Primary Stability of a Hybrid Implant Compared with Tapered and Cylindrical Implants in an Ex Vivo Model.

Takeshi Toyoshima1, Hideaki Tanaka1, Yasunori Ayukawa2, Miori Howashi2, Tomohiro Masuzaki2, Takahiro Kiyosue1,3, Kiyoshi Koyano2,3, Seiji Nakamura1.   

Abstract

BACKGROUND: Hybrid implants are thought to offer good stability and bone-implant contact.
PURPOSE: This study sought to evaluate the primary stability of a hybrid implant compared with tapered and cylindrical implants in an ex vivo model and compare undersized drilling with standard drilling in attaining primary stability.
MATERIALS AND METHODS: Hybrid (Straumann® Bone Level; BL), tapered (Straumann® Tapered Effect; TE), and cylindrical (Straumann® Standard Plus; SP) implants were inserted into 15 mm-thick porcine iliac crest blocks using standard (corticocancellous or cancellous bone) or undersized (cancellous bone only) drilling (n = 7). Stability was evaluated using maximum insertion torque (MIT), implant stability quotient (ISQ), and push-out tests.
RESULTS: The mean MIT for BL and TE was significantly higher than that for SP (p < 0.05). With standard drilling in corticocancellous bone, the mean ISQ for TE was significantly higher than that for SP (p < .05). The mean MIT, ISQ, and push-out test scores for BL, TE, and SP with standard drilling in corticocancellous bone were significantly higher than those with undersized or standard drilling in cancellous bone (p < .05).
CONCLUSIONS: In this ex vivo model, the hybrid implant demonstrated promising primary stability compared with the cylindrical implant. The improved stability of the hybrid implant was especially evident in cancellous bone. The influence of the cortical layer should be recognized when selecting implants for primary stability.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  animal model; dental implant; implant stability quotient; maximum insertion torque; primary stability; push-out test

Mesh:

Substances:

Year:  2014        PMID: 24506148     DOI: 10.1111/cid.12205

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  7 in total

1.  Implant stability in different techniques of surgical sites preparation: an in vitro study.

Authors:  C Rastelli; G Falisi; R Gatto; M Galli; E Saccone; M Severino; C DI Paolo
Journal:  Oral Implantol (Rome)       Date:  2014-12-27

2.  Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Authors:  Pedro Ferrás Fernandes; Liliana Grenho; Maria Helena Fernandes; João Carlos Sampaio-Fernandes; Pedro Sousa Gomes
Journal:  Odontology       Date:  2021-08-27       Impact factor: 2.634

3.  The effects of surgical preparation techniques and implant macro-geometry on primary stability: An in vitro study.

Authors:  G Falisi; M Severino; C Rastelli; S Bernardi; S Caruso; M Galli; L Lamazza; C Di Paolo
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2017-03-01

4.  Correlation between Implant Geometry, Bone Density, and the Insertion Torque/Depth Integral: A Study on Bovine Ribs.

Authors:  Danilo Alessio Di Stefano; Paolo Arosio; Vittoria Perrotti; Giovanna Iezzi; Antonio Scarano; Adriano Piattelli
Journal:  Dent J (Basel)       Date:  2019-03-05

5.  Primary stability of different implant macrodesigns in a sinus floor elevation simulated model: an ex vivo study.

Authors:  Mikio Imai; Yoichiro Ogino; Hideaki Tanaka; Kiyoshi Koyano; Yasunori Ayukawa; Takeshi Toyoshima
Journal:  BMC Oral Health       Date:  2022-08-08       Impact factor: 3.747

6.  Effect of implant macro-design on primary stability: A prospective clinical study.

Authors:  Naroa Lozano-Carrascal; Oscar Salomó-Coll; Marta Gilabert-Cerdà; Nuria Farré-Pagés; Jordi Gargallo-Albiol; Federico Hernández-Alfaro
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2016-03-01

Review 7.  Influence of different implant placement techniques to improve primary implant stability in low-density bone: A systematic review.

Authors:  Jayant N Palaskar; Nikhil Joshi; Pooja M Shah; Poorva Gullapalli; Vineet Vinay
Journal:  J Indian Prosthodont Soc       Date:  2020-01-27
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.