Characteristics of the m2000 automated sample preparation and multiplex real-time PCR system for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

R Marshall1, M Chernesky, D Jang, E W Hook, C P Cartwright, B Howell-Adams, S Ho, J Welk, J Lai-Zhang, J Brashear, B Diedrich, K Otis, E Webb, J Robinson, H Yu.   

Abstract

We evaluated a new real-time PCR-based prototype assay for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae developed by Abbott Molecular Inc. This assay is designed to be performed on an Abbott m2000 real-time instrument system, which consists of an m2000sp instrument for sample preparation and an m2000rt instrument for real-time PCR amplification and detection. The limit of detection of this prototype assay was determined to be 20 copies of target DNA for both C. trachomatis and N. gonorrhoeae, using serially diluted linearized plasmids. No cross-reactivity could be detected when 55 nongonococcal Neisseria isolates and 3 non-C. trachomatis Chlamydia isolates were tested at 1 million genome equivalents per reaction. Concordance with the Roche Amplicor, BDProbeTec ET, and Gen-Probe APTIMA Combo 2 tests was assessed using unlinked/deidentified surplus clinical specimens previously analyzed with these tests. For C. trachomatis, concordance for positive results ranged from 93.7% to 100%, while concordance for negative results ranged from 98.2% to 100%. For N. gonorrhoeae, concordance for positive and negative results ranged from 91.4% to 100% and 99.3% to 100%, respectively. A workflow analysis of the prototype assay was conducted to obtain information on throughput under laboratory conditions. At 48 samples/run, the time to first result for both C. trachomatis and N. gonorrhoeae was 4.5 h. A total of 135 patient specimens could be analyzed in 8.9 h, with 75 min of hands-on time. This study demonstrated the technical and clinical feasibility of the new Abbott real-time PCR C. trachomatis/N. gonorrhoeae assay.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17202273      PMCID: PMC1829145          DOI: 10.1128/JCM.01956-06

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  20 in total

1.  Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by strand displacement amplification and relevance of the amplification control for use with vaginal swab specimens.

Authors:  Lisa A Cosentino; Daniel V Landers; Sharon L Hillier
Journal:  J Clin Microbiol       Date:  2003-08       Impact factor: 5.948

2.  Evaluation of an automated liquid-handling system (Tecan Genesis RSP 100) in the Abbott LCx assay for Chlamydia trachomatis.

Authors:  K L Hanson; C P Cartwright
Journal:  J Clin Microbiol       Date:  2001-05       Impact factor: 5.948

3.  Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by enzyme immunoassay, culture, and three nucleic acid amplification tests.

Authors:  E Van Dyck; M Ieven; S Pattyn; L Van Damme; M Laga
Journal:  J Clin Microbiol       Date:  2001-05       Impact factor: 5.948

4.  Multicenter evaluation of the BDProbeTec ET System for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens, female endocervical swabs, and male urethral swabs.

Authors:  B Van Der Pol; D V Ferrero; L Buck-Barrington; E Hook; C Lenderman; T Quinn; C A Gaydos; J Lovchik; J Schachter; J Moncada; G Hall; M J Tuohy; R B Jones
Journal:  J Clin Microbiol       Date:  2001-03       Impact factor: 5.948

5.  High analytical sensitivity and low rates of inhibition may contribute to detection of Chlamydia trachomatis in significantly more women by the APTIMA Combo 2 assay.

Authors:  Max Chernesky; Dan Jang; Kathy Luinstra; Sylvia Chong; Marek Smieja; Wenjie Cai; Beth Hayhoe; Eder Portillo; Cindy Macritchie; Cheryl Main; Ruth Ewert
Journal:  J Clin Microbiol       Date:  2006-02       Impact factor: 5.948

6.  Multicenter evaluation of the AMPLICOR and automated COBAS AMPLICOR CT/NG tests for detection of Chlamydia trachomatis.

Authors:  B Van Der Pol; T C Quinn; C A Gaydos; K Crotchfelt; J Schachter; J Moncada; D Jungkind; D H Martin; B Turner; C Peyton; R B Jones
Journal:  J Clin Microbiol       Date:  2000-03       Impact factor: 5.948

7.  Performance of the APTIMA Combo 2 assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in female urine and endocervical swab specimens.

Authors:  C A Gaydos; T C Quinn; D Willis; A Weissfeld; E W Hook; D H Martin; D V Ferrero; J Schachter
Journal:  J Clin Microbiol       Date:  2003-01       Impact factor: 5.948

8.  Evaluation of the Digene Hybrid Capture II Assay with the Rapid Capture System for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

Authors:  B Van Der Pol; J A Williams; N J Smith; B E Batteiger; A P Cullen; H Erdman; T Edens; K Davis; H Salim-Hammad; V W Chou; L Scearce; J Blutman; W J Payne
Journal:  J Clin Microbiol       Date:  2002-10       Impact factor: 5.948

9.  Vaginal swabs are appropriate specimens for diagnosis of genital tract infection with Chlamydia trachomatis.

Authors:  Julius Schachter; William M McCormack; Max A Chernesky; David H Martin; Barbara Van Der Pol; Peter A Rice; Edward W Hook; Walter E Stamm; Thomas C Quinn; Joan M Chow
Journal:  J Clin Microbiol       Date:  2003-08       Impact factor: 5.948

10.  Specimen processing and concentration of Chlamydia trachomatis added can influence false-negative rates in the LCx assay but not in the APTIMA Combo 2 assay when testing for inhibitors.

Authors:  S Chong; D Jang; X Song; J Mahony; A Petrich; P Barriga; M Chernesky
Journal:  J Clin Microbiol       Date:  2003-02       Impact factor: 5.948

View more
  17 in total

1.  [Diagnosis and treatment of Chlamydia trachomatis infections].

Authors:  T Meyer
Journal:  Hautarzt       Date:  2012-01       Impact factor: 0.751

2.  Immiscible phase nucleic acid purification eliminates PCR inhibitors with a single pass of paramagnetic particles through a hydrophobic liquid.

Authors:  Kunal Sur; Sally M McFall; Emilie T Yeh; Sujit R Jangam; Mark A Hayden; Stephen D Stroupe; David M Kelso
Journal:  J Mol Diagn       Date:  2010-06-25       Impact factor: 5.568

3.  Evaluation of three automated nucleic acid amplification systems for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in first-void urine specimens.

Authors:  P N Levett; K Brandt; K Olenius; C Brown; K Montgomery; G B Horsman
Journal:  J Clin Microbiol       Date:  2008-04-09       Impact factor: 5.948

4.  Workflow and maintenance characteristics of five automated laboratory instruments for the diagnosis of sexually transmitted infections.

Authors:  Sam Ratnam; Dan Jang; Jodi Gilchrist; Marek Smieja; Andre Poirier; Todd Hatchette; Jean-Frederic Flandin; Max Chernesky
Journal:  J Clin Microbiol       Date:  2014-04-16       Impact factor: 5.948

5.  Sensitivity and Specificity of Empiric Treatment for Sexually Transmitted Infections in a Pediatric Emergency Department.

Authors:  Kristen Breslin; Lisa Tuchman; Katie L Hayes; Gia Badolato; Monika K Goyal
Journal:  J Pediatr       Date:  2017-06-17       Impact factor: 4.406

6.  Molecular detection and confirmation of Neisseria gonorrhoeae in urogenital and extragenital specimens using the Abbott CT/NG RealTime assay and an in-house assay targeting the porA pseudogene.

Authors:  A Walsh; F O Rourke; B Crowley
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2010-12-19       Impact factor: 3.267

Review 7.  Chlamydia trachomatis today: treatment, detection, immunogenetics and the need for a greater global understanding of chlamydial disease pathogenesis.

Authors:  D Dean
Journal:  Drugs Today (Barc)       Date:  2009-11       Impact factor: 2.245

8.  Taqman real-time PCR detects Avipoxvirus DNA in blood of Hawai'i 'amakihi (Hemignathus virens).

Authors:  Margaret E M Farias; Dennis A LaPointe; Carter T Atkinson; Christopher Czerwonka; Rajesh Shrestha; Susan I Jarvi
Journal:  PLoS One       Date:  2010-05-27       Impact factor: 3.240

9.  Rapid and specific detection of amantadine-resistant influenza A viruses with a Ser31Asn mutation by the cycling probe method.

Authors:  Yasushi Suzuki; Reiko Saito; Hassan Zaraket; Clyde Dapat; Isolde Caperig-Dapat; Hiroshi Suzuki
Journal:  J Clin Microbiol       Date:  2009-11-04       Impact factor: 5.948

10.  Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays.

Authors:  Lesley E Scott; Lara D Noble; Jackie Moloi; Linda Erasmus; Willem D F Venter; Wendy Stevens
Journal:  J Clin Microbiol       Date:  2009-05-06       Impact factor: 5.948

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.