PURPOSE: To assess the safety, efficacy and performance of the RELAY NBS PLUS stent-graft in patients with aortic arch pathology.
METHODS: From July 2010 to December 2011, the RELAY NBS PLUS was used to treat 13 patients (eight males; mean age 59.8 years, range 29-78 years) suffering from aortic arch pathology. The distribution of the proximal landing zone was Zone 0 in one case, Zone 1 in three cases and Zone 2 in nine cases. Bypass of the cervical branches was performed in 11 patients.
RESULTS: The surgery-related mortality within 30 days and postoperative morbidity were 0%. Postoperative computed tomography at discharge confirmed 100% clinical success without any endoleaks. The mean radius of the inner curve in the aortic arch was 16.2 mm (range 5.8-25.7 mm). We detected two bird-beak configurations with a severely angulated aortic arch with a radius of the inner curve <15 mm. At the mid-term follow-up (mean 22.1 months; range 13.8-28.2 months), all patients were alive without aorta- or device-related complications.
CONCLUSIONS: RELAY NBS PLUS stent-grafts can be safely and effectively implanted with controlled deployment to satisfactorily repair aortic arch pathology, even in the cases considered to present with an unfavorable anatomy for the first generation commercially available stent-graft devices. These encouraging outcomes will need to be confirmed in a larger series with a longer follow-up.

Entities:  

Mesh:

Year:  2014        PMID: 24496982     DOI: 10.1007/s00595-014-0851-2

Source DB:  PubMed          Journal:  Surg Today        ISSN: 0941-1291            Impact factor:   2.549


  14 in total

1.  New operative method for acute type B dissection: left carotid artery-left subclavian artery bypass combined with endovascular stent-graft implantation.

Authors:  N Shigemura; M Kato; T Kuratani; Y Funakoshi; M Kaneko
Journal:  J Thorac Cardiovasc Surg       Date:  2000-08       Impact factor: 5.209

2.  Relay NBS graft with the plus delivery system to improve deployment in aortic Arch with small radius curve.

Authors:  Carlo Ferro; Umberto G Rossi; Sara Seitun; Andrea Guastavino; Flavio Scarano; Gian Carlo Passerone
Journal:  Cardiovasc Intervent Radiol       Date:  2010-04-16       Impact factor: 2.740

3.  Endovascular stent grafting versus open surgical repair of descending thoracic aortic aneurysms in low-risk patients: a multicenter comparative trial.

Authors:  Joseph E Bavaria; Jehangir J Appoo; Michel S Makaroun; Joel Verter; Zi-Fan Yu; R Scott Mitchell
Journal:  J Thorac Cardiovasc Surg       Date:  2007-01-08       Impact factor: 5.209

4.  Final operative and midterm results of the European experience in the RELAY Endovascular Registry for Thoracic Disease (RESTORE) study.

Authors:  Vincent Riambau; Burkhart Zipfel; Gioacchino Coppi; Martin Czerny; Domenico G Tealdi; Carlo Ferro; Roberto Chiesa; Carlo Sassi; Herve Rousseau; Sergio Berti
Journal:  J Vasc Surg       Date:  2010-11-26       Impact factor: 4.268

Review 5.  Current strategy of endovascular aortic repair for thoracic aortic aneurysms.

Authors:  Toru Kuratani; Yoshiki Sawa
Journal:  Gen Thorac Cardiovasc Surg       Date:  2010-08-12

6.  Endovascular repair of descending thoracic aneurysms: results with "on-label" application in the post Food and Drug Administration approval era.

Authors:  G Chad Hughes; Sean M Lee; Mani A Daneshmand; Syamal D Bhattacharya; Judson B Williams; Sonny W Tucker; Richard L McCann
Journal:  Ann Thorac Surg       Date:  2010-07       Impact factor: 4.330

7.  Arch and visceral/renal debranching combined with endovascular repair for thoracic and thoracoabdominal aortic aneurysms.

Authors:  Sung Wan Ham; Terry Chong; John Moos; Vincent L Rowe; Robbin G Cohen; Mark J Cunningham; Alison Wilcox; Fred A Weaver
Journal:  J Vasc Surg       Date:  2011-02-21       Impact factor: 4.268

8.  Thoracic endovascular repair (TEVAR) in the management of aortic arch pathology.

Authors:  Erin H Murphy; Gregory A Stanley; Mihaiela Ilves; Martyn Knowles; J Michael Dimaio; Michael E Jessen; Frank R Arko
Journal:  Ann Vasc Surg       Date:  2012-01       Impact factor: 1.466

9.  Incomplete endograft apposition to the aortic arch: bird-beak configuration increases risk of endoleak formation after thoracic endovascular aortic repair.

Authors:  Takuya Ueda; Dominik Fleischmann; Michael D Dake; Geoffrey D Rubin; Daniel Y Sze
Journal:  Radiology       Date:  2010-05       Impact factor: 11.105

10.  Five-year results of endovascular treatment with the Gore TAG device compared with open repair of thoracic aortic aneurysms.

Authors:  Michel S Makaroun; Ellen D Dillavou; Grayson H Wheatley; Richard P Cambria
Journal:  J Vasc Surg       Date:  2008-03-19       Impact factor: 4.268

View more
  2 in total

Review 1.  Application of the Bolton Relay Device for Thoracic Endografting In or Near the Aortic Arch.

Authors:  Vincent Riambau
Journal:  Aorta (Stamford)       Date:  2015-02-01

2.  Strategy for thoracic endovascular aortic repair based on collateral circulation to the artery of Adamkiewicz.

Authors:  Takeshi Kamada; Kunihiro Yoshioka; Ryoichi Tanaka; Shinji Makita; Akihiko Abiko; Masayuki Mukaida; Akio Ikai; Hitoshi Okabayashi
Journal:  Surg Today       Date:  2015-11-13       Impact factor: 2.549

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.