Continuous registration based on computed tomography for breathing motion compensation.

Dominik Spinczyk1, Jaroslaw Zyłkowski2, Tadeusz Wróblewski3.   

Abstract

INTRODUCTION: Image guidance for intervention is applied for complex and difficult anatomical regions. Nowadays, it is typically used in neurosurgery, otolaryngology, orthopedics and dentistry. The application of the image-guided system for soft tissues is challenging due to various deformations caused by respiratory motion, tissue elasticity and peristalsis. AIM: The main task for the presented approach is continuous registration of preoperative computed tomography (CT) and patient position in the operating room (OR) without touching the patient and compensation of breathing motion. This approach is being developed as a step to image-guided percutaneous liver RF tumor ablation.
MATERIAL AND METHODS: Up to ten integrated radiological markers are placed on the patient's skin before CT scans. Then the anatomical model based on CT images is calculated. Point-to-point registration based on the Horn algorithm during a few breathing cycles is performed using a videometric tracking system. The transformation which corresponds to the minimum fiducial registration error (FRE) is found during the registration and it is treated as the initial transformation for calculating local deformation field of breathing motion compensation based on the spline approach.
RESULTS: For manual registration of the abdominal phantom, the mean values of target registration error (TRE), fiducial localization error (FLE) and FRE are all below 4 mm for the rigid transformation and are below 1 mm for the affine transformation. For the patient's data they are all below 9 mm and 6 mm, respectively. For the automatic method, different marker configurations have been evaluated while dividing the respiratory cycle into inhale and exhale. Average median values for FRE, TRE rigid estimation and TRE based on spline deformation were 15.56 mm, 0.82 mm and 7.21 mm respectively.
CONCLUSIONS: In this application two registration methods of abdominal preoperative CT anatomical model and physical patient position in OR were presented and compared. The presented approach is being developed as a step to image-guided percutaneous liver radiofrequency ablation tumor ablation. Implementation of the automated registration method to clinical practice is easier because of shortening of preparation time in OR, no necessity of touching the patient, and no dependency on the physician's experience.

Entities:  

Keywords:  image-guided intervention; percutaneous ablation; preoperative anatomical model registration; respiratory motion compensation

Year:  2013        PMID: 24501595      PMCID: PMC3908646          DOI: 10.5114/wiitm.2013.39505

Source DB:  PubMed          Journal:  Wideochir Inne Tech Maloinwazyjne        ISSN: 1895-4588            Impact factor:   1.195


Introduction

Image guidance for intervention is applied for complex and difficult anatomical regions. Nowadays, it is typically used in neurosurgery, otolaryngology, orthopedics and dentistry [1-4]. The application of the image-guided system for soft tissues is challenging due to various deformations caused by respiratory motion, tissue elasticity and peristalsis [5, 6]. Most patients with liver tumors are not candidates for surgical resections, because of unfavorable pathological sites of tumor, widespread metastases and poor general physical condition. For these patients there is still a chance in minimally invasive, percutaneous ablations of lesions. The major task of these procedures is accurate needle placement. The conventional method for guiding needle insertion is ultrasound imaging, but small and deeply situated lesions are difficult to identify. The main task for the presented approach is continuous registration of preoperative computed tomography (CT) and patient position in the operating room (OR) without touching the patient and compensation of breathing motion. Nowadays, this kind of system is at the development stage [7]. One application of the Image-Guided Surgery Toolkit [8] is CT-based liver tumor biopsy [9]. Rigid transformation and the Horn algorithm [10] are used to register the preoperative model and the patient's position. The needle is tracked by an optical tracking system. Fiducial markers, entry and target points are manually selected in CT images. Target registration error based on Fitzpatrick estimation [11] is calculated. Banovac et al. [12] use an electromagnetic tracking system to traverse needles into the radiopaque tumor target as well as point-to-point registration algorithm. The average registration error between the CT coordinate system and electromagnetic coordinate system was 1.4 ±0.3 mm) in the phantom and 1.9 ±0.3 mm) in the swine. Maier-Hein et al. [13, 14] proposed a CT-based, optical tracking system and a real-time deformation model to continuously estimate the position of the target in the liver. A spline-based approach was used to capture local deformations between the preoperative CT anatomical model and the patient's position in the OR. The proposed approach was tested combining internal and external markers in a custom-designed respiratory liver motion simulator and in swine tissues. Diagonal arrangement of the needle-shaped fiducials yielded the best results. The rest of the article presents the proposed registration method, which is used to compensate respiratory movements. The proposed method consists of two steps: rigid and non-rigid registration.

Aim

The main task for the presented approach is continuous registration of preoperative CT and patient position in the OR without touching the patient and compensation of breathing motion. This approach is being developed as a step to image-guided percutaneous liver radiofrequency ablation (RF) tumor ablation.

Material and methods

Registering a preoperative anatomical model and the patient's position starts with a rigid registration. In our approach, the initial registration is based on radiological markers. The presented method uses Beekley CT Spots skin markers which are attached to the patient's abdominal skin before the CT examination and then are used for patient registration. The selected markers were used because they do not produce artifacts on CT images and it is possible to pinpoint their position. Markers are manually selected: in the CT abdominal examination and on the patient's skin. At the very beginning (it was a pilot study) the Polaris Vicra tracking system was used and markers on the patient's skin were manually selected. We used up to ten markers to test various configurations of the marker's positions relative to the target point. West et al. [9] demonstrated the importance of configuration of markers’ positions relative to the target point to the registration error. The description of anatomical position of the skin markers is presented in Figure 1 and Table I. The following configuration of markers against the central marker were tested: all, up-down, left-right and diagonal. For each configuration, the central marker was set as a target. To find rigid mapping between two Cartesian coordinate systems, data must include three or more corresponding non-collinear points. Horn [10] proposed a closed form solution based on a least-squares formulation. Optimal rotation and translation matrices are found using singular value decomposition (SVD) of a correlation matrix:
Figure 1

A–C. Diagram of patient's abdominal marker positions

Table I

Anatomical position of skin markers for manual registration

Marker numberAnatomical locationGrounds
1Xiphoid processPoint fixed to bony structures in the midline of the body
2Intersection of midclavicular line and right costal marginAnatomical point fixed to bony structures
3Intersection of anterior axillary line and right costal marginAnatomical point fixed to bony structures
4Intersection of midclavicular line and left costal marginAnatomical point fixed to bony structures
5Intersection of anterior axillary line and left costal marginAnatomical point fixed to bony structures
6First control point on the abdominal surfaceRegistration accuracy validation
7UmbilicusAnatomical point not fixed to bony structures
8Second control point on the abdominal surfaceRegistration accuracy validation
A–C. Diagram of patient's abdominal marker positions Anatomical position of skin markers for manual registration where is the correlation matrix, Si are points in the first Cartesian coordinate system, Ti are points in the second Cartesian coordinate system, σ are non-negative singular values of the correlation matrix and U and V are orthonormal matrices,and where , are average values of the point coordinates in the first and second coordinate systems, respectively. The method has been evaluated on imaging data and patients with any liver diseases from the Second Department of Clinical Radiology for patients on the Chair and Department of General, Transplant and Liver Surgery at the Medical University of Warsaw after acceptance of the ethics committee of the Medical University of Warsaw (GE Light Speed 16 Pro CT Scanner; Contrast medium: Iomeron 400 or Ultravist 370; Intravenous bolus: 10 ml of 0.9% NaCl, 1 ml/kg contrast medium and thereafter 30 ml of 0.9% NaCl all with 6 ml/s infusion speed). In order to assess the results, FLE, FRE and TRE were used. Following the Fitzpatrick estimation of registration error [10], different markers were evaluated as targets. The results of this study are presented in the next section. After the pilot study we found a few disadvantages of the manual registration: the necessity of touching the patient's skin manually, prolonging the time of the procedure, and dependence of accuracy on the physician's experience. Due to the disadvantages we decided to implement an automatic continuous registration. The automated method in this case means no need for multiple manual indication of the position of markers on the patient. Position of markers is determined not once, but several times during the respiratory cycle. We used the videometric tracking system of Micron Tracker Claron Hx. 40. Claron Hx 40 is stereo camera which tracks a unique black-white template with an error of less than 0.5 mm after calibration. We put the markers on the patient before CT scans but unlike in a manual registration all markers were placed on one (right) side of the abdomen for visibility of the tracking system. We constantly grab the position of the marker (a few times a second) and calculate the FRE error between the DICOM coordinate system and patient (tracker) coordinate system. The resulting transform is organized by the error FRE and the transform which corresponds to the minimum is treated as the result of rigid registration.

Breathing motion compensation

Breathing motion compensation is necessary to select a proper breathing phase for needle insertion and estimate the target movement. For breathing motion compensation, a local deformation field based on displacement of tracking skin markers is calculated. The set of marker positions (in the tracker coordinate system) during tracking time is treated as reference points for the local deformation field, based on a spline approach [15, 16]. The deformation equation: calculates a new position of point x in space, based on spline coefficients matrix D and affine transformation Ax + b. Based on the position of tracking skin markers in the specific time and after a time interval, unknown spline coefficients of the current deformation field are found and then this field can be used to estimate the new target position. To evaluate the correctness of deformation one marker is treated as a target. Due to TRE estimation [11] usually the central marker is treated as a target. The transformation obtained for minimum FRE in the rigid registration step is used to translate the target from the DICOM coordinate system to the tracker coordinate system.

Selecting proper breathing phase for needle insertion

For tracking breathing, the marker positions are continuously measured by the videometric tracking system. Time course of marker positions is analyzed. Breathing cycles are detected based on local extremes. The breath is the process variable over time, so there is a need for averaging it [17, 18]. Amplitude of respiratory motion is not measured directly, but through the respiratory phase. The filtering signal derived from the so-called breathing bag with a Gaussian filter to unambiguously determine the minimum and maximum is applied. A mean of the maximum breathing phase is assigned by the formula (assuming zero breathing phase for minima) [18]: where N max is the number of the observed breathing cycles and and are moments of time corresponding to the next extremes of marker position, and respiratory phase, for any time tj, is interpolated linearly [18]: Calculating the breathing phase allows breathing data to be grouped from many breathing cycles and also data to be compared between patients.

Results

Manual rigid registration

The rigid registration results of a preoperative CT anatomical model and a physical patient position are presented both for the phantom (Table II) and patient data (Table III). To check the dependence of the results on the operator and at the same time not to extend too much time with the patient, the method was evaluated by two operators.
Table II

Errors of phantom markers registration using the rigid or affine transformation in millimeters (Polaris Vicra)

Kind of error Error typeError value
Rigid registrationAffine registration
Operator 1Operator 2Operator 1Operator 2
Case 1Case 2Case 1Case 2Case 1Case 2Case 1Case 2
FRE2.081.911.564.330.520.520.871.04
TRE marker 13.292.772.607.270.520.690.871.04
TRE marker 51.561.211.042.940.350.520.350.52
TRE marker 91.732.772.256.580.520.520.871.04
TRE marker 32.422.771.915.370.520.520.870.87
TRE marker 84.503.813.127.620.690.691.041.21
FLE2.602.422.085.540.350.520.690.69
Table III

Errors of patient's markers registration using the rigid or affine transformation for markers configuration: 8-4-5-2-6 in millimeters (using Polaris Vicra)

Kind of error Error typeError value
Rigid registrationAffine registration
Operator 1Operator 2Operator 1Operator 2
FRE7.15.04.73.1
TRE marker 111.47.87.44.8
TRE marker 57.35.04.73.1
TRE marker 911.88.17.65.0
TRE marker 39.06.65.74.0
TRE marker 85.03.53.32.3
FLE9.26.65.94.0
Errors of phantom markers registration using the rigid or affine transformation in millimeters (Polaris Vicra) Errors of patient's markers registration using the rigid or affine transformation for markers configuration: 8-4-5-2-6 in millimeters (using Polaris Vicra) For the manual registration of the abdominal phantom, the mean values of target registration error TRE, FLE, FRE are all below 4 mm for the rigid transformation and are below 1 mm for the affine transformation. For the patient's data they are all below 9 mm and 6 mm, respectively. The smallest errors are obtained for the target position located in the center of the marker coordinate system (target 5 in Figure 2 – Table II and target 6 in Figure 2 – Table III).
Figure 2

A, B. Diagram of phantom abdominal marker positions

A, B. Diagram of phantom abdominal marker positions

Automatic continuous rigid registration and respiratory motion compensation

The results of automatic registration of the preoperative CT anatomical model and a physical patient position are presented below. The anatomical location of markers is presented in Table III. Due to the field of view for the Micron Tracking system we restricted the area of marker placement to one side of the patient's body. The estimation of a new target position is calculated following the deformation field equation. Table IV presents two cases for a breathing phase divided into three groups:
Table IV

Errors of patient's markers registration for three breathing phase groups

Breathing phaseError typeMedian value
Case 3Case 4
ExhaleFRE_rigid7.044.26
TRE_spline4.361.08
TRE_rigid0.570.80
MiddleFRE_rigid7.057.05
TRE_spline4.371.57
TRE_rigid1.390.49
InhaleFRE_rigid7.17.73
TRE_spline4.441.91
TRE_rigid1.81.09
exhale (0.85 ≤ breathing_phase ≤ 0.15), middle (0.15 < breathing_phase ≤ 0.3 or 0.7 ≤ breathing_phase < 0.85), inhale (0.3 < breathing_phase < 0.7). Errors of patient's markers registration for three breathing phase groups

Discussion

There are some patients who demonstrate a little difference in rigid registration error in the whole breathing cycle. The error of target position estimation for a specific patient and for rigid and spline estimations is similar for the whole breathing cycle, but generally it rises with an increase in respiratory phase difference between preoperative data and a temporary breathing phase. Median of rigid fiducial registration error was used as a criterion to distinguish an inhale and exhale phase [13, 14]. Table V presents results based on this criterion (for the same patient as in Table IV).
Table V

Errors of patient's markers registration for median FRE

Median FRE breathing phaseError typeMedian value
Case 3Case 4
ExhaleFRE_rigid6.812.44
TRE_spline4.541.18
TRE_rigid1.410.80
InhaleFRE_rigid7.189.76
TRE_spline4.162.63
TRE_rigid1.230.85
Errors of patient's markers registration for median FRE Table VI presents registration and target position estimation errors for different marker configurations. Reconfiguring the markers will change registration errors.
Table VI

Errors of patient's markers registration for different marker configuration

Markers configurationMedian FRE breathing PhaseError typeMedian value
Case 3Case 4
2_updownExhaleFRE_rigid8.166.67
TRE_spline2.281.17
TRE_rigid0.120.74
InhaleFRE_rigid9.969.82
TRE_spline3.180.94
TRE_rigid3.320.65
2_diagonalExhaleFRE_rigid6.581.66
TRE_spline4.540.80
TRE_rigid1.961.30
InhaleFRE_rigid7.131.86
TRE_spline2.950.94
TRE_rigid1.441.94
2_leftrightExhaleFRE_rigid4.794.07
TRE_spline0.851.36
TRE_rigid0.60.86
InhaleFRE_rigid4.934.21
TRE_spline0.711.34
TRE_rigid0.680.94
AllExhaleFRE_rigid7.098.58
TRE_spline4.4710.24
TRE_rigid0.471.04
InhaleFRE_rigid7.189.78
TRE_spline4.589.07
TRE_rigid1.081.14
Errors of patient's markers registration for different marker configuration Meier et al. [13, 14] report worse performance of target position estimation for left-right marker configuration for studies in swine. The results obtained generally do not support this rule. Due to big differences between patients (Table VII) it is necessary to treat each patient individually. One possible solution is the initial use of more markers (e.g. 8 to 10) during CT examination and then after nodule localization in the CT preoperative anatomical model it is possible to test different configurations of markers before intervention to select the best one. The final number of proposed markers is limited by the line of sight between markers and the camera, which is distorted by an ultrasound head and ablation needle.
Table VII

Errors of patient's markers registration for median FRE

Error typeMedian value
MinimumMaximumMean
ExhaleInhaleTotalExhaleInhaleTotalExhaleInhaleTotal
FRE_rigid2.273.093.0732.85251.699.3716.7515.56
TRE_spline1.071.421.396.92112.88522.9712.827.21
TRE_rigid0.340.30.291.781.231.690.852.430.82
Errors of patient's markers registration for median FRE Implementation of the automated registration method to clinical practice is easier because of shortening of preparation time in OR, no necessity of touching the patient, and no dependency on the physician's experience. In OR liver tumor ablation is usually performed in general anesthesia, so breathing is more regular than free breathing. As regards the image-guided system for percutaneous liver RF tumor ablation, the problem is to synchronize a respiratory phase for which the static preoperative CT anatomical model was generated. Using a shape conserving mattress, the observation of marker movement and real time FRE calculation helps to select the best respiratory phase for needle insertion. In this application two registration methods of abdominal preoperative CT and physical patient position in OR were presented and compared. This approach is being developed as a step to image-guided percutaneous liver RF tumor ablation. The proposed equation of deformation is calculated in real time. Assuming regular breathing conducted by the respirator, recurrence of patient position and short time between CT study and procedure, the proposed solution could be used to synchronize the respiratory phase for which the static preoperative CT anatomical model was generated, which is one of the most challenging steps in RF ablation. In the absence of a direct correlation between the movements of the abdominal area and movement of internal organs to get more reliable registration results in the next step we want to apply 4D magnetic resonance or computed tomography imaging and build an internal organ motion model to correlate internal and external breathing motion.
  16 in total

1.  Inter-subject modelling of liver deformation during radiation therapy.

Authors:  M von Siebenthal; G Székely; A Lomax; Ph Cattin
Journal:  Med Image Comput Comput Assist Interv       Date:  2007

2.  Navigation of bone alignment osteotomies of the tibia.

Authors:  S Weber; T C Lueth
Journal:  Int J Med Robot       Date:  2005-12       Impact factor: 2.547

3.  Real-time computed tomography image update for endoscopic skull base surgery.

Authors:  B A Woodworth; A G Chiu; N A Cohen; D W Kennedy; B W O'Malley; J N Palmer
Journal:  J Laryngol Otol       Date:  2007-08-16       Impact factor: 1.469

4.  Respiratory motion compensation for CT-guided interventions in the liver.

Authors:  Lena Maier-Hein; Sascha A Müller; Frank Pianka; Stefan Wörz; Beat P Müller-Stich; Alexander Seitel; Karl Rohr; Hans-Peter Meinzer; Bruno M Schmied; Ivo Wolf
Journal:  Comput Aided Surg       Date:  2008-05

5.  Predicting error in rigid-body point-based registration.

Authors:  J M Fitzpatrick; J B West; C R Maurer
Journal:  IEEE Trans Med Imaging       Date:  1998-10       Impact factor: 10.048

6.  Precision targeting of liver lesions using a novel electromagnetic navigation device in physiologic phantom and swine.

Authors:  Filip Banovac; Jonathan Tang; Sheng Xu; David Lindisch; Ho Young Chung; Elliot B Levy; Thomas Chang; Michael F McCullough; Ziv Yaniv; Bradford J Wood; Kevin Cleary
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

7.  Systematic errors in respiratory gating due to intrafraction deformations of the liver.

Authors:  Martin von Siebenthal; Gábor Székely; Antony J Lomax; Philippe C Cattin
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

8.  Continuous intrathecal treatment with methotrexate via subcutaneous port: implication for leptomeningeal dissemination of malignant tumors.

Authors:  Nobusada Shinoura; Yusuke Tabei; Ryozi Yamada; Kuniaki Saito; Masamichi Takahashi
Journal:  J Neurooncol       Date:  2007-12-12       Impact factor: 4.130

Review 9.  Interventional navigation systems for treatment of unresectable liver tumor.

Authors:  Soo Jay Phee; Kai Yang
Journal:  Med Biol Eng Comput       Date:  2009-12-30       Impact factor: 2.602

10.  The image-guided surgery toolkit IGSTK: an open source C++ software toolkit.

Authors:  Andinet Enquobahrie; Patrick Cheng; Kevin Gary; Luis Ibanez; David Gobbi; Frank Lindseth; Ziv Yaniv; Stephen Aylward; Julien Jomier; Kevin Cleary
Journal:  J Digit Imaging       Date:  2007-08-17       Impact factor: 4.056

View more
  2 in total

1.  Image-based guidance of percutaneous abdomen intervention based on markers for semi-automatic rigid registration.

Authors:  Dominik Spinczyk
Journal:  Wideochir Inne Tech Maloinwazyjne       Date:  2014-09-05       Impact factor: 1.195

2.  Robotic stereotactic body radiation therapy for liver-limited malignant tumors.

Authors:  Mariusz Frączek; Jacek Sobocki; Katarzyna Pędziwiatr; Edward Skrocki; Norbert Piotrkowicz; Dobromira Tyc-Szczepaniak; Elżbieta Korab-Chrzanowska; Piotr Hevelke; Maciej Krasnodębski; Waldemar Koszewski
Journal:  Wideochir Inne Tech Maloinwazyjne       Date:  2014-07-23       Impact factor: 1.195

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.