| Literature DB >> 24524074 |
Trygve Eftestøl1, Lawrence D Sherman2.
Abstract
BACKGROUND: During resuscitation of cardiac arrest victims a variety of information in electronic format is recorded as part of the documentation of the patient care contact and in order to be provided for case review for quality improvement. Such review requires considerable effort and resources. There is also the problem of interobserver effects.Entities:
Mesh:
Year: 2014 PMID: 24524074 PMCID: PMC3913461 DOI: 10.1155/2014/276965
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Signals and data recorded by three different automated external defibrillators: (a) Philips Forerunner 2, (b) Philips MRx, and (c) Physiocontrol Lifepak 12. The blue and red tracings show the electrocardiogram and thoracic impedance, respectively. Examples of information recorded in the defibrillator's electronic log are shown above each plot window. Annotations of rhythm transitions and therapeutic events are shown at the top and bottom inside each plot window.
Figure 2The process of deriving the variables from the manual review; the generation of the representations used for the further algorithmic reasoning for deriving the variables automatically.
The time variables with descriptive names, variable names, coding, and explanation of each variable.
| Variable | Possible values | Description | |
|---|---|---|---|
| ECG start time | ecghr | (0–23, 99) | The hour of actual start time of the ECG |
| ECG start time | ecgmn | (0–59, 99) | The minute of the actual start time of the ECG |
| ECG start time | ecgsc | (0–59, 99) | The second of the actual start time of the ECG |
| First compression time | fchr | (0–23, 88, 99) | Hour of first compression, prior to shock. |
| First compression time | fcmn | (0–59, 88, 99) | Minute of first compression, prior to shock. |
| First compression time | fcsc | (0–59, 88, 99) | Second of first compression, prior to shock. |
| Last compression time | lchr | (0–23, 88, 99) | Hour of last compression before shock |
| Last compression time | lcmn | (0–59, 88, 99) | Minute of last compression before shock |
| Last compression time | lcsc | (0–59, 88, 99) | Second of last compression before shock |
| Shock time | shkhr | (0–23, 99) | Actual hour of shock delivery |
| Shock time | shkmn | (0–59, 99) | Actual minute of shock delivery |
| Shock time | shksc | (0–59, 99) | Actual second of shock delivery |
| VF onset time | vfonsethr | (0–23, 99) | Hour of VF onset, best estimate when underneath CPR artifact |
| VF onset time | vfonsetmm | (0–59, 99) | Minute of VF onset, best estimate when underneath CPR artifact |
| VF onset time | vfonsetss | (0–59, 99) | Second of VF onset, best estimate when underneath CPR artifact |
| ROSC time | rschr | (0–23, 99) 99 = unknown | Actual hour of ROSC |
| ROSC time | rscmn | (0–59, 99) 99 = unknown | Actual minute of ROSC |
| ROSC time | rscsc | (0–59, 99) 99 = unknown | Actual second of ROSC |
The variables describing device operation and therapy with descriptive names, variable names, coding, and explanation of each variable.
| Variable | Possible values | Description | |
|---|---|---|---|
| Shock number | shkn | ≥0 and <21 | Shock sequence number. |
| Number of shock sequences recorded | ssrecord | (>0 and <31) or 99 | This describes the number of electrical shocks delivered to the patient as recorded on the AED total shocks received by the patient. |
| Number of shocks in sequence | shks | 1–31 | The number of shocks without intervening CPR. For example, 1 = no stacked shocks while > 1 = stacked shocks. After 2005 no case should have stacked shock. |
| Mode | mode | 1 = manual | This is the mode of the AED at the time of the shock |
| EMS CPR prior to shock? | cpr | 1 = yes | Description of if EMS CPR was administered to patient before shock was delivered. |
| Impedance at 1st shock | imp | Measured in Ohms | Impedence at time of first shock sequence (50–200, 999) 999 = unknown. |
| Energy of 1st shock | enrgy | Measured in Joules | Energy of first shock in sequence. |
The response variables with descriptive names, variable names, coding, and explanation of each variable.
| Variable | Possible values | Description | |
|---|---|---|---|
| Initial rhythm | init_rhy | 1 = asystole, | Description of what the initial rhythm recorded was as determined from the AED. Codes in parenthesis refer to indeterminate between the two stated rhythms. |
|
| |||
| Preshock rhythm | rhyb4 | 1 = asystole, | Description of what the preshock rhythm recorded was. Codes in parenthesis refer to indeterminate between the two stated rhythms. |
|
| |||
| Rhythm 10 secs after the last shock | r10 | 1 = asystole, | Description of what rhythm recorded 10 seconds after the shock was. Codes in parenthesis refer to indeterminate between the two stated rhythms. |
|
| |||
| Rhythm 30 secs after the last shock | r30 | 1 = asystole, | Description of what rhythm recorded 30 seconds after the shock was. Codes in parenthesis refer to indeterminate between the two stated rhythms. |
|
| |||
| Rhythm 60 secs after the last shock | r60 | 1 = asystole, | Description of what rhythm recorded 60 seconds after the shock was. Codes in parenthesis refer to indeterminate between the two stated rhythms |
|
| |||
| Rhythm 120 sec after the last shock | r120 | 1 = asystole, | Description of what rhythm recorded 120 seconds after the shock was. Codes in parenthesis refer to indeterminate between the two stated rhythms. |
|
| |||
| VF prior to next shock? | vfpr | 1 = yes, 2 = no, and 9 = unknown | Description of whether or not there was VF between shocks, or between the last shock and the end of this recording as determined from audio or the MRIF. Not to capture VF at any time after device turned off. |
|
| |||
| Organized rhythm prior to next shock? | orgpr | 1 = yes, 2 = no, and 9 = unknown | Description of whether or not there was an organized rhythm between shocks, or between the last shock and the end of the recording. |
|
| |||
| ROSC | rosc | 1 = yes, 2 = no, and 9 = unknown | Description of whether return of spontaneous circulation occurred. |
The primary information objects (PIO). These variables constitute the primary information objects (PIO) and are used to model the entire resuscitation episode. The therapy and response domains are described independently. For each domain the entire time span of the episode is described as a sequence of interchanging states. For each state change, the corresponding transition times are given which specify the times entering and leaving the state. The entering time of a state corresponds to the leaving time of the prior state unless it is the beginning of the episode. The leaving time of a state corresponds to the entering time for the next state unless it is the end of the episode.
| Variable domain | ||||
|---|---|---|---|---|
| Transition time | Patient's response | Therapy | ||
| State | Code | State | Code | |
| Seconds | Ventricular fibrillation | VF | Chest compressions | C |
| Seconds | Ventricular tachycardia | VT | Hands off interval | H |
| Seconds | Asystole | AS | Defibrillation | D |
| Seconds | Pulseless electrical activity | PE | Unknown | U |
| Seconds | Pulse generating rhythm | PR | ||
| Seconds | Unknown | UN | ||
The variables in the original database which will be automatically replicated. The table is organized in columns to highlight the type of information the variables provide: time events, patient's response, and device operation and therapy. Each variable is labeled according to it being a primary information object (PIO) or secondary information object (SIO). All the variables labeled as PIO can also be found in the table listing the PIO variables (Table 4). The remaining variables labeled as SIO can be automatically derived from the PIO variables (listed in table PIO) by designing proper reasoning algorithms.
| Variable domain | |||||
|---|---|---|---|---|---|
| Time events | Patient's response | Device operation and therapy | |||
| First compression time | PIO | Initial rhythm | SIO | Shock number | SIO |
| Last compression time | PIO | Preshock rhythm | SIO | Number of shock seq. recorded | SIO |
| VF onset time | PIO | Rhythm 10 sec after last shock | SIO | Number of shocks in sequence | SIO |
| ROSC time | PIO | Rhythm 30 sec after last shock | SIO | Mode | SIO |
| Shock time | PIO | Rhythm 60 sec after last shock | SIO | EMS CPR prior to shock? | SIO |
| Rhythm 120 sec after last shock | SIO | Impedance at 1st shock | SIO | ||
| ECG start time | SIO | VF prior to next shock? | SIO | Energy of 1st shock | SIO |
| Org. rhythm prior to next shock? | SIO | ||||
| ROSC | SIO | ||||
Figure 3Case recording from MRX defibrillator. The blue tracing is the ECG. Impedance tracing is shown in red. The defibrillator log events (shock advised, charge, shock delivered, analysis started, and artifact detected), the rhythm transitions (vf: ventricular fibrillation, as: asystole, and pe: pulseless electrical activity), and annotations for start and end of compressions (c1 and c2) and defibrillations (d1 and d2) are shown. Refer to Figure 4 for the corresponding RORE representation.
Figure 4The representation of the resuscitation episode (RORE) including the therapy domain representation, the response domain representation, and the episode representation (for detailed explanation see Appendix C). UN = unknown, C = compressions, H = hands off chest, D = shock, CVF = compressions during VF, CAS = compressions during asystole, DVF = defibrillation for VF, PE = pulseless electrical activity, and PR = pulsatile rhythm.
Figure 5Summary of the match rates comparing the manually and automatically derived variable values. (a) shows match rates for rhythms at the preselected times from the database. (b) shows data regarding the defibrillator logs of shock data. (c) shows the times for ECG start, shock times, first compression and last compression times, VF onset times, and time of return of spontaneous circulation. All results are given in percentages of the ratio: correct/(correct + error). For the time variables, the lower row shows the ratio of automatically generated codes for missing values matching the manually given missing codes (gray bars).
Results for the comparison between manual and automatic recording of time event variables. The table counts the number of correct, wrong, and missing values. An automatic recording is considered correct if the deviation from the manual recording is less than or equal to one second. Otherwise the recording is considered wrong. A recording is considered missing if the code 66 : 66 : 66, 88 : 88 : 88, 99 : 99 : 99 is used for either of the recordings or the manual recording was changed to 00 : 00 : 00 or 12 : 00 : 00.
| Time variables | ||||||
|---|---|---|---|---|---|---|
| Time point | ECG start | Defib shock | First compression | Last compression | VF onset | ROSC pulse |
| Abbreviated | ecgtm | shktm | fctm | lctm | vfonsettm | rsctm |
| Correct | 33 | 138 | 90 | 108 | 45 | 15 |
| Wrong | 4 | 2 | 0 | 1 | 5 | 0 |
| Missing | 17 | 0 | 50 | 31 | 90 | 125 |
Results for the comparison between manual and automatic recording of patient response variables. The table counts the number of correct and wrong values. An automatic recording is considered correct if the automatic recording is identical to the manual recording. Otherwise the recording is considered wrong.
| Response variables | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Variable name | Initial rhythm | Preshock rhythm | Rhythm 10 secs after the last shock | Rhythm 30 secs after the last shock | Rhythm 60 secs after the last shock | Rhythm 120 secs after the last shock | VF prior to next shock? | Organized rhythm prior to next shock | ROSC |
| Abbreviated | init_rhy | rhyb4 | r10 | r30 | r60 | r120 | vfpr | orgpr | rosc |
| Correct | 140 | 136 | 135 | 136 | 137 | 134 | 138 | 133 | 127 |
| Wrong | 0 | 4 | 5 | 4 | 3 | 6 | 2 | 7 | 3 |
Results for the comparison between manual and automatic recording of device operation and therapy variables for the data set. The table counts the number of correct and wrong values. For all variables except imp and enrgy an automatic recording is considered correct if the automatic recording is identical to the manual recording. Otherwise the recording is considered wrong. For imp and enrgy the numeric deviations between the manual and automatic recordings are considered. A deviation of zero is considered correct, and larger deviations are counted as errors.
| Therapy variables | |||||||
|---|---|---|---|---|---|---|---|
| Variable name | Shock number | Number of shock sequences recorded | Number of shocks in sequence | Mode | EMS CPR prior to shock? | Impedance at 1st shock | Energy of 1st shock |
| Abbreviated | shkn | ssrecord | shks | mode | cpr | imp | enrgy |
| Correct | 140 | 140 | 140 | 119 | 137 | 123 | 120 |
| Wrong | 0 | 0 | 0 | 21 | 3 | 17 | 20 |