Incidence of nonphysiologic short VV intervals detected by the sensing integrity counter with integrated bipolar compared with true bipolar leads: clinically inconsequential or cause for concern?

Justin Ng1, Nagesh Chopra, Chirag Barbhaiya, Tobias Reichlin, Eyal Nof, Koichi Nagashima, Thomas Tadros, Bruce A Koplan.   

Abstract

INTRODUCTION: The need to detect impending implantable cardiac defibrillator (ICD) lead failure has grown. Automated sensing diagnostics have been developed for this reason. The sensing integrity counter (SIC) is one such oversensing diagnostic, which forms an integral part of the Medtronic™ lead integrity alert (LIA) feature on implantable defibrillators. It records nonphysiologic short VV intervals (NPSVVIs). It is unclear whether SIC data derived from integrated bipolar (IBP) leads need to be interpreted differently when compared to true bipolar (TBP) leads. We hypothesized that IBP ICD leads by virtue of a larger "antennae" may generate more NPSVVIs on than TBP leads, leading to more false-positive SIC counts.
METHODS: Equal durations of remote monitoring records of 44 patients (mean age of 65.9 ± 2.2 years, 52 % female) with IBP ICD leads and Medtronic (MDT) generators (IBP group) were compared with those of 44 randomly selected patients (64.0 ± 2.2 years, 24 % female) who had TBP ICD leads and MDT generators (TBP group). Mean surveillance time, defined as the time over which the cumulative SIC count was acquired, was 614 ± 44 days (TBP group) vs. 620 ± 49 days (IBP group, p = ns). The mean time of follow-up following the first documented short VV interval was 115.2 months in the integrated bipolar group and 66.9 months in the true bipolar group. Leads on advisory were excluded from the study.
RESULTS: A total of 26/44 patients in the IBP group displayed NPSVVI compared to 11/44 patients in the TBP group (59 vs. 25 %; p = 0.002, Fisher exact test). When adjusted for gender and lead age, the difference was still significant (p = 0.008). When evaluating the clinical consequence of NPSVVI in this cohort, 3/11 TBP leads with NPSVVI of >0 were eventually extracted due to additional abnormalities vs. 0/26 IBP leads with NPSVVI (p = 0.02, Fisher exact test). None of the IBP group patients with NPSVVI have developed inappropriate therapy from lead noise or a need for abandonment or extraction.
CONCLUSION: Integrated bipolar ICD leads are more likely to have elevated SIC counts than true bipolar leads despite revealing no other evidence of lead failure. There does not appear to be a need for heightened surveillance in IBP leads with observed elevated SIC counts that have no other findings to suggest lead malfunction.

Entities:  

Mesh:

Year:  2014        PMID: 24532110     DOI: 10.1007/s10840-013-9866-7

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  7 in total

Review 1.  Ventricular arrhythmias in patients with implanted defibrillators.

Authors:  William G Stevenson; Roy M John
Journal:  Circulation       Date:  2011-10-18       Impact factor: 29.690

2.  The contradiction of appropriate shocks in primary prevention ICDs: increasing and decreasing the risk of death.

Authors:  Michael O Sweeney
Journal:  Circulation       Date:  2010-12-06       Impact factor: 29.690

3.  Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study.

Authors:  Charles D Swerdlow; Bruce D Gunderson; Kevin T Ousdigian; Athula Abeyratne; Haresh Sachanandani; Kenneth A Ellenbogen
Journal:  Circulation       Date:  2010-09-27       Impact factor: 29.690

4.  Detection of a Riata™ insulation failure by the Medtronic Lead Integrity Alert™.

Authors:  James Kneller; William Delacey; Mark A Wood; Kenneth A Ellenbogen
Journal:  Europace       Date:  2012-02-26       Impact factor: 5.214

5.  Causes of ventricular oversensing in implantable cardioverter-defibrillators: implications for diagnosis of lead fracture.

Authors:  Bruce D Gunderson; Charles D Swerdlow; Jay M Wilcox; Jean E Hayman; Kevin T Ousdigian; Kenneth A Ellenbogen
Journal:  Heart Rhythm       Date:  2010-01-15       Impact factor: 6.343

6.  Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures.

Authors:  Linda M Kallinen; Robert G Hauser; Chuen Tang; Daniel P Melby; Adrian K Almquist; William T Katsiyiannis; Charles C Gornick
Journal:  Heart Rhythm       Date:  2010-05-21       Impact factor: 6.343

7.  Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads.

Authors:  Charles D Swerdlow; Bruce D Gunderson; Kevin T Ousdigian; Athula Abeyratne; Robert W Stadler; Jeffrey M Gillberg; Amisha S Patel; Kenneth A Ellenbogen
Journal:  Circulation       Date:  2008-11-03       Impact factor: 29.690

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.