Effects of humidity on foil and vial packaging to preserve glucose and lactate test strips for disaster readiness.

Anh-Thu Truong1, Richard F Louie1, John H Vy1, Corbin M Curtis1, William J Ferguson1, Mandy Lam1, Stephanie Sumner1, Gerald J Kost1.   

Abstract

OBJECTIVE: Efficient emergency and disaster response is challenged by environmental conditions exceeding test reagent storage and operating specifications. We assessed the effectiveness of vial and foil packaging in preserving point-of-care (POC) glucose and lactate test strip performance in humid conditions.
METHODS: Glucose and lactate test strips in both packaging were exposed to mean relative humidity of 97.0 ± 1.1% in an environmental chamber for up to 168 hours. At defined time points, stressed strips were removed and tested in pairs with unstressed strips using whole blood samples spiked to glucose concentrations of 60, 100, and 250 mg/dL (n = 20 paired measurements per level). A Wilcoxon signed rank test was used to compare stressed and unstressed test strip measurements.
RESULTS: Stressed glucose and lactate test strip measurements differed significantly from unstressed strips, and were inconsistent between experimental trials. Median glucose paired difference was as high as 12.5 mg/dL at the high glucose test concentration. Median lactate bias was -0.2 mmol/L. Stressed strips from vial (3) and foil (7) packaging failed to produce results.
CONCLUSIONS: Both packaging designs appeared to protect glucose and lactate test strips for at least 1 week of high humidity stress. Documented strip failures revealed the need for improved manufacturing process.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24588956     DOI: 10.1017/dmp.2014.13

Source DB:  PubMed          Journal:  Disaster Med Public Health Prep        ISSN: 1935-7893            Impact factor:   1.385


  2 in total

1.  The Effects of Temperature and Relative Humidity on Point-of-Care Glucose Measurements in Hospital Practice in a Tropical Clinical Setting.

Authors:  Busadee Pratumvinit; Nattakom Charoenkoop; Soamsiri Niwattisaiwong; Gerald J Kost; Panutsaya Tientadakul
Journal:  J Diabetes Sci Technol       Date:  2016-08-22

2.  Computing the surveillance error grid analysis: procedure and examples.

Authors:  Boris P Kovatchev; Christian A Wakeman; Marc D Breton; Gerald J Kost; Richard F Louie; Nam K Tran; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2014-06-13
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.