% (22/989, 95% CI, 1.43–3.41%), respectively. The co-infection prevalence of NG and CT were 1.01 (10/986, 95% CI, 0.52–1.92). Three co-infections of NG and CT were missed by the Abbott system and Qiagen system.
Table 2

The results of Qiagen care CT&NG PCR assay for discrepant results in Roche and Abbott assays.

Qiagen CTQiagen NG
NegativePositiveTotalNegativePositiveTotal
Roche Pos/Abbott Neg13122524125
Roche Neg/Abbott Pos000101
Total13122525126
The performance of Abbott m2000 CT/NG system compared to the reference results (Table 3), the sensitivity and specificity of the Abbott m2000 assay were 92.59% (175/187, 95% CI, 87.63–95.74%) and 100% (799/799, 95% CI, 99.40–100%) for CT and 95.45% (21/22, 95% CI, 75.12–99.76%) and 99.90% (966/967, 95% CI, 99.33–99.99%) for NG, respectively. Its PPV and NPV reached 100% (175/175, 95% CI, 97.32–100%) and 98.52% (799/811, 95% CI, 97.35–99.20%) for CT and 95.5% (21/22, 95% CI, 75.12–99.76%) and 99.90% (966/967, 95% CI, 99.33–99.99%) for NG, respectively. In comparison, The Roche Cobas Amplicor CT/NG assay showed a sensitivity and specificity of 100% (187/187, 95% CI, 97.49–100%) and 98.37% (786/799, 95% CI, 97.16–99.09%) for CT and 100% (22/22, 95% CI, 81.50–100%) and 97.52% (943/967, 95% CI, 96.27–98.37%) for NG, respectively. The PPV and NPV of Roche Cobas Amplicor CT/NG assay were 100% (786/786, 95% CI, 99.39–100%) and 93.50% (187/200, 88.90–96.35%), respectively; for CT and 52.17% (95% CI, 37.12–66.86%) and 100% (943/943, 95% CI, 99.49–100%) for NG.
Table 3

The performance of Abbott m2000 CT/NG system compared to the reference results.

Abbott CTAbbott NG
NegativePositiveTotalNegativePositiveTotal
Reference Negative79907999661967
Reference Positive1217518712122
Total81117598696722989
No candidate nvCT was found among the FSW population in this study.

Discussion

This is the first independent evaluation study for Abbott RealTime CT/NG assay in China without the support of manufacturers. In the present research, 20.2% and 17.7% specimens were detected CT DNA, 4.7% and 2.2% were detected NG DNA by Roche Cobas Amplicor CT/NG assay and Abbott RealTime CT/NG assay respectively. Given the difference between those assays might be exist, both of them still showed a higher prevalence of CT/NG among the studied population compared with general population [11]–[14]. In addition, a high rate of CT-NG co-infection among NG infected participants partly support the suggestion of the STD Treatment Guideline [15], in which simultaneous treatment to CT among NG-infected patients should be recommended. The ratio of positive agreement (87.5%) and negative agreement (100%) for CT in this study were comparable to the manufacturer's instruction of the Abbott RealTime CT/NG assay (91.3%, (95% CI, 85.03–95.60%)), (98.9%, (95% CI, 97.38–99.55%).The negative agreement ratio for NG (99.89%) from our study was similar to the manufacturer's instruction of the Abbott RealTime CT/NG assay too. However, the positive agreement ratio for NG was only 45.65% between the Abbott and Roche methods in the population with low prevalence of NG (2.2%), which was significantly lower than the manufacturer's instruction (97.8%, (95%CI, 92.45–99.74%)), and the evaluation study which was supported in part by grant support received from by Abbott Laboratories (96.6% (95%CI, 88.3–99.1%)) among the population with high prevalence of NG (8.2%) [4]. Three co-infections of NG and CT were missed by the Abbott system and Qiagen system. This was probably because the nucleic acid concentration of one target greatly exceeds the other, so PCR reaction failed to amplify the less concentrated target [16]. Alternatively, these three specimens may be false positive. Unfortunately, we could not distinguish between these two possibilities because there was insufficient amount of sample for further analysis. Abbott RealTime CT/NG assay were more specify for CT and equivalent for NG detection than Roche Cobas Amplicor CT/NG assay. However, Abbott RealTime CT/NG assay sensitivity for CT and NG were some lower than Roche Cobas Amplicor CT/NG assay. Abbott RealTime CT/NG assay had higher PPV for NG detection than Roche Cobas Amplicor CT/NG assay; it would be more suitable for screening for the population with low-prevalence NG. The PPV of Roche Cobas Amplicor CT/NG assay for NG in this study was similar to the study conducted by Diemert 31.6% (13.6–56.5%) in low prevalence (1.19%, 6/503) population [17]. The low PPV in NG of Roche Cobas Amplicor CT/NG assay may be caused by sequence-related issues with the Roche assay. Fortunately, this company has developed the new generation product (Roche 4800) to replace this old assay, the Roche 4800 having a much better performance profile [18], [19]. The nvCT with a 377 bp deletion plasmid has been described in Sweden in 2006. It can cause diseases and has the same biological fitness as wild type CT. It may be able to spread from one country to another in an unnoticed way [20]. However, a range of CT NAAT platforms have been established and utilized in China, which makes the rapid transmission of nvCT unlikely. However, when using the CT NAAT platforms whose primary method for testing CT has no capacity to detect nvCT, epidemiological laboratories need to remain vigilant for obvious decreases in the number of positive CT cases. Some limitations of this study should be acknowledged. First, the study subjects recruited by convenience sampling were not truly representative of the sex worker population. Second, failure to use Abbott or Roche Standard operation procedures (SOPs) for collection of specimens and storage may affect the performance of these two assays. Third, lack of positive CT variant controls may affect the evaluation of Abbott RealTime CT/NG assay performance for detecting CT variant. Fourth, the Qiagen Care assay is a new assay that has not been widely evaluated and may have low sensitivity, especially for NG detection. In conclusion, the Abbott RealTime CT/NG assay was comparable for Roche Cobas Amplicor CT/NG assay for CT and NG detection, especially for NG detection in the population with low-prevalence NG. There is currently no evidence that nvCT is present in FSWs in China.
  17 in total

1.  Prevalence of Neisseria gonorrhoeae among persons 14 to 39 years of age, United States, 1999 to 2008.

Authors:  Elizabeth A Torrone; Robert E Johnson; Lin H Tian; John R Papp; S Deblina Datta; Hillard S Weinstock
Journal:  Sex Transm Dis       Date:  2013-03       Impact factor: 2.830

2.  Comparison between the LCx Probe system and the COBAS AMPLICOR system for detection of Chlamydia trachomatis and Neisseria gonorrhoeae infections in patients attending a clinic for treatment of sexually transmitted diseases in Amsterdam, The Netherlands.

Authors:  G J van Doornum; L M Schouls; A Pijl; I Cairo; M Buimer; S Bruisten
Journal:  J Clin Microbiol       Date:  2001-03       Impact factor: 5.948

3.  Evaluation of the Roche cobas® CT/NG test for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in male urine.

Authors:  Stephanie N Taylor; Oliver Liesenfeld; Rebecca A Lillis; Barbara A Body; Melinda Nye; James Williams; Carol Eisenhut; Edward W Hook; Barbara Van Der Pol
Journal:  Sex Transm Dis       Date:  2012-07       Impact factor: 2.830

Review 4.  The Swedish new variant of Chlamydia trachomatis.

Authors:  Magnus Unemo; Ian N Clarke
Journal:  Curr Opin Infect Dis       Date:  2011-02       Impact factor: 4.915

Review 5.  The epidemic of sexually transmitted infections in China: implications for control and future perspectives.

Authors:  Xiang-Sheng Chen; Rosanna W Peeling; Yue-Ping Yin; David C Mabey
Journal:  BMC Med       Date:  2011-10-06       Impact factor: 8.775

6.  Evaluation of the Abbott RealTime CT/NG assay in comparison to the Roche Cobas Amplicor CT/NG assay.

Authors:  Annie Cheng; Qinfang Qian; James E Kirby
Journal:  J Clin Microbiol       Date:  2011-02-16       Impact factor: 5.948

7.  Population-based study of asymptomatic infection with Chlamydia trachomatis among female and male students.

Authors:  Hirohisa Imai; H Nakao; H Shinohara; Y Fujii; H Tsukino; R Hamasuna; Y Osada; K Fukushima; M Inamori; T Ikenoue; T Katoh
Journal:  Int J STD AIDS       Date:  2010-05       Impact factor: 1.359

8.  Prevalence of Chlamydia trachomatis: results from the first national population-based survey in France.

Authors:  V Goulet; B de Barbeyrac; S Raherison; M Prudhomme; C Semaille; J Warszawski
Journal:  Sex Transm Infect       Date:  2010-08       Impact factor: 3.519

9.  Performance of cobas® 4800 and m2000 real-time™ assays for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in rectal and self-collected vaginal specimen.

Authors:  Tanja H Geelen; John W Rossen; Antoine M Beerens; Linda Poort; Servaas A Morré; Wilma S Ritmeester; Harry E van Kruchten; Masja M van de Pas; Paul H M Savelkoul
Journal:  Diagn Microbiol Infect Dis       Date:  2013-07-23       Impact factor: 2.803

10.  Prevalences of sexually transmitted infections in young adults and female sex workers in Peru: a national population-based survey.

Authors:  César P Cárcamo; Pablo E Campos; Patricia J García; James P Hughes; Geoff P Garnett; King K Holmes
Journal:  Lancet Infect Dis       Date:  2012-08-08       Impact factor: 25.071

View more
  3 in total

1.  Evaluation of the Predictive Value of Urine Leukocyte Esterase Test in Chlamydia trachomatis and Neisseria gonorrhoeae Infection Among Males Attending HIV/STI Clinics in Guangdong Province, China.

Authors:  Xueying Yu; Peizhen Zhao; Zhida Mai; Qingqing Xu; Wentao Chen; Zhiqiao Wu; Xiaojuan Luo; Zhizhou Wu; Xiaofeng Liu; Qian Wu; Heping Zheng; Yaohua Xue
Journal:  Front Med (Lausanne)       Date:  2022-03-21

2.  Precision of the Abbott RealTime Assay in the Detection of Ocular Chlamydia trachomatis in a Trachoma-Endemic Area of Ethiopia.

Authors:  Kieran S O'Brien; Ambahun Chernet; Jeanne Moncada; Julius Schachter; Paul M Emerson; Scott D Nash; Melsew Chanyalew; Zerihun Tadesse; Zhaoxia Zhou; Charles E McCulloch; Thomas M Lietman; Jeremy D Keenan
Journal:  Am J Trop Med Hyg       Date:  2020-05-07       Impact factor: 2.345

3.  Evaluation of a Hydrogel-Based Diagnostic Approach for the Point-of-Care Based Detection of Neisseria gonorrhoeae.

Authors:  Sumudu R Perera; Ali Taheri; Nurul H Khan; Rajinder P Parti; Stephanie Stefura; Pauline Skiba; Jason P Acker; Irene Martin; Anthony Kusalik; Jo-Anne R Dillon
Journal:  Antibiotics (Basel)       Date:  2018-08-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.