A note on arterial to venous oxygen saturation as reference for NIRS-determined frontal lobe oxygen saturation in healthy humans.

Henrik Sørensen1, Niels H Secher1, Peter Rasmussen1.   

Abstract

Entities:  

Keywords:  NIRS; cerebral oxygenation; humans; hypercapnia; hypocapnia; hypoxia; jugular vein

Year:  2014        PMID: 24478709      PMCID: PMC3897873          DOI: 10.3389/fphys.2013.00403

Source DB:  PubMed          Journal:  Front Physiol        ISSN: 1664-042X            Impact factor:   4.566


× No keyword cloud information.
Near infrared spectroscopy (NIRS) offers non-invasive assessment of oxygenation within the human brain (ScO2) by appreciating the different absorption of near infrared light by hemoglobin and oxyhemoglobin (Jobsis, 1977). Since the length the light has passed when traveling from the skin to the cortex and then returning to the skin remains unknown, there is a need to adjust the signal according to an assumed ratio between the arterial vs. venous blood that is appreciated. Apparently, most NIRS devices use a fixed reference ratio between the arterial (25%) and venous contribution (75%) to the signal despite not eligible mismatch (Bickler et al., 2013). This assumption is based on anatomical evidence (Pollard et al., 1996), but might be confounded by changes in cerebral blood volume during, e.g., hypoxia and changes in the arterial carbon dioxide tension (PaCO2) (Ito et al., 2005). A standard reference arterial to venous ratio may therefore not exist for application of NIRS to determine ScO2 in humans. Thus, an estimate of cerebral capillary hemoglobin oxygen saturation (ScapO2) to express cerebral oxygenation is based on 50% jugular and arterial saturations (Gjedde et al., 2005) and has been reported to follow changes in ScO2 (Rasmussen et al., 2007). In this report we made a meta-analysis on published data (Rasmussen et al., 2007; Sørensen et al., 2012, in press) in order to evaluate which ratio between arterial and internal jugular venous hemoglobin saturation that fits best to the concomitant determined ScO2 in healthy humans exposed to a wide range of interventions. Thirty seven subjects [age 27(9) years, height 181(10) cm, mass 79(13) kg, mean (SD)] were catheterized in the right internal jugular vein with the tip of the catheter advanced to the bulb of the vein and with a catheter in the brachial artery of the non-dominant arm, while ScO2 was monitored by the Invos Cerebral Oximetry (Covidien, Mansfield, MI). Arterial partial pressure for oxygen and carbon dioxide, and oxygen saturations were measured in the jugular and arterial blood (PaO2; PaCO2; SaO2; SjO2) (ABL-800, Radiometer, Brønshøj, Denmark). The subjects were exposed to hypoxia (FiO2 = 10%; n = 23), inspiration of 100% oxygen (n = 8), atmospheric air (n = 37), hypercapnia (FiCO2 = 5%; n = 8) and asked to hyperventilate (~2–3 kPa reduction in PaCO2; n = 32) with separate controls. By linear regression the contribution from arterial and jugular blood to ScO2 was estimated and R-squared (R2) and root mean square error (RMSE) between ScO2 and the arterial fraction in the reference saturation were calculated (SAS Institute Inc., Cary, NC). Only data points where SjO2 ≤ ScO2SaO2 were included. All reference saturations were calculated, e.g., ScapO2 = 0.50 · SaO2 + 0.50 · SjO2. The following equation, 0 = ScO2 – [a · SaO2 + (a − 1) · SjO2], was used to calculate the arterial fraction (a) for which the difference between ScO2 and the reference saturation was zero. SaO2 and SjO2 ranged from 70 to 100% and 33 to 87%, respectively, while ScO2 ranged from 48 to 95%. PaCO2 was manipulated from 1.6 to 6.3 kPa. According to the linear regression analysis, ScO2 demonstrated a correlation to a wide range of ratios between the arterial and venous hemoglobin saturations. The highest RMSE was obtained when it was considered that there was no arterial contribution to ScO2 and the RMSE became gradually lower when the arterial contribution was considered to increase (Figure 1A). The lowest RMSE was observed for a 75% arterial and 25% jugular venous blood contribution to ScO2(RMSE = 2.70; R2 = 0.644; P < 0.0001; Figure 1A). For the often-used calibration ratio (25% arterial) R2 was 0.505 (P < 0.0001) with a RMSE of 4.233. In contrast, ScapO2 (50% arterial) had a R2 of 0.606 (P < 0.0001; RMSE = 3.276; Figure 1A). Zero was within the 95% confidence interval only with a calculated 40–50% arterial contribution to the reference ratio. When ScO2 was compared with the calibration ratio, the mean difference was zero in 2.9% of the blood samples, whereas it was 10.5% when ScapO2 was used as reference (Figure 1B).
Figure 1

(A) Root mean square error for the arterial fraction determined by the linear regression analysis. Values are mean (95% confidence interval). (B) Zero difference between the Invos-determined cerebral oxygenation (ScO2) and the calculated reference saturation related to the considered arterial contribution for 171 blood samples; average arterial contribution 46 ± 17% (SD). Solid line represents the mean distribution

(A) Root mean square error for the arterial fraction determined by the linear regression analysis. Values are mean (95% confidence interval). (B) Zero difference between the Invos-determined cerebral oxygenation (ScO2) and the calculated reference saturation related to the considered arterial contribution for 171 blood samples; average arterial contribution 46 ± 17% (SD). Solid line represents the mean distribution This meta-analysis of published data suggests that the optimal reference ratio has a larger arterial contribution than the ratio defined by anatomical models and likely incorporated in most NIRS devices (Pollard et al., 1996; Bickler et al., 2013). The subjects were not only exposed to conditions that changes the venous oxygen saturation, i.e., hyperventilation, but also to conditions known to affect arterial and venous cerebral blood volume and oxygen content without influencing extra-cerebral blood (Ito et al., 2005). From these interventions, linear regression analysis demonstrated that the correlation between ScO2 and the arterial contribution to the reference saturation was highest when a 75% arterial contribution was considered (R2 = 0.644) and also for that ratio the RMSE was the lowest (2.70) (Figure 1A). The calibration ratio (25% arterial) typically incorporated in the algorithms had a R2 of 0.505 and a RMSE of 4.233 while ScapO2 demonstrated a stronger correlation with a R2 of 0.606 (P < 0.0001; RMSE = 3.276). In addition, the mismatch between ScO2 and the reference ratio was more likely to be zero when a more arterial weighted reference was applied (Figure 1B). Thus, our findings demonstrate that ScapO2 is an accurate reference for ScO2 at least when determined with the Invos apparatus. Similar to these findings, no eligible mismatch between ScO2 values and the calibration ratio (25% arterial) is observed in five different NIRS devices used to determine cerebral oxygenation in healthy subjects exposed to isocapnic hypoxemia (Bickler et al., 2013). Hypercapnia and isocapnic hypoxemia induce cerebral vasodilation and affect cerebral blood volume mainly by an increase in the arterial fraction (Ito et al., 2005). Thus, it is likely that the illuminated area of the brain encompasses more arterial blood and that could explain the aggravated mismatch between ScO2 and the calibration ratio during hypoxia (Bickler et al., 2013). Cerebral blood flow and jugular venous oxygen saturation are decreased with hyperventilation that reduce the calibration ratio more compared to ScapO2. Interestingly, when healthy humans hyperventilate the ScO2 overestimated the calibration ratio by 11.2% while the mean bias was only 0.2% for ScapO2 (Sørensen et al. unpublished). Thus, the mismatch between the calibration ratio and ScO2 is aggravated when only jugular venous oxygen saturation is altered, which indicates that ScO2 accounts for more than 25% arterial blood and that evaluations of reference saturations for NIRS must also involve conditions known to affect arterial oxygen content. The arterial to venous balance within the brain may differ between individuals and explains the inter-individual variation in absolute ScO2 readings (Rasmussen et al., 2007; Bickler et al., 2013), especially the heterogeneity of blood vessels in the illuminated area of the brain seems to affect light absorption because the photons are “lost” in major blood vessel, e.g., the sagittal sinuses (Kishi et al., 2003). Other factors affecting light absorption and thereby the ScO2 readings may be variation in skull thickness and amount of cerebrospinal fluid (Yoshitani et al., 2007; Strangman et al., 2014). Also, skin pigmentation (Bickler et al., 2013) and degradation products of heme can affect ScO2 because of competitive absorption of light (Madsen et al., 2000). Despite absolute ScO2 values are not comparable and exhibit large variations when compared to the calibration ratio, NIRS offers a unique non-invasive method for assessment of cerebral oxygen delivery vs. consumption and its clinical utility relies on changes from baseline rather than on absolute values (Murkin et al., 2007). Yet, at present the NIRS technology for evaluation of ScO2 is limited when skin blood flow is affected either by scalp ischemia or administration of sympathomimetic agents that affect skin blood flow (Davie and Grocott, 2012; Sørensen et al., 2012). In summary, this report suggests that the reference saturations applied when using Invos cerebral oximetry should be weighted more to arterial hemoglobin saturation than accepted by anatomical models.

Author contributions

All authors contributed equally to the design, data analysis, and interpretation, drafting the manuscript and critical revision. All authors approved the final version before submission.
  14 in total

1.  Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies.

Authors:  Sophie N Davie; Hilary P Grocott
Journal:  Anesthesiology       Date:  2012-04       Impact factor: 7.892

2.  Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers.

Authors:  Philip E Bickler; John R Feiner; Mark D Rollins
Journal:  Anesth Analg       Date:  2013-09-10       Impact factor: 5.108

3.  Cerebral metabolic response to low blood flow: possible role of cytochrome oxidase inhibition.

Authors:  Albert Gjedde; Peter Johannsen; Georg E Cold; Leif Ostergaard
Journal:  J Cereb Blood Flow Metab       Date:  2005-09       Impact factor: 6.200

4.  Interference of cerebral near-infrared oximetry in patients with icterus.

Authors:  P L Madsen; C Skak; A Rasmussen; N H Secher
Journal:  Anesth Analg       Date:  2000-02       Impact factor: 5.108

5.  Cutaneous vasoconstriction affects near-infrared spectroscopy determined cerebral oxygen saturation during administration of norepinephrine.

Authors:  Henrik Sørensen; Niels H Secher; Christoph Siebenmann; Henning B Nielsen; Matthias Kohl-Bareis; Carsten Lundby; Peter Rasmussen
Journal:  Anesthesiology       Date:  2012-08       Impact factor: 7.892

6.  External carotid artery flow maintains near infrared spectroscopy-determined frontal lobe oxygenation during ephedrine administration.

Authors:  H Sørensen; P Rasmussen; K Sato; S Persson; N D Olesen; H B Nielsen; N V Olsen; S Ogoh; N H Secher
Journal:  Br J Anaesth       Date:  2014-02-06       Impact factor: 9.166

7.  Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study.

Authors:  John M Murkin; Sandra J Adams; Richard J Novick; Mackenzie Quantz; Daniel Bainbridge; Ivan Iglesias; Andrew Cleland; Betsy Schaefer; Beverly Irwin; Stephanie Fox
Journal:  Anesth Analg       Date:  2007-01       Impact factor: 5.108

8.  Validation in volunteers of a near-infrared spectroscope for monitoring brain oxygenation in vivo.

Authors:  V Pollard; D S Prough; A E DeMelo; D J Deyo; T Uchida; H F Stoddart
Journal:  Anesth Analg       Date:  1996-02       Impact factor: 5.108

9.  Influence of patient variables and sensor location on regional cerebral oxygen saturation measured by INVOS 4100 near-infrared spectrophotometers.

Authors:  Katsuyoshi Kishi; Masahiko Kawaguchi; Kenji Yoshitani; Toshihiro Nagahata; Hitoshi Furuya
Journal:  J Neurosurg Anesthesiol       Date:  2003-10       Impact factor: 3.956

10.  Scalp and skull influence on near infrared photon propagation in the Colin27 brain template.

Authors:  Gary E Strangman; Quan Zhang; Zhi Li
Journal:  Neuroimage       Date:  2013-05-07       Impact factor: 6.556

View more
  7 in total

1.  In response to: phenylephrine and paradoxically increased muscle tissue oxygenation: is the mechanism related to local venoconstriction or augmented venous return?

Authors:  H Sørensen; A H Egesborg; J H Thomsen; A S P Meyer; D Terzic; L Hilsted; J Kjærgaard; J P Goetze; T C Barbosa; N H Secher
Journal:  J Clin Monit Comput       Date:  2018-01-31       Impact factor: 2.502

2.  Comparison of tissue oximeters on a liquid phantom with adjustable optical properties: an extension.

Authors:  S Kleiser; D Ostojic; B Andresen; N Nasseri; H Isler; F Scholkmann; T Karen; G Greisen; M Wolf
Journal:  Biomed Opt Express       Date:  2017-12-05       Impact factor: 3.732

3.  Cerebral oxygenation in health and disease.

Authors:  Patrice Brassard; Philip N Ainslie; Niels H Secher
Journal:  Front Physiol       Date:  2014-11-24       Impact factor: 4.566

Review 4.  Journal of clinical monitoring and computing 2016 end of year summary: monitoring cerebral oxygenation and autoregulation.

Authors:  Thomas W L Scheeren; Bernd Saugel
Journal:  J Clin Monit Comput       Date:  2017-01-24       Impact factor: 2.502

5.  Effects of Arterial Carbon Dioxide Tension on Cerebral and Somatic Regional Tissue Oxygenation and Blood Flow in Neonates After the Norwood Procedure With Deep Hypothermic Cardiopulmonary Bypass.

Authors:  George M Hoffman; John P Scott; Eckehard A Stuth
Journal:  Front Pediatr       Date:  2022-02-11       Impact factor: 3.418

6.  Comparing near-infrared spectroscopy-measured cerebral oxygen saturation and corresponding venous oxygen saturations in children with congenital heart disease: a systematic review and meta-analysis.

Authors:  Yiqi Ma; Lihong Zhao; Jiafu Wei; Ziwei Wang; Su Lui; Bin Song; Qiyong Gong; Pu Wang; Min Wu
Journal:  Transl Pediatr       Date:  2022-08

7.  Alteration in Brain Functional and Effective Connectivity in Subjects With Hypertension.

Authors:  Lingguo Bu; Congcong Huo; Gongcheng Xu; Ying Liu; Zengyong Li; Yubo Fan; Jianfeng Li
Journal:  Front Physiol       Date:  2018-05-31       Impact factor: 4.566

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.