Osseointegration of machined, injection moulded and oxygen plasma modified PEEK implants in a sheep model.

Alexandra H C Poulsson1, David Eglin2, Stefan Zeiter2, Karin Camenisch2, Christoph Sprecher2, Yash Agarwal2, Dirk Nehrbass2, Joanne Wilson3, Robert G Richards4.   

Abstract

Machined and injection moulded polyetheretherketone (PEEK) implants with and without an oxygen plasma modification were prepared and implanted in sheep cancellous and cortical bone. After 4, 12 and 26 weeks, osseointegration was evaluated through mechanical push-out tests and histomorphometry. In the cancellous bone, push-out force increased with time, a trend toward higher force was observed for machined compared to moulded, and oxygen plasma modified compared to unmodified. On-going remodelling of the bone was detected in the periphery of the implants at 4 weeks. Minimal or no inflammation was observed with all the implants at all locations and time-points. Bone-implant contact (BIC) was quantified at all-time points and locations for all the four PEEK implant surfaces. The BIC values ranged from 15 to 75% with an average of 29 ± 13% in the cancellous bone and 25-65% with an average of 50 ± 12% in the cortical bone. In the cortical bone the BIC increased significantly from 4 to 26 weeks. This in vivo study has identified that surface topography of PEEK implants influences osseointegration. In addition, oxygen plasma has the potential to increase bone-implant interface stability. This study provides a unique reference for further modifications and in vivo assessment of PEEK implants.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Animal model; Biocompatibility; Histomorphometry; Osseointegration; Plasma; Polyetheretherketone (PEEK)

Mesh:

Substances:

Year:  2014        PMID: 24485795     DOI: 10.1016/j.biomaterials.2013.12.056

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  20 in total

1.  Getting PEEK to Stick to Bone: The Development of Porous PEEK for Interbody Fusion Devices.

Authors:  F Brennan Torstrick; David L Safranski; J Kenneth Burkus; James L Chappuis; Christopher S D Lee; Robert E Guldberg; Ken Gall; Kathryn E Smith
Journal:  Tech Orthop       Date:  2017-09-01

2.  Do Surface Porosity and Pore Size Influence Mechanical Properties and Cellular Response to PEEK?

Authors:  F Brennan Torstrick; Nathan T Evans; Hazel Y Stevens; Ken Gall; Robert E Guldberg
Journal:  Clin Orthop Relat Res       Date:  2016-11       Impact factor: 4.176

3.  High-strength, surface-porous polyether-ether-ketone for load-bearing orthopedic implants.

Authors:  Nathan T Evans; F Brennan Torstrick; Christopher S D Lee; Kenneth M Dupont; David L Safranski; W Allen Chang; Annie E Macedo; Angela S P Lin; Jennifer M Boothby; Daniel C Whittingslow; Robert A Carson; Robert E Guldberg; Ken Gall
Journal:  Acta Biomater       Date:  2014-11-24       Impact factor: 8.947

4.  Assessment of a polyelectrolyte multilayer film coating loaded with BMP-2 on titanium and PEEK implants in the rabbit femoral condyle.

Authors:  R Guillot; I Pignot-Paintrand; J Lavaud; A Decambron; E Bourgeois; V Josserand; D Logeart-Avramoglou; E Viguier; C Picart
Journal:  Acta Biomater       Date:  2016-03-07       Impact factor: 8.947

5.  In vitro evaluation of bioactivity of chemically deposited hydroxyapatite on polyether ether ketone.

Authors:  D Almasi; S Izman; M Sadeghi; N Iqbal; F Roozbahani; G Krishnamurithy; T Kamarul; M R Abdul Kadir
Journal:  Int J Biomater       Date:  2015-03-09

6.  Implant materials generate different peri-implant inflammatory factors: poly-ether-ether-ketone promotes fibrosis and microtextured titanium promotes osteogenic factors.

Authors:  Rene Olivares-Navarrete; Sharon L Hyzy; Paul J Slosar; Jennifer M Schneider; Zvi Schwartz; Barbara D Boyan
Journal:  Spine (Phila Pa 1976)       Date:  2015-03-15       Impact factor: 3.468

Review 7.  PEEK with Reinforced Materials and Modifications for Dental Implant Applications.

Authors:  Fitria Rahmitasari; Yuichi Ishida; Kosuke Kurahashi; Takashi Matsuda; Megumi Watanabe; Tetsuo Ichikawa
Journal:  Dent J (Basel)       Date:  2017-12-15

8.  Evaluation of the Grafting Efficacy of Active Biomolecules of Phosphatidylcholine and Type I Collagen on Polyether Ether Ketone: In Vitro and In Vivo.

Authors:  Jian-Chih Chen; Chih-Hua Chen; Kai-Chi Chang; Shih-Ming Liu; Chia-Ling Ko; Chi-Jen Shih; Ying-Sui Sun; Wen-Cheng Chen
Journal:  Polymers (Basel)       Date:  2021-06-24       Impact factor: 4.329

Review 9.  Nanomodified Peek Dental Implants: Bioactive Composites and Surface Modification-A Review.

Authors:  Shariq Najeeb; Zohaib Khurshid; Jukka Pekka Matinlinna; Fahad Siddiqui; Mohammad Zakaria Nassani; Kusai Baroudi
Journal:  Int J Dent       Date:  2015-10-01

Review 10.  Preparation Methods for Improving PEEK's Bioactivity for Orthopedic and Dental Application: A Review.

Authors:  Davood Almasi; Nida Iqbal; Maliheh Sadeghi; Izman Sudin; Mohammed Rafiq Abdul Kadir; Tunku Kamarul
Journal:  Int J Biomater       Date:  2016-04-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.