Initial experience in Japan with HeartWare ventricular assist system.

Hiroyuki Nishi1, Koichi Toda, Shigeru Miyagawa, Yasushi Yoshikawa, Satsuki Fukushima, Shunsuke Saito, Daisuke Yoshioka, Tetsuya Saito, Takayoshi Ueno, Taichi Sakaguchi, Yoshiki Sawa.   

Abstract

We describe the first clinical experiences in Japan with the HeartWare ventricular assist device (HVAD: HeartWare Inc., Miami Lakes, FL, USA) in patients awaiting heart transplantation. Nine patients (6 males, 3 females; mean 33.5 ± 7.8 years; New York Heart Association class III or IV) received the HVAD as a bridge to transplantation between 2011 and 2012. Six had dilated cardiomyopathy, 2 secondary cardiomyopathy, and 1 dilated phase hypertrophic cardiomyopathy. All operations were uneventful, with a mean operation time of 269 ± 77 min and cardiopulmonary bypass time of 121 ± 40 min. One required a temporary right ventricular assist device and was weaned on postoperative day 20, while another required pump exchange due to foreign tissue in the inflow. Mean support duration was 245 ± 162 days (range: 50-535 days) and mean pump blood flow at 1 month postoperatively was 4.8 ± 0.8 l/min. There was no mortality after 30 days, though 1 patient died during support due to cerebral hemorrhage. Presently, the others are waiting for heart transplantation without problems, except 1 who suffered from an active infection. There was no pump mechanical failure in any case. The HeartWare pump enables quick implantation with acceptable morbidity and mortality. Our preliminary results indicate that this left ventricular assist device is safe for circulatory assistance for heart transplant candidates in Japan.

Entities:  

Mesh:

Year:  2014        PMID: 24464393     DOI: 10.1007/s10047-013-0753-x

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  17 in total

1.  Long-term use of a left ventricular assist device for end-stage heart failure.

Authors:  E A Rose; A C Gelijns; A J Moskowitz; D F Heitjan; L W Stevenson; W Dembitsky; J W Long; D D Ascheim; A R Tierney; R G Levitan; J T Watson; P Meier; N S Ronan; P A Shapiro; R M Lazar; L W Miller; L Gupta; O H Frazier; P Desvigne-Nickens; M C Oz; V L Poirier
Journal:  N Engl J Med       Date:  2001-11-15       Impact factor: 91.245

2.  Initial clinical experience with a novel left ventricular assist device with a magnetically levitated rotor in a multi-institutional trial.

Authors:  Georg M Wieselthaler; Gerry O Driscoll; Paul Jansz; Asghar Khaghani; Martin Strueber
Journal:  J Heart Lung Transplant       Date:  2010-06-19       Impact factor: 10.247

Review 3.  Mechanical circulatory support devices (MCSD) in Japan: current status and future directions.

Authors:  Setsuo Takatani; Hikaru Matsuda; Akihisa Hanatani; Chisato Nojiri; Kenji Yamazaki; Tadashi Motomura; Katsuhiro Ohuchi; Tohru Sakamoto; Takashi Yamane
Journal:  J Artif Organs       Date:  2005       Impact factor: 1.731

4.  Use of an intrapericardial, continuous-flow, centrifugal pump in patients awaiting heart transplantation.

Authors:  Keith D Aaronson; Mark S Slaughter; Leslie W Miller; Edwin C McGee; William G Cotts; Michael A Acker; Mariell L Jessup; Igor D Gregoric; Pranav Loyalka; O H Frazier; Valluvan Jeevanandam; Allen S Anderson; Robert L Kormos; Jeffrey J Teuteberg; Wayne C Levy; David C Naftel; Richard M Bittman; Francis D Pagani; David R Hathaway; Steven W Boyce
Journal:  Circulation       Date:  2012-05-22       Impact factor: 29.690

5.  Acquired von Willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation.

Authors:  Nir Uriel; Sang-Woo Pak; Ulrich P Jorde; Brigitte Jude; Sophie Susen; Andre Vincentelli; Pierre-Vladimir Ennezat; Sarah Cappleman; Yoshifumi Naka; Donna Mancini
Journal:  J Am Coll Cardiol       Date:  2010-07-02       Impact factor: 24.094

6.  EVAHEART: an implantable centrifugal blood pump for long-term circulatory support.

Authors:  Kenji Yamazaki; Shinichiro Kihara; Takehide Akimoto; Osamu Tagusari; Akihiko Kawai; Mitsuo Umezu; Jun Tomioka; Robert L Kormos; Bartley P Griffith; Hiromi Kurosawa
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2002-11

7.  Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantation.

Authors:  O H Frazier; E A Rose; M C Oz; W Dembitsky; P McCarthy; B Radovancevic; V L Poirier; K A Dasse
Journal:  J Thorac Cardiovasc Surg       Date:  2001-12       Impact factor: 5.209

8.  Acquired von Willebrand syndrome in patients with an axial flow left ventricular assist device.

Authors:  Anna L Meyer; Doris Malehsa; Christoph Bara; Ulrich Budde; Mark S Slaughter; Axel Haverich; Martin Strueber
Journal:  Circ Heart Fail       Date:  2010-08-25       Impact factor: 8.790

9.  Initial experience of conversion of Toyobo paracorporeal left ventricular assist device to DuraHeart left ventricular assist device.

Authors:  Daisuke Yoshioka; Taichi Sakaguchi; Shunsuke Saito; Shigeru Miyagawa; Hiroyuki Nishi; Yasushi Yoshikawa; Satsuki Fukushima; Takayoshi Ueno; Toru Kuratani; Yoshiki Sawa
Journal:  Circ J       Date:  2011-11-27       Impact factor: 2.993

10.  Initial clinical experience with the Jarvik 2000 implantable axial-flow left ventricular assist system.

Authors:  O H Frazier; Timothy J Myers; Igor D Gregoric; Tehreen Khan; Reynolds Delgado; Mihai Croitoru; Kathy Miller; Robert Jarvik; Stephen Westaby
Journal:  Circulation       Date:  2002-06-18       Impact factor: 29.690

View more
  5 in total

1.  Journal of Artificial Organs 2014: the year in review.

Authors:  Y Sawa; K Matsuda; E Tatsumi; G Matsumiya; T Abe; K Fukunaga; A Kishida; K Kokubo; T Masuzawa; A Myoui; M Nishimura; T Nishimura; T Nishinaka; E Okamoto; S Tokunaga; T Tomo; T Tsukiya; Y Yagi; T Yamaoka
Journal:  J Artif Organs       Date:  2015-02-21       Impact factor: 1.731

Review 2.  Artificial hearts-recent progress: republication of the article published in the Japanese Journal of Artificial Organs.

Authors:  Masahiro Nishida
Journal:  J Artif Organs       Date:  2017-06-15       Impact factor: 1.731

3.  Left Ventricular Assist Device Infections: A Systematic Review.

Authors:  John C O'Horo; Omar M Abu Saleh; John M Stulak; Mark P Wilhelm; Larry M Baddour; M Rizwan Sohail
Journal:  ASAIO J       Date:  2018 May/Jun       Impact factor: 2.872

4.  Clinical effectiveness of therapy with continuous-flow left ventricular assist devices in nonischemic versus ischemic cardiomyopathy: a systematic review and meta-analysis.

Authors:  Christopher Wavell; Andrew Sokolowski; Michelle L Klingel; Charles Yin; A Dave Nagpal
Journal:  Can J Surg       Date:  2021-01-26       Impact factor: 2.089

5.  Use of durable left ventricular assist devices for high-risk patients: Korean experience before insurance coverage.

Authors:  Jun Ho Lee; Ilkun Park; Heemoon Lee; Kiick Sung; Young Tak Lee; Darae Kim; Jeong Hoon Yang; Jin-Oh Choi; Eun-Seok Jeon; Yang Hyun Cho
Journal:  J Thorac Dis       Date:  2020-12       Impact factor: 2.895

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.