Abstract
Transcranial Direct Current Stimulation (tDCS) is a neuromodulatory device often publicized for its ability to enhance cognitive and behavioral performance. These enhancement claims, however, are predicated upon electrophysiological evidence and descriptions which are far from conclusive. In fact, a review of the literature reveals a number of important experimental and technical issues inherent with this device that are simply not being discussed in any meaningful manner. In this paper, we will consider five of these topics. The first, inter-subject variability, explores the extensive between- and within-group differences found within the tDCS literature and highlights the need to properly examine stimulatory response at the individual level. The second, intra-subject reliability, reviews the lack of data concerning tDCS response reliability over time and emphasizes the importance of this knowledge for appropriate stimulatory application. The third, sham stimulation and blinding, draws attention to the importance (yet relative lack) of proper control and blinding practices in the tDCS literature. The fourth, motor and cognitive interference, highlights the often overlooked body of research that suggests typical behaviors and cognitions undertaken during or following tDCS can impair or abolish the effects of stimulation. Finally, the fifth, electric current influences, underscores several largely ignored variables (such as hair thickness and electrode attachments methods) influential to tDCS electric current density and flow. Through this paper, we hope to increase awareness and start an ongoing dialog of these important issues which speak to the efficacy, reliability, and mechanistic foundations of tDCS.Entities:
Keywords: efficacy; mechanisms of action; reliability; transcranial direct current stimulation (tDCS); variability
Year: 2014 PMID: 24478640 PMCID: PMC3901383 DOI: 10.3389/fnsys.2014.00002
Source DB: PubMed Journal: Front Syst Neurosci ISSN: 1662-5137
Studies exploring the effects of 0.0286 mA/cm.
| Study | tDCS duration | Anode | Cathode | Control | |
|---|---|---|---|---|---|
| Nitsche and Paulus, | 10 and 9 ( | 4 s ( | |||
| Nitsche and Paulus, | 12 ( | 5, 7, 9, 11, and 13 min ( | |||
| Liebetanz et al., | 11 | 5 min | |||
| Nitsche et al., | 12 ( | 4 s ( | |||
| Nitsche et al., | 12 ( | 5, 7, and 9 min ( | |||
| Lang et al., | 8 | 10 min | |||
| Lang et al., | 5 and 10 ( | 10 min ( | |||
| Siebner et al., | 5 and 8 ( | 10 min ( | |||
| Nitsche et al., | 6 ( | 4 s ( | |||
| Nitsche et al., | 12, 9, and 10 ( | 4 s ( | |||
| Nitsche et al., | 12 | 9 min C/13 min A | |||
| Quartarone et al., | 7 and 21 ( | 5 min ( | |||
| Quartarone et al., | 8 | 10 min | |||
| Nitsche et al., | 12 | 9 min C/13 min A | |||
| Power et al., | 10 | 10 min | |||
| Nitsche et al., | 12 ( | 4 s ( | |||
| Nitsche et al., | 12 | 7 min | |||
| Kuo et al., | 7 | 9 min C/13 min A | |||
| Antal et al., | 12 | 10 min | |||
| Boros et al., | 17 | 13 min | |||
| Kuo et al., | 7 | 9 min C/13 min A | |||
| Nitsche et al., | 12 | 9 min C/13 min A | |||
| Monte-Silva et al., | 12 ( | 9 min C/13 min A ( | |||
| Monte-Silva et al., | 12 ( | 9 and 18 min ( | |||
| Bradnam et al., | 18 | 15 min | |||
| Fricke et al., | 8–12 ( | 5 min ( | |||
| List et al., | 12 | 10 min | |||
| McCambridge et al., | 7 Active/5 Sham | 10 min | |||
| Munneke et al., | 10 ( | 7, 11, and 15 min ( | |||
| Scelzo et al., | 12 | 13 min | |||
| Thirugnanasambandam et al., | 16 | 20 min | |||
| Di Lazzaro et al., | 30 | 20 min | |||
| Hasan et al., | 18 | 9 min | |||
| Schade et al., | 8 ( | 5 min ( | |||
| Suzuki et al., | 9 | 10 min | |||
| Monte-Silva et al., | 15 ( | 13 and 26 min ( | |||
| Hasan et al., | 20 | 9 min | |||
| Batsikadze et al., | 9 and 8 ( | 20 min ( | |||
| Schabrun et al., | 21 A, 9 C, 13 s | 20 min ( | |||
| Simis et al., | 11 | 20 min | |||
| TOTAL | 62 | 67 | 10 | ||
A sham condition from Quartarone et al. (2004) was excluded due to reporting MEP modulation during motor imagery only (not at rest). One study from Monte-Silva et al., 2009 was excluded due to presenting replication data. A, Anode; C, Cathode; S, Sham.
Studies include the placebo condition/s in any drug or device interaction studies.
Figure 1(A) Under typical conditions, one will be able to build an electric circuit using two saline soaked tDCS sponges placed on a piece of clean skin. (B) When a piece of non-conductive plastic is placed underneath one electrode, the circuit will be broken. (C) When a small hole is cut in the plastic and a small stream of saline is used to connect the skin under the hole to the electrode, the circuit will be re-built. This demonstrates that, even if a tDCS electrode is not in contact with the scalp, excessive saline can be used to bridge between the electrode and the skin. Unfortunately, with excessive saline, determining the location of the circuit connection and electrical density at this point is extremely difficult.