Signal to noise ratio comparisons for ultrasound attenuation slope estimation algorithms.

Eenas A Omari1, Tomy Varghese1.   

Abstract

PURPOSE: Attenuation imaging has a promising role in the detection of tissue abnormalities. The authors have previously compared three different frequency domain ultrasound attenuation estimation methods, for accuracy and bias. The mean estimated attenuation value in a region of interest has been the determining factor of how well a method performs; however, the noise level has not been quantified for attenuation estimated using different methods.
METHODS: The authors compare three different frequency domain ultrasound attenuation estimation methods [the reference phantom method (RPM), the centroid downshift method (CEN), and the hybrid method (HYB)] using the signal to noise ratio (SNR) metric. Both simulated and experimental tissue-mimicking phantoms are used in the performance comparison study, evaluating the impact of the variation in acoustical properties.
RESULTS: For attenuation estimation in a tissue-mimicking phantom with a known attenuation coefficient of 0.5 dB/cm/MHz, all the three methods estimated the attenuation coefficient to be ≈ 0.49 dB/cm/MHz for a transmit center frequency of 6 MHz, however, the signal to noise ratio obtained was found to be 8.5, 5.7, and 2.2 for the HYB, RPM, and CEN methods, respectively. These results demonstrate the need for the SNR metric in the comparison of different algorithms and to evaluate the impact of varying different ultrasound system and tissue parameters.
CONCLUSIONS: In this paper, the authors demonstrate that although the estimated mean attenuation value with a region of interest may be closely estimated using different methods, the signal to noise ratio obtained of the estimates can vary significantly. The centroid downshift method presented with the lowest signal-to-noise ratio of the methods compared. The hybrid method was the least susceptible to changes in the acoustical properties and provided unbiased attenuation coefficient estimates with the highest signal-to-noise ratios.

Mesh:

Year:  2014        PMID: 24593741      PMCID: PMC3978318          DOI: 10.1118/1.4865781

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  31 in total

1.  Interlaboratory comparison of ultrasonic backscatter, attenuation, and speed measurements.

Authors:  E L Madsen; F Dong; G R Frank; B S Garra; K A Wear; T Wilson; J A Zagzebski; H L Miller; K K Shung; S H Wang; E J Feleppa; T Liu; W D O'Brien; K A Topp; N T Sanghvi; A V Zaitsev; T J Hall; J B Fowlkes; O D Kripfgans; J G Miller
Journal:  J Ultrasound Med       Date:  1999-09       Impact factor: 2.153

2.  Direct strain estimation in elastography using spectral cross-correlation.

Authors:  T Varghese; E E Konofagou; J Ophir; S K Alam; M Bilgen
Journal:  Ultrasound Med Biol       Date:  2000-11       Impact factor: 2.998

3.  Estimation of ultrasound attenuation and dispersion using short time Fourier transform.

Authors:  B Zhao; O A Basir; G S Mittal
Journal:  Ultrasonics       Date:  2005-03       Impact factor: 2.890

4.  Ultrasonic radio-frequency spectrum analysis of normal brain tissue.

Authors:  Martin Strowitzki; Sebastian Brand; Klaus-Vitold Jenderka
Journal:  Ultrasound Med Biol       Date:  2007-04       Impact factor: 2.998

5.  Attenuation estimation using spectral cross-correlation.

Authors:  Hyungsuk Kim; Tomy Varghese
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2007-03       Impact factor: 2.725

6.  A frequency domain model for generating B-mode images with array transducers.

Authors:  Y Li; J A Zagzebski
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  1999       Impact factor: 2.725

7.  Hybrid spectral domain method for attenuation slope estimation.

Authors:  Hyungsuk Kim; Tomy Varghese
Journal:  Ultrasound Med Biol       Date:  2008-07-14       Impact factor: 2.998

8.  Quantitative ultrasound assessment of the rat cervix.

Authors:  Barbara L McFarlin; William D O'Brien; Michael L Oelze; James F Zachary; Rosemary C White-Traut
Journal:  J Ultrasound Med       Date:  2006-08       Impact factor: 2.153

9.  Ultrasound backscatter and attenuation in human liver with diffuse disease.

Authors:  Z F Lu; J A Zagzebski; F T Lee
Journal:  Ultrasound Med Biol       Date:  1999-09       Impact factor: 2.998

10.  Exvivo ultrasound attenuation coefficient for human cervical and uterine tissue from 5 to 10 MHz.

Authors:  Miklos Z Kiss; Tomy Varghese; M A Kliewer
Journal:  Ultrasonics       Date:  2010-11-27       Impact factor: 2.890

View more
  4 in total

1.  System-Independent Ultrasound Attenuation Coefficient Estimation Using Spectra Normalization.

Authors:  Ping Gong; Pengfei Song; Chengwu Huang; Joshua Trzasko; Shigao Chen
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2019-03-05       Impact factor: 2.725

2.  Performance evaluation of the spectral centroid downshift method for attenuation estimation.

Authors:  Kayvan Samimi; Tomy Varghese
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2015-05       Impact factor: 2.725

3.  Optimum Diffraction-Corrected Frequency-Shift Estimator of the Ultrasonic Attenuation Coefficient.

Authors:  Kayvan Samimi; Tomy Varghese
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2016-03-04       Impact factor: 2.725

4.  Reverberation clutter signal suppression in ultrasound attenuation estimation using wavelet-based robust principal component analysis.

Authors:  U-Wai Lok; Ping Gong; Chengwu Huang; Shanshan Tang; Chenyun Zhou; Lulu Yang; Kymberly D Watt; Matthew Callstrom; Joshua D Trzasko; Shigao Chen
Journal:  Phys Med Biol       Date:  2022-04-28       Impact factor: 4.174