A new adjustable glaucoma drainage device.

Adan Villamarin1, Sylvain Roy, Stéphane Bigler, Nikos Stergiopulos.   

Abstract

PURPOSE: This work is focused on the testing of a new experimental noninvasively adjustable glaucoma drainage device (AGDD) that allows for the control of its outflow resistance to modulate intraocular pressure (IOP) in a customized fashion.
METHODS: Six AGDDs were directly connected to a pressure transducer and a perfusion system continuously delivering saline solution at rate of 2 μL/min. The steady-state pressure was measured and reported as a function of the angular position of the AGDD disk. Ex vivo experiments were conducted on six freshly enucleated rabbit eyes. The IOP was measured, and the flow rate was increased with a syringe pump to simulate elevated IOP associated with glaucoma. After insertion of the implant in the anterior chamber, the position of the disk was sequentially adjusted.
RESULTS: The relation between the pressure drop and the angular position of the AGDD disk is nonlinear. The functional range lies between 80° and 130°, which allows for four or five different reproducible adjustment positions. Above 130° the implant is considered to be closed (no outflow), and below 80° it is considered to be open (minimum resistance to flow).
CONCLUSIONS: The resistance to outflow of the experimental AGDD can be adjusted to keep IOP in the desired physiological range. This feature could be useful for addressing the risk of hypotony in the early postoperative stages and could provide a means to achieve optimal IOP under a wide range of postoperative conditions.

Entities:  

Keywords:  adjustable glaucoma; drainage device; filtering surgery; glaucoma; shunt tube

Mesh:

Year:  2014        PMID: 24557347     DOI: 10.1167/iovs.13-12626

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  7 in total

1.  Control and optimisation of fluid flow in glaucoma drainage device surgery.

Authors:  Kin Sheng Lim
Journal:  Eye (Lond)       Date:  2018-01-19       Impact factor: 3.775

2.  eyeWatch™ System Combined with Non-plated Intraorbital Tube Insertion for the Management of Refractory Glaucoma: A Case Series.

Authors:  Efstathios T Detorakis; Adan Villamarin; Sylvain Roy; Stephane Bigler; Georgios Bontzos; Constantinos Stergiopulos; Nikos Stergiopulos
Journal:  J Curr Glaucoma Pract       Date:  2020 May-Aug

3.  iSTAT (Intraluminal Suture Transfixed and Titratable) technique for ligature-free Reversal of long-term drainage implant hypotony.

Authors:  Khin P Kilgore; Ronald L Fellman; Davinder S Grover
Journal:  Am J Ophthalmol Case Rep       Date:  2022-05-04

4.  The Inflammatory and Foreign Body Reaction of Polymethyl Methacrylate Glaucoma Drainage Device in the Rabbit Eye.

Authors:  Virna D Oktariana Asrory; Ratna Sitompul; Widya Artini; Sri Estuningsih; Deni Noviana; William H Morgan
Journal:  Transl Vis Sci Technol       Date:  2020-02-18       Impact factor: 3.283

5.  Comparison of two prototypes of a magnetically adjustable glaucoma implant in rabbits.

Authors:  Birthe Schwerk; Lisa Harder; Claudia Windhövel; Marion Hewicker-Trautwein; Anna Wagner; Jan-Peter Bach; Lena Carolin Voigt; Ulf Hinze; Boris Chichkov; Heinz Haferkamp; Holger Lubatschowski; Stephan Nikolic; Ingo Nolte
Journal:  PLoS One       Date:  2019-04-11       Impact factor: 3.240

6.  Disintegrin-Integrin Binding for Attachment of Polymer Substrate to the Retina.

Authors:  Alejandra Gonzalez-Calle; Rodrigo Brant; Bruno Diniz; Steven Swenson; Frank Markland; Mark S Humayun; James D Weiland
Journal:  J Clin Exp Ophthalmol       Date:  2018-10-08

7.  Outcomes of Deep Sclerectomy following Failed XEN Gel Stent Implantation in Open-Angle Glaucoma: A Prospective Study.

Authors:  Giorgio Enrico Bravetti; Kevin Gillmann; Harsha L Rao; André Mermoud; Kaweh Mansouri
Journal:  J Clin Med       Date:  2022-08-16       Impact factor: 4.964

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.