Literature DB >> 24527185

Occurrence of longitudinal stent compression before stent deployment: Two case studies.

Adel Aminian1, Jacques Lalmand1, Dariouch Dolatabadi1.   

Abstract

Several recent reports have described the occurrence of longitudinal stent deformation (LSD, defined as the distortion or shortening of a stent along the longitudinal axis), following its successful deployment. However, few reports have described LSD prior to any stent deployment. This previously unrecognized complication is the result of modifications to stent design. It has been noted that the new-generation stent platforms have a reduced number of connectors, which in turn causes a reduction in longitudinal stent strength. To corroborate previous findings by our lab and others (Vijayvergiya et al, 2013), we describe here two cases of LSD prior to stent deployment that occurred due to crushing of the proximal stent edge by the guide catheter while attempting to withdraw the crimped stent. In addition, we discuss the associated risk factors, such as the length of the stent, and specific management strategies, including technical guidelines and use of fluoroscopic guidance for maneuvering the stent during the procedure.

Entities:  

Keywords:  Complication; Longitudinal stent compression; Longitudinal stent deformation; Percutaneous coronary intervention; Stent structure

Year:  2014        PMID: 24527185      PMCID: PMC3920163          DOI: 10.4330/wjc.v6.i1.23

Source DB:  PubMed          Journal:  World J Cardiol


  3 in total

1.  Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration Manufacturer and User Facility Device Experience database.

Authors:  Mamas A Mamas; Paul D Williams
Journal:  EuroIntervention       Date:  2012-06-20       Impact factor: 6.534

2.  Longitudinal stent compression of everolimus-eluting stent: A report of 2 cases.

Authors:  Rajesh Vijayvergiya; Alok Kumar; Smit Shrivastava; Naveen K Kamana
Journal:  World J Cardiol       Date:  2013-08-26

3.  Major longitudinal deformation of a new-generation drug-eluting stent during withdrawal into the guide catheter.

Authors:  Adel Aminian; Jacques Lalmand
Journal:  J Invasive Cardiol       Date:  2012-12       Impact factor: 2.022