Literature DB >> 24569391

A widely applicable plasma renin activity assay by LC-MS/MS with offline solid phase extraction.

Laura J Owen1, Joanne Adaway, Karen Morris, Stephen Lockhart, Brian G Keevil.   

Abstract

BACKGROUND: The measurement of plasma renin activity using LC-MS/MS is an attractive alternative to radioimmunoassay techniques. A published method for renin activity by LC-MS/MS uses equipment that is not available in all LC-MS/MS laboratories. Here, we present an offline modification to allow users of LC-MS/MS to perform this analysis without specialist equipment.
METHODS: Samples were prepared in duplicate and incubated for 6.5 h and 24 h. Solid phase extraction was performed offline using Waters Oasis MAX ion-exchange 96-well plate. A method comparison was performed between this assay and a previously published assay using on-line solid phase extraction.
RESULTS: The offline method for plasma rennin activity gave similar results to the on-line method across the concentration range. Analytical performance was also found to be comparable and the use of a 24 h incubation for low samples was also deemed unnecessary. DISCUSSION: The method described is an acceptable alternative to a previously published assay which does not require the use of highly specialist equipment and will be applicable to the majority of LC-MS/MS users.

Entities:  

Keywords:  LC-MS/MS; Renin activity; solid phase extraction

Mesh:

Substances:

Year:  2014        PMID: 24569391     DOI: 10.1177/0004563214524558

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  1 in total

1.  Quantitation of Renin Activity in Plasma Using Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS).

Authors:  J Grace van der Gugten; Daniel T Holmes
Journal:  Methods Mol Biol       Date:  2022
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.