The analytical performance evaluation of Freelite™ Human Kappa Free and Human Lambda Free on the SPAPLUS™ immunoturbidimetric analyzer.

Kyong-Ho Cha1, Yang Bo Sim, Hyojin Chae, Hae-il Park, Myungshin Kim, Yonggoo Kim.   

Abstract

BACKGROUND: SPAPLUS™ is a turbidimetric immunoassay analyzer for detection of excess free light chain (FLC) antigens in serum. Here, we evaluated the analytical performance of Freelite™ Human Kappa Free and Lambda Free on a SPAPLUS™ instrument.
METHODS: We evaluated the precision, linearity, sample carryover, and drift of the SPAPLUS™ instrument and compared it with Hitachi 7600 and BN™ II instruments. We evaluated the detection of antigen excess for 12 specimens from patients with monoclonal gammopathy.
RESULTS: The coefficients of variations of κFLC and λFLC were below 5.0%. Linearity was shown in the range of 9.68-152.25 mg/l for κFLC and 4.96-171.09 mg/l for λFLC, and no drift was observed. The κFLC sample carryover was statistically significant, but much smaller than the optimum allowable bias. Agreement rates with the two comparative methods were 87.1, 87.1, and 97.1% or higher for κFLC, λFLC, and the κ/λ ratio, respectively. Antigen excess signals were observed for all 12 antigen excess specimens.
CONCLUSIONS: The Freelite™ on the SPAPLUS™ exhibited appropriate precision, linearity, and relative comparability to the reagents on the other instruments. It was good at detecting specimens that had previously demonstrated the hook effect due to antigen excess.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  analytical performance; evaluation; free light chain; immunoassay; turbidimetry

Mesh:

Substances:

Year:  2014        PMID: 24478145      PMCID: PMC6807592          DOI: 10.1002/jcla.21671

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  21 in total

1.  Highly sensitive, automated immunoassay for immunoglobulin free light chains in serum and urine.

Authors:  A R Bradwell; H D Carr-Smith; G P Mead; L X Tang; P J Showell; M T Drayson; R Drew
Journal:  Clin Chem       Date:  2001-04       Impact factor: 8.327

2.  Hook effect with lambda free light chain in serum free light chain assay.

Authors:  Stanley S Levinson
Journal:  Clin Chim Acta       Date:  2010-07-25       Impact factor: 3.786

3.  Overestimation of serum kappa free light chain concentration by immunonephelometry.

Authors:  Corrie M de Kat Angelino; Reinier Raymakers; Maria A Teunesen; Joannes F M Jacobs; Ina S Klasen
Journal:  Clin Chem       Date:  2010-04-22       Impact factor: 8.327

4.  Analytical performance of serum free light-chain assay during monitoring of patients with monoclonal light-chain diseases.

Authors:  Jillian R Tate; Peter Mollee; Goce Dimeski; Andrew C Carter; Devinder Gill
Journal:  Clin Chim Acta       Date:  2006-07-14       Impact factor: 3.786

5.  Effect of sample dilution on two free light chain nephelometric assays.

Authors:  Joannes F M Jacobs; Rein M J Hoedemakers; Elisa Teunissen; Renate G van der Molen; Henk te Velthuis
Journal:  Clin Chim Acta       Date:  2012-05-03       Impact factor: 3.786

6.  Regarding the overestimation of serum κ free light chains.

Authors:  Stanley S Levinson
Journal:  Clin Chem       Date:  2011-02-08       Impact factor: 8.327

7.  Effect of sample dilution on serum free light chain concentration by immunonephelometric assay.

Authors:  Martine Vercammen; Patricia Meirlaen; Linda Broodtaerts; Isabelle Vande Broek; Xavier Bossuyt
Journal:  Clin Chim Acta       Date:  2011-06-29       Impact factor: 3.786

8.  Clinical and preclinical validation of the serum free light chain assay: identification of the critical difference for optimized clinical use.

Authors:  Charlotte T Hansen; Anna-Marie Münster; Lars Nielsen; Per Pedersen; Niels Abildgaard
Journal:  Eur J Haematol       Date:  2012-10-16       Impact factor: 2.997

9.  Immunoglobulin free light chain assay using latex agglutination.

Authors:  K Wakasugi; H Suzuki; A Imai; S Konishi; H Kishioka
Journal:  Int J Clin Lab Res       Date:  1995

10.  A case of hook effect in the serum free light chain assay using the Olympus AU400e.

Authors:  Christopher R McCudden; Peter M Voorhees; Catherine A Hammett-Stabler
Journal:  Clin Biochem       Date:  2008-10-25       Impact factor: 3.281

View more
  1 in total

1.  Effect of specimen type on free immunoglobulin light chains analysis on the Roche Diagnostics cobas 8000 analyzer.

Authors:  Louis S Nelson; Bryan Steussy; Cory S Morris; Matthew D Krasowski
Journal:  Springerplus       Date:  2015-12-08
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.