Analytical performance of a dry chemistry analyzer designed for in-clinic use.

Bente Flatland1, Liesl C Breickner, Michael M Fry.   

Abstract

BACKGROUND: The Heska Dri-Chem 4000 uses dry slide technology to evaluate serum or plasma. No previous independent performance evaluation is published to the authors' knowledge.
OBJECTIVES: The objectives were to (1) characterize analytical performance of a Dri-Chem 4000 by measuring precision and bias, (2) compare analytical performance of that Dri-Chem 4000 unit with a predetermined quality requirement, and (3) determine whether statistical QC of the Dri-Chem 4000 is possible using the 13s control rule.
METHODS: Sixteen analytes were measured using plasma from dogs, cats, and horses. Coefficient of variation (CV), bias, and observed total error (TEobs ) were calculated. TEobs was compared with allowable total error (TEa ). Sigma metric and quality goal index were calculated where relevant. QC validation was performed.
RESULTS: Bias and TEobs calculated using quality control material (QCM) data were smaller than those calculated using method comparison data. Using TEobs calculated from species-specific CV and QCM-based bias, 100% of analytes in each species met ASVCP-recommended TEa . Desired error detection and false rejection rates were achievable using the 13s control rule and ASVCP-recommended TEa values for 9/16 (56%) of analytes in dogs, 9/14 (64%) of analytes in cats, and 8/13 (62%) of analytes in horses.
CONCLUSIONS: Analytical performance of the Dri-Chem 4000 is comparable to that reported by other authors for other small benchtop biochemistry analyzers. Statistical QC using a simple control rule is possible for most analytes in dogs, cats, and horses.
© 2014 American Society for Veterinary Clinical Pathology and European Society for Veterinary Clinical Pathology.

Entities:  

Keywords:  Canine; dry chemistry; equine; feline; method comparison; point-of-care

Mesh:

Year:  2014        PMID: 24588622     DOI: 10.1111/vcp.12122

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  3 in total

1.  Total observed error, total allowable error, and QC rules for canine serum and urine cortisol achievable with the Immulite 2000 Xpi cortisol immunoassay.

Authors:  Jeremie Korchia; Kathleen P Freeman
Journal:  J Vet Diagn Invest       Date:  2022-03       Impact factor: 1.569

2.  Assesment of severity and changes in C-reactive protein concentration and various biomarkers in dogs with pancreatitis.

Authors:  Toru Sato; Koichi Ohno; Takashi Tamamoto; Mariko Oishi; Hideyuki Kanemoto; Kenjiro Fukushima; Yuko Goto-Koshino; Masashi Takahashi; Hajime Tsujimoto
Journal:  J Vet Med Sci       Date:  2016-09-23       Impact factor: 1.267

Review 3.  Current and emerging concepts in biological and analytical variation applied in clinical practice.

Authors:  Bente Flatland; Randolph M Baral; Kathleen P Freeman
Journal:  J Vet Intern Med       Date:  2020-10-21       Impact factor: 3.333

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.